European Multicentre Evaluation of the Super Aution SA-4220 Urinalysis Analyser

Román Galimany, Marta Arambarri, Carmen Biosca, Ferruccio Ceriotti, Jean Pierre Chapelle, Rosa Lopez, John H. M. Souverijn, Giovanni Spagnuolo, Francesco Scopacasa and Lourdes Almirall
From the journal

Abstract

A multicentre evaluation of the urine test strip analyser Super Aution-4220 was carried out in six laboratories. The analytical performance of the instrument with regard to imprecision, linearity, detection limit, drift, carry-over and method comparison was studied. Using the Aution stick 8 test strip the pH, glucose, protein, ketones, bilirubin, blood, urobilinogen and leukocyte esterase were analysed. Specific gravity measurements were performed by refractive index method. Within-run and between-run imprecision determined at three levels of analyte were good. No carry-over was observed. Obtained results were linear through all the described analytical range. No significant drift was detected. Method comparison with some quantitative methods was performed and showed a good correlation with most of the analytes. The study of interferences showed minor interferences by common therapeutic drugs with the measurement of some analytes. During the assessment period of about 6 months no breakdown occurred in any laboratory. The Super Aution urine analyser appeared to be a highly automated analyser of urinary test strips. The operation was simple and the maintenance required only a few minutes a day.

: