Jump to ContentJump to Main Navigation

Online

249,00 € / $374.00*

* Prices subject to change. Shipping costs will be added if applicable.
Publication Date:
February 2011
ISSN:
1437-4331
DOI:
10.1515/cclm.2011.096

See all formats and pricing

Online
Individual Subscription Online only
Euro [D] 249.00
RRP for USA, Canada, Mexico
US$ 374.00 *
Print
Individual Subscription Online only
Euro [D] 1577.00
RRP for USA, Canada, Mexico
US$ 2365.00 *
Print + Online
Individual Subscription Online only
Euro [D] 1893.00
RRP for USA, Canada, Mexico
US$ 2838.00 *
*Prices subject to change. Shipping costs will be added if applicable.

Clinical Chemistry and Laboratory Medicine (CCLM)

Published in Association with the International Federation of Clinical Chemistry and Laboratory Medicine and the European Federation of Clinical Chemistry and Laboratory Medicine

Editor-in-Chief: Plebani, Mario

Editorial Board Member: Lippi, Giuseppe / Gillery, Philippe / Kazmierczak, Steven / Lackner, Karl J. / Melichar, Bohuslav / Siest, Gérard / Whitfield, John B. / Abi Fadel, Marianne / Alvarez Menendez, Francisco V. / Azzazy, Hassan M.E. / Diamandis, Eleftherios P. / Eckardstein, Arnold / Favaloro, Emmanuel J. / Griesmacher, Andrea / Herrmann, Wolfgang / Hoffmann, Johannes J.M.L. / Hooijkaas, Herbert / Ichihara, Kiyoshi / Kaabachi, Naziha / Kim, Jeong-Ho / Korte, Wolfgang / Kroupis, Christos / Lai, Leslie Charles / Lam, Wai Kei Christopher / Marc, Janja / Miyoshi, Eiji / Özben, Tomris / Palicka, Vladimir / Panteghini, Mauro / Queralto, Jose M. / Scartezini, Marileia / Simundic, Ana-Maria / Tsongalis, Gregory J. / Wallemacq, Pierre E. / Yan, Shengkai / Young, Ian S. / Chiu, Rossa Wai Kwun / Ghosh, Debabrata / Kappelmayer, Janos / Lehmann, Sylvain / Sypniewska, Grazyna

12 Issues per year

Increased IMPACT FACTOR 2011: 2.150
Rank 10 out of 32 in category Medical Laboratory Technology in the 2011 Thomson Reuters Journal Citation Report/Science Edition

VolumeIssuePage

Issues

Evaluation of the Menarini/ARKRAY ADAMS A1c HA-8180V analyser for HbA1c

1 / Erna Lenters-Westra2 / Hans van der Vuurst1 / Robbert Slingerland2 / Carla Siebelder1 / Willeke Visser-Dekkers1

1European Reference Laboratory, Streekziekenhuis Koningin Beatrix, Winterswijk, The Netherlands

2European Reference Laboratory, Isala Klinieken, Zwolle, The Netherlands

Corresponding author: Cas Weykamp, Streekziekenhuis Koningin Beatrix, Beatrixpark 1, 7101 BN Winterswijk, The Netherlands Phone: +31 543 544758, Fax: +31 543 524265

Citation Information: Clinical Chemistry and Laboratory Medicine. Volume 49, Issue 4, Pages 647–651, ISSN (Online) 1437-4331, ISSN (Print) 1434-6621, DOI: 10.1515/cclm.2011.096, February 2011

Publication History:
Received:
2010-08-31
Accepted:
2010-10-12
Published Online:
2011-02-17

Abstract

Background: We report an evaluation of the Menarini/ARKRAY ADAMS A1c HA-8180V analyser (HA-8180V), the fifth generation Menarini/ARKRAY ion-exchange HPLC for the measurement of HbA1c.

Methods: We evaluated the analytical performance, the measurement of haemoglobin variants and the performance in comparison to major analytical methods.

Results: Within-run, between-run and total CV were 0.2%, 0.4% and 0.7% at low HbA1c concentrations and 0.2%, 0.2% and 0.4% at high HbA1c concentrations, respectively. Trueness revealed a maximum deviation of 0.8 mmol/mol (IFCC units) or 0.1% (NGSP units) over the relevant analytical range. Linearity, carry-over and linear drift were excellent. Labile-HbA1c, carbamylated haemoglobin, icteric samples and variation in hematocrit did not affect HbA1c outcome. Haemoglobin variants AS, AC and F do not affect HbA1c outcome and are explicitly identified and correctly quantified. HbA1c can not be measured in samples with AE and AD, but these variants are identified correctly. In comparison to other methods used at present, the HA-8180V shows excellent performance.

Conclusions: The HA-8180V performs at a high level and is fit for any clinical application.

Keywords: diabetes; evaluation; HbA1c; Hb variants; Menarini/ARKRAY ADAMS A1c HA-8180V analyser