Jump to ContentJump to Main Navigation

Online

192,00 € / $269.00*

* Prices subject to change. Shipping costs will be added.
Publication Date:
28 09 2011
ISSN:
1437-4331
DOI:
10.1515/cclm.2011.721

See all formats and pricing

Print
List price
Euro [D] 1487.00
RRP for USA, Canada, Mexico
US$ 2231.00 *
Online
List price
Euro [D] 192.00
RRP for USA, Canada, Mexico
US$ 269.00 *
Print + Online
List price
Euro [D] 1710.00
RRP for USA, Canada, Mexico
US$ 2565.00 *
*Prices subject to change. Shipping costs will be added.

Clinical Chemistry and Laboratory Medicine (CCLM)

Published in Association with the International Federation of Clinical Chemistry and Laboratory Medicine and the European Federation of Clinical Chemistry and Laboratory Medicine

Editor-in-Chief: Plebani, Mario

null Lippi, Giuseppe / Gillery, Philippe / Kazmierczak, Steven / Lackner, Karl J. / Melichar, Bohuslav / Siest, Gérard / Whitfield, John B. / Abi Fadel, Marianne / Alvarez Menendez, Francisco V. / Azzazy, Hassan M.E. / Diamandis, Eleftherios P. / Eckardstein, Arnold / Favaloro, Emmanuel J. / Griesmacher, Andrea / Herrmann, Wolfgang / Hoffmann, Johannes J.M.L. / Hooijkaas, Herbert / Ichihara, Kiyoshi / Kaabachi, Naziha / Kim, Jeong-Ho / Korte, Wolfgang / Kroupis, Christos / Lai, Leslie Charles / Lam, Wai Kei Christopher / Marc, Janja / Miyoshi, Eiji / Özben, Tomris / Palicka, Vladimir / Panteghini, Mauro / Queralto, Jose M. / Scartezini, Marileia / Simundic, Ana-Maria / Tsongalis, Gregory J. / Wallemacq, Pierre E. / Yan, Shengkai / Young, Ian S. / Chiu, Rossa Wai Kwun / Ghosh, Debabrata / Kappelmayer, Janos / Lehmann, Sylvain / Sypniewska, Grazyna

12 Issues per year

Increased IMPACT FACTOR 2010: 2.069
Rank 9 out of 30 in category Medical Laboratory Technology in the 2010 Thomson Reuters Journal Citation Report/Science Edition

VolumeIssuePage

Issues

Comparison of two immunoassays for determining hepatitis B virus serum markers

Xu, Wanzhou 1 / 1 / Wang, Ming 1 / Gu, Jian 1

1Department of Clinical Laboratory, Renmin Hospital of Wuhan University, Wuhan, Hubei Province 430060, China

Corresponding author: Yan Li, Department of Clinical Laboratory, Renmim Hospital of Wuhan University, 99 Ziyang Road, Wuchang, Wuhan, Hubei Province 430060, P.R. China Phone: +86-27-880 71553, Fax: +86-27-880 71553

Citation Information: Clinical Chemistry and Laboratory Medicine. Volume 50, Issue 1, Pages 153–157, ISSN (Online) 1437-4331, ISSN (Print) 1434-6621, DOI: 10.1515/cclm.2011.721, September 2011

Publication History:

Received: 03/07/2011;
Accepted: 28/08/2011;
Published Online: 22/03/2012

Abstract

Background: To evaluate and compare the detection efficacy of serum hepatitis B virus (HBV) markers by two immunoassays: electrochemiluminescence immunoassay (ECLIA) and enzyme-linked immunosorbent assay (ELISA).

Methods: ECLIA and ELISA were used to analyze 359 serum samples, including 64 HBsAg/anti-HBs coexistence serological pattern samples (samples positive for both HBsAg and anti-HBs), 24 HBeAg/anti-HBe coexistence serological pattern samples (samples positive for both HBeAg and anti-HBe), and 271 normal serological pattern samples (negative for either of the combinations).

Results: In the normal serological pattern samples, the concordance rates of the two methods in detecting serum HBV markers were as follows: 97.05% for HBsAg, 92.62% for anti-HBs, 100% for HBeAg, 76.75% for anti-HBe, and 58.67% for anti-HBc. The differences in the qualitative criteria for anti-HBc and anti-HBe were primarily responsible for the discrepancy between the two methods (κ-values of 0.657 and 0.253, respectively). Most weak positive results, determined by ECLIA, were negative determined by ELISA, whereas the results of HBsAg, anti-HBs, and HBeAg detection were generally consistent. In the HBsAg/anti-HBs coexistence serological pattern samples, the concordance rates of HBsAg and anti-HBs detection were 98.44% and 34.38%, respectively. The positive rate of ELISA does not vary as the COI (cut-off index) varies which was determined by ECLIA; in the HBeAg/anti-HBe coexistence serological pattern samples, the concordance rates of HBeAg and anti-HBe detection were 45.83% and 79.17%, respectively. Most weak positive results, determined by ECLIA, were negative when determined by ELISA.

Conclusion: The discrepancies between the two assays in normal serological patterns samples and HBeAg/anti-HBe coexistence serological pattern samples were mostly due to the different sensitivity of the two assays, but for the HBsAg/anti-HBs coexistence serological pattern samples, the discrepancy was not caused by the different sensitivity. It is the difference in determining anti-HBs which led to the discrepancy.

Keywords: anti-HBe; anti-HBs; electrochemiluminescence immunoassay; enzyme-linked immunosorbent assay; HBeAg; HBsAg; serum HBV markers