Application of WLS strips for position determination in strip PET tomograph based on plastic scintillators

Jerzy Smyrski
  • Corresponding author
  • Institute of Physics, Jagiellonian University, Cracow, Poland
  • Email:
/ Paweł Moskal
  • Institute of Physics, Jagiellonian University, Cracow, Poland
/ Tomasz Bednarski
  • Institute of Physics, Jagiellonian University, Cracow, Poland
/ Piotr Białas
  • Institute of Physics, Jagiellonian University, Cracow, Poland
/ Eryk Czerwiński
  • Institute of Physics, Jagiellonian University, Cracow, Poland
/ łukasz Kapłon
  • Institute of Physics, Jagiellonian University, Cracow, Poland
  • Faculty of Chemistry, Jagiellonian University, Cracow, Poland
/ Andrzej Kochanowski
  • Faculty of Chemistry, Jagiellonian University, Cracow, Poland
/ Grzegorz Korcyl
  • Institute of Physics, Jagiellonian University, Cracow, Poland
/ Jakub Kowal
  • Institute of Physics, Jagiellonian University, Cracow, Poland
/ Paweł Kowalski
  • Świerk Computing Centre, National Centre for Nuclear Research, Otwock-Świerk, Poland
/ Tomasz Kozik
  • Institute of Physics, Jagiellonian University, Cracow, Poland
/ Wojciech Krzemień
  • Institute of Physics, Jagiellonian University, Cracow, Poland
/ Marcin Molenda
  • Faculty of Chemistry, Jagiellonian University, Cracow, Poland
/ Szymon Niedźwiecki
  • Institute of Physics, Jagiellonian University, Cracow, Poland
/ Marek Pałka
  • Institute of Physics, Jagiellonian University, Cracow, Poland
/ Monika Pawlik-Niedzwiecka
  • Institute of Physics, Jagiellonian University, Cracow, Poland
/ Lech Raczyński
  • Świerk Computing Centre, National Centre for Nuclear Research, Otwock-Świerk, Poland
/ Zbigniew Rudy
  • Institute of Physics, Jagiellonian University, Cracow, Poland
/ Piotr Salabura
  • Institute of Physics, Jagiellonian University, Cracow, Poland
/ Neha Gupta-Sharma
  • Institute of Physics, Jagiellonian University, Cracow, Poland
/ Michał Silarski
  • Institute of Physics, Jagiellonian University, Cracow, Poland
/ Artur Słomski
  • Institute of Physics, Jagiellonian University, Cracow, Poland
/ Adam Strzelecki
  • Institute of Physics, Jagiellonian University, Cracow, Poland
/ Wojciech Wiślicki
  • Świerk Computing Centre, National Centre for Nuclear Research, Otwock-Świerk, Poland
/ Marcin Zieliński
  • Institute of Physics, Jagiellonian University, Cracow, Poland
/ Natalia Zoń
  • Institute of Physics, Jagiellonian University, Cracow, Poland
Published Online: 2014-05-24 | DOI: https://doi.org/10.1515/bams-2013-0103

Abstract

A method of the determination of a γ-quantum absorption point in a plastic scintillator block using a matrix of wavelength-shifting (WLS) strips is proposed. An application of this method for the improvement of position resolution in newly proposed positron emission tomography (PET) detectors based on plastic scintillators is presented. The method enables to reduce parallax errors in the reconstruction of images, which occurs in the presently used PET scanners.

Keywords: scintillator detectors; time-of-flight positron emission tomography

References

  • 1.

    Denisov S, Dzierba A, Heinz R, Klimenko A, Samoylenko V, Scott E, et al. Timing characteristics of scintillator bars. Nucl Instrum Methods A 2002;478:440–3.

  • 2.

    Anton G, Arends J, Beulertz W, Hey J. Amadeus – a new type of large area scintillation detector with position-, energy- and time-of-flight determination. Nucl Instrum Methods A 1991;310:631–5.

  • 3.

    Brauksiepe S, Grzonka D, Kilian K, Oelert W, Roderburg E, Rook M, et al. COSY-11, an internal experimental facility for threshold measurements. Nucl Instrum Methods A 1996;376:397–410.

  • 4.

    Moskal P, Salabura P, Silarski M, Smyrski J, Zdebik J, Zieliński M. Novel detector systems for the positron emission tomography. Bio-Algorithms Med-Systems 2011;7:73; arXiv:1305.5187.

  • 5.

    Moskal P, Bednarski T, Bialas P, Ciszewska M, Czerwinski E, Heczko A, et al. Strip-PET: a novel detector concept for the TOF-PET scanner, Nucl Med Rev 2012;15:C68; arXiv: 1305.5562.

  • 6.

    Moskal P, Bednarski T, Bialas P, Ciszewska M, Czerwinski E, Heczko A, et al. TOF-PET detector concept based on organic scintillators. Nucl Med Rev 2012;15:C81; arXiv:1305.5559.

  • 7.

    Awes TC, Baktash C, Cumby RP, Ferguson RL, Franz A, Gabriel TA, et al. The mid-rapidity calorimeter for the relativistic heavy-ion experiment WA80 at CERN. Nucl Instrum Methods A 1989;279:479–502.

  • 8.

    Abdallah J, Adragna P, Alexa C, Alves R, Amaral P, Ananiev A, et al. Mechanical construction and installation of the ATLAS tile calorimeter. J Instrum 2013;8:T11001. [Web of Science] [Crossref]

  • 9.

    Mineev O, Kudenko Y, Musienko Y, Polyansky I, Yershov N, et al. Scintillator detectors with long WLS fibers and multi-pixel photodiodes. arXiv:1110.2651v1. [Web of Science]

  • 10.

    Maeda J, Matsubara T, Nitta K, Kuze M. Development of cosmic-ray tracker for KASKA neutrino oscillation experiment. Physica E 2007;40:425–9. [Web of Science]

  • 11.

    Stephan AC, Dai S, Wallace SA. Neutronics aspects of position-sensitive neutron scintillator detectors using wavelength-shifting readout fibers. Appl Radiat Isotopes 2004;61:1375–82.

  • 12.

    Nakamara T, Kawasaki T, Hosoya T, Toh K, Ebine M, Birumachi A, et al. A wavelength-shifting-fibre-based scintillator neutron detector implemented with the median point calculation method. J Instrum 2012;7:C02003. [Crossref] [Web of Science]

  • 13.

    Belcari N, Damiani C, Del Guerra A, Di Domenico G, Motta A, Sabba N, et al. Measurement of photoelectron yield from scintillating fibers coupled to a YAP:Ce matrix. Nucl Instrum Methods A 2001;461:413–5.

  • 14.

    Braem A, Chesi E, Joram C, Séguinot J, Weilhammer P, De Leo R, et al. High precision coordinate readout for an axial 3-D PET detector module using a wave length shifter strip matrix. Nucl Instrum Methods A 2007;580:1513–21.

  • 15.

    Saint-Gobain Company. Product Data Sheets. Available at: http://www.detectors.saint-gobain.com/Plastic-Scintillator.aspx.

  • 16.

    Knoll GK. Radiation detection and measurement, 3rd ed. New York: John Wiley & Sons, 1999:251.

About the article

Corresponding author: Jerzy Smyrski, Institute of Physics, Jagiellonian University, 30-059 Cracow, Poland, E-mail:


Received: 2013-10-22

Accepted: 2014-02-26

Published Online: 2014-05-24

Published in Print: 2014-06-30


Citation Information: Bio-Algorithms and Med-Systems, ISSN (Online) 1896-530X, ISSN (Print) 1895-9091, DOI: https://doi.org/10.1515/bams-2013-0103. Export Citation

Citing Articles

Here you can find all Crossref-listed publications in which this article is cited. If you would like to receive automatic email messages as soon as this article is cited in other publications, simply activate the “Citation Alert” on the top of this page.

[1]
P. Moskal, N. Zoń, T. Bednarski, P. Białas, E. Czerwiński, A. Gajos, D. Kamińska, Ł. Kapłon, A. Kochanowski, G. Korcyl, J. Kowal, P. Kowalski, T. Kozik, W. Krzemień, E. Kubicz, Sz. Niedźwiecki, M. Pałka, L. Raczyński, Z. Rudy, O. Rundel, P. Salabura, N.G. Sharma, M. Silarski, A. Słomski, J. Smyrski, A. Strzelecki, A. Wieczorek, W. Wiślicki, and M. Zieliński
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2015, Volume 775, Page 54

Comments (0)

Please log in or register to comment.
Log in