Jump to ContentJump to Main Navigation

Online

249,00 € / $374.00*

* Prices subject to change. Shipping costs will be added if applicable.
Publication Date:
February 2011
ISSN:
1437-4331
DOI:
10.1515/cclm.2011.113

See all formats and pricing

Online
Individual Subscription Online only
Euro [D] 249.00
RRP for USA, Canada, Mexico
US$ 374.00 *
Print
Individual Subscription Online only
Euro [D] 1577.00
RRP for USA, Canada, Mexico
US$ 2365.00 *
Print + Online
Individual Subscription Online only
Euro [D] 1893.00
RRP for USA, Canada, Mexico
US$ 2838.00 *
*Prices subject to change. Shipping costs will be added if applicable.

Clinical Chemistry and Laboratory Medicine (CCLM)

Published in Association with the International Federation of Clinical Chemistry and Laboratory Medicine and the European Federation of Clinical Chemistry and Laboratory Medicine

Editor-in-Chief: Plebani, Mario

Editorial Board Member: Lippi, Giuseppe / Gillery, Philippe / Kazmierczak, Steven / Lackner, Karl J. / Melichar, Bohuslav / Siest, Gérard / Whitfield, John B. / Abi Fadel, Marianne / Alvarez Menendez, Francisco V. / Azzazy, Hassan M.E. / Diamandis, Eleftherios P. / Eckardstein, Arnold / Favaloro, Emmanuel J. / Griesmacher, Andrea / Herrmann, Wolfgang / Hoffmann, Johannes J.M.L. / Hooijkaas, Herbert / Ichihara, Kiyoshi / Kaabachi, Naziha / Kim, Jeong-Ho / Korte, Wolfgang / Kroupis, Christos / Lai, Leslie Charles / Lam, Wai Kei Christopher / Marc, Janja / Miyoshi, Eiji / Özben, Tomris / Palicka, Vladimir / Panteghini, Mauro / Queralto, Jose M. / Scartezini, Marileia / Simundic, Ana-Maria / Tsongalis, Gregory J. / Wallemacq, Pierre E. / Yan, Shengkai / Young, Ian S. / Chiu, Rossa Wai Kwun / Ghosh, Debabrata / Kappelmayer, Janos / Lehmann, Sylvain / Sypniewska, Grazyna

12 Issues per year

Increased IMPACT FACTOR 2011: 2.150
Rank 10 out of 32 in category Medical Laboratory Technology in the 2011 Thomson Reuters Journal Citation Report/Science Edition

VolumeIssuePage

Issues

Pseudohypocalcemia caused by perchlorate (Irenat®)

Jürgen Durner1 / Ursula Winkler-Budenhofer2 / Simon Gahr2 / Walter Samtleben2 / 2

1Institute of Clinical Chemistry, University Hospital Munich-Grosshadern, Ludwig-Maximilians-University, Munich, Germany

2Nephrology Division, Department I of Internal Medicine, University Hospital Munich-Grosshadern, Ludwig-Maximilians-University, Munich, Germany

Corresponding author: Ulf Schönermarck, MD, Department I of Internal Medicine, Nephrology Division, University Hospital Munich, Campus Grosshadern, Ludwig-Maximilians-University Munich, Marchioninistr. 15, 81377 Munich, Germany Phone: +49-89-7095-2212, Fax: +49-89-7095-5317

Citation Information: Clinical Chemistry and Laboratory Medicine. Volume 49, Issue 4, Pages 665–667, ISSN (Online) 1437-4331, ISSN (Print) 1434-6621, DOI: 10.1515/cclm.2011.113, February 2011

Publication History:
Received:
2010-09-17
Accepted:
2010-11-17
Published Online:
2011-02-23

Abstract

Background: Blood gas analysis (BGA), including measurement of ionized calcium, is performed routinely in patients with end stage renal disease on renal replacement therapy, especially when using citrate for regional anticoagulation. After installation of a new blood gas analyzer (RAPIDpoint® 405; BGA), we observed lower ionized calcium concentrations in a few patients without signs of hypocalcemia, whereas calcium concentrations were normal using a standard laboratory method. Pseudohypocalcemia was of limited duration and correlated with the short-term intake of sodium perchlorate monohydrate (Irenat®).

Methods: We prepared dilution series from whole blood samples and stock solutions of calcium and perchlorate with different concentrations of ionized calcium and perchlorate. Measurement of ionized calcium concentrations was performed using two different blood gas analyzers (RAPIDpoint® 405; BGA and Roche AVL 9180; standard laboratory method).

Results: After addition of different amounts of perchlorate, significant lower ionized calcium concentrations were measured with BGA compared to the standard laboratory method using either preparations from whole blood samples or stock solutions. The addition of potassium or methylene blue known to complex perchlorate had no effect on the concentrations of ionized calcium measured with BGA. Using different mathematical methods, a calculation of the “real” ionized calcium concentration from the value measured with BGA was not possible.

Conclusions: Based on our experiments, we confirm the hypothesis that perchlorate can influence the measurement of ionized calcium by BGA. As the effect depends on the ion selective electrode that is used, it is advisable to test the blood gas analyzer with calcium and perchlorate solutions.

Keywords: calcium; citrate anticoagulation; hemodialysis; perchlorate; pseudohypocalcemia

Comments (0)