Evaluation of 18 quality indicators from the external quality assurance preanalytical programme of the Spanish Society of Laboratory Medicine (SEQCML)

Andrea Caballero ORCID logo, Rubén Gómez-Rioja ORCID logo, Montserrat Ventura, María Antonia Llopis, Josep Miquel Bauça, Carolina Gómez-Gómez, Itziar Marzana ORCID logo and Mercedes Ibarz ORCID logo

Abstract

Objectives

Most errors in laboratory medicine occur in the pre- and post-analytical phases of the total testing process (TTP). In 2014, the Spanish Society of Laboratory Medicine (SEQCML) started the current Preanalytical Phase EQA Programme, with the objective of providing a tool for the improvement of the preanalytical phase. The aim of this study was to review the evolution of quality indicators (QI) and the comparability of established performance specifications (PS) with other EQA programmes.

Methods

In the SEQCML programme, participants were asked to register rejections of the main specimens and the causes for rejections. Data collected from 2014 to 2017, and then reviewed biennially (2018–2019), was used to calculate the percentiles; p25, p50, p75, and p90 for every round, and their means were set as PS. These PS were compared with the results of other programmes.

Results

The evolution of QI results for 2018–2019 period showed general maintenance or improvement, e.g., a significant decrease in the number of serum samples with a haemolytic index ≥0.5 g/L, except for EDTA and citrate samples handle, maybe for an improvement in detection. The comparison with PS for the QI of the IFCC Working Group “Laboratory Errors and Patient Safety” and the Key Incident Management and Monitoring System (KIMMS) programme of the RCPA showed comparable results, supporting the validity of the established specifications.

Conclusions

The PS obtained are a helpful tool for benchmarking and to identify processes of the preanalytical phase whose improvement should be set as a priority.

Introduction

A substantial body of evidence has demonstrated that most errors in laboratory medicine occur in the pre- and post-analytical phases of the total testing process (TTP) which, in turn, translate into a higher risk for the patient [1]. Although the concept of the brain-to-brain laboratory loop was described more than 45 years ago [2], awareness and consensus on the importance of extra-analytical aspects in laboratory quality are a more recent achievement [3, 4]. All steps of the TTP should be continuously monitored and evaluated to ensure that high quality laboratory results are provided [5].

The establishment of internal quality monitoring systems [6] and the participation in external quality assurance (EQA) programmes represent essential tools for the continuous improvement of these processes [7, 8]. Quality indicators (QI), managed as a part of laboratory improvement strategy, have proven to be a suitable tool in monitoring and improving performance in the extra-analytical phase [9, 10]. In fact, according to the ISO 15189:2012, the identification and use of effective QIs in all phases of the TTP is an essential requirement for laboratory accreditation [8], ideally checking the entire examination process, including pre and post-examination procedures [11].

In the last decades, three different types of EQA schemes for the preanalytical phase have been developed in some countries (Table 1) [12]: type I; registration of procedures, type II; circulation of samples simulating errors, and type III; registration of errors/adverse events. In this latter case, the EQA organizer should also lead the harmonization of QIs for a valid comparison of error rates and provide performance specifications (PS). Three of the programmes included in this type have established PS based on the state of the art; the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) [13], the Royal College of Pathologists of Australasia (RCPA) [14], and the Spanish Society of Laboratory Medicine (SEQCML) programme.

Table 1:

EQA programmes for the extra-analytical phase (modified from Kristensen et al. 2014 [12]).

Type Programme Organization Country Starting year Web site
Type I Preanalytical phase EQA programme SEQCML Spain 2001a
NOKLUS preanalytical EQAs NOKLUSb Norway 2008 http://www.noklus.no
Four preanalytical EQAs programmes (online training) INSTAND e.V Germany 2011 https://www.instand-ev.de/en
Preanalytics programmes in clinical chemistry, anatomic pathology, microbiology, urine, and blood sample collection and POCTc Labquality Finland 2014 https://www.labquality.fi/en/eqas/?lang=en
Type II Preanalytical serum indices scheme WEQASd UK 2010 http://www.weqas.com
EQA programme for serum indices SEQCML Spain 2018 https://www.seqc.es/es/programas-garantia-calidad
EQA programme addressing the important preanalytical workflows applied to personalized medicine SPIDIAe and SPIDIA4P Partners from Denmark, UK, Switzerland, Sweden, Italy, Austria, Luxembourg, France, Netherlands, and Spain 2011 https://www.spidia.eu
Type III The Q-track programmes CAPf USA 1998 https://www.cap.org/laboratory-improvement/quality-management-Programmes
Key incident monitoring and management systems quality assurance (KIMMS) RCPA Australia 2009 https://rcpaqap.com.au
Preanalytical phase EQA programme SEQCML Spain 2014 https://www.seqc.es/es/programas-garantia-calidad
NOKLUS preanalytical EQAs NOKLUS Norway 2017 http://www.noklus.no
UK NEQAS pre- and post-analytical quality monitoring service UK NEQASg United Kingdom 2019 https://birminghamquality.org.uk/prepq/
Model of quality indicators of WG-LEPS IFCC Worldwide 2017 https://www.ifcc.org/ifcc-education-division/working-groups-special-projects/laboratory-errors-and-patient-safety-wg-leps/quality-indicators-project/

  1. aFinished in 2013. bNorwegian Organization for Quality Improvement of Laboratory Examinations, cPoint-of-care testing, dWales External Quality Assurance Scheme, eStandardization and Improvement of Generic Pre-analytical Tools and Procedures for In Vitro Diagnostics, fCollege of American Pathologists, gUnited Kingdom National External Quality Assessment Service.

A PS is a quantitative value that requires implementation of corrective measures when it is exceeded [15] and in line with the claims of the first Strategic Conference of the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) held in Milan in 2015, the definition of PS for the extra-analytical phase should follow the same hierarchical models as those for the analytical PS: clinical outcome, biological variability, and state of the art [9, 1518]. Nevertheless, in this phase of the TTP, the state of the art is the most feasible criterion to be readily applicable [1, 10]. Three levels of PS were also suggested to be defined, in accordance with the model developed by Fraser et al. for analytical specifications [19]: first level: individual results below the 25th percentile (p25) of value distribution, representing the best performance; second level: 50th percentile (p50) value, representing the more frequent/common performance, and third level: 75th percentile (p75), representing the worst performance [17]. Other levels might be also used, such as the 10th percentile (p10) to designate “first class laboratories”; the majority of laboratories being grouped between the p10 and 90th percentile (p90); and those with PS values outside the p90, being considered as having the “worst performance”, or alternatively, a one limit model set at the p25 to define the acceptable or unacceptable performance [20].

In 1998 the Commission on Extra-analytical Quality of the SEQCML started a pilot type I programme for the quality assurance of the preanalytical phase, which was later consolidated in 2001. Until 2013, the programme focused on the analysis of the causes of blood and urine sample rejection. The results for the first five years of the Programme (2001–2005) were published in 2006 for urine [21] and in 2008 for blood samples [22], while an evaluation of the whole 13 year period (2001–2013) was published in 2016 [7]. This programme provided information on the main modes of error in the preanalytical process in Spanish laboratories, which were haemolysis, followed by non-received and clotted for blood samples, and non-received for urine samples.

In 2014 the programme was redesigned as type III to improve user accessibility, thus enabling easier data collection and submission, and to enhance robustness by increasing the length of time for each laboratory to register rejections. Since then, data from participating laboratories has been processed and analysed, and p25, p50, p75, and p90 have been used as PS. The main objective of the programme is to offer an easy tool for the detection of the most frequent errors in the preanalytical phase, as well as the promotion of cross-comparison of results and, thus, stimulate continuous improvement [23]. Each laboratory is encouraged to implement corrective measures when the results of their QI exceed the proposed PS [15].

The aim of this study was to review the consistency and evolution of QI, first stablished in a 4 year period (2014–2017) and then reviewed biennially (2018–2019), and assess the comparability of established PS with the EQA programmes of the Working Group “Laboratory Errors and Patient Safety” (WG-LEPS) of the IFCC and the Key Incident Management and Monitoring System (KIMMS) Programme of the RCPA.

Materials and methods

Programme design

In the SEQCML preanalytical programme, participants are asked to register rejections and their causes for the main specimens observed for one month, four times per year. Then, QIs are calculated as percentage of rejections respect to the total number of the most frequently measured test in each type of sample (creatinine for serum, complete blood count for whole-blood EDTA, and prothrombin time for plasma citrate samples), although in some cases they are referred to the total number of requests when there is not a main type of specimen or test performed. QI definitions and the mean value of p25, p50, p75, and p90 used to calculate PS are described in Table 2.

Table 2:

List of quality indicators (QI) of the SEQCML preanalytical programme and their formula for calculation. Performance specifications (PS) calculated from the results of the programme from 2014 to 2017 and from 2018 to 2019, median difference between them (negative when there is an improvement in p50 and positive when there is a worsening) are shown.

Type of QI QI Formula PS 2014–2017 PS 2018–2019 Median difference
p25 p50 p75 p90 p25 p50 p75 p90
General rejections PRE-01 Total rejections/Total requests (%) 1.409 2.266 3.272 4.695 1.393 2.182 3.215 4.608 −0.084
PRE-02 Unlabelled samples/Total requests (%) 0.000 0.009 0.041 0.086 0.000 0.014 0.049 0.102 0.005
PRE-03 Misidentified samples/Total requests (%) 0.001 0.010 0.030 0.068 0.000 0.009 0.020 0.045 −0.001
Serum sample rejections PRE-04 Total serum sample rejections/Creatinine tests (%) 0.452 1.091 2.224 3.534 0.418 0.853 1.737 3.696 −0.238 a
PRE-05 Not received serum samples/Creatinine tests (%) 0.102 0.188 0.328 0.678 0.104 0.210 0.350 0.616 0.022
PRE-06 Haemolysed serum samples/Creatinine tests (%) 0.222 0.718 1.791 2.876 0.146 0.483 1.397 3.365 −0.235 a
PRE-07 Insufficient serum samples/Creatinine tests (%) 0.004 0.037 0.107 0.232 0.003 0.035 0.111 0.252 −0.002
Whole-blood EDTA sample rejections PRE-08 Total whole-blood EDTA sample rejections/Complete blood counts (%) 0.288 0.481 0.766 1.143 0.317 0.519 0.769 1.066 0.038 a
PRE-09 Not received whole-blood EDTA samples/Complete blood counts (%) 0.134 0.248 0.414 0.751 0.167 0.266 0.401 0.624 0.018
PRE-10 Insufficient whole-blood EDTA samples/Complete blood counts (%) 0.005 0.026 0.068 0.146 0.005 0.022 0.062 0.113 −0.004
PRE-11 Clotted whole-blood EDTA samples/Complete blood counts (%) 0.072 0.150 0.252 0.443 0.087 0.162 0.288 0.456 0.012
Plasma coagulation citrate rejections PRE-12 Total plasma coagulation citrate sample rejections/Prothrombin tests (%) 0.811 1.748 2.863 4.734 0.937 1.887 3.298 5.471 0.139
PRE-13 Not received plasma coagulation citrate samples/Prothrombin tests (%) 0.371 0.833 1.565 2.273 0.388 0.743 1.468 2.441 −0.090
PRE-14 Insufficient plasma coagulation citrate samples/Prothrombin tests (%) 0.112 0.430 1.028 2.028 0.150 0.465 1.230 2.629 0.035
PRE-15 Clotted plasma coagulation citrate samples/Prothrombin tests (%) 0.015 0.146 0.381 0.822 0.042 0.204 0.474 0.920 0.058
PRE-16 Haemolysed plasma coagulation citrate samples/Prothrombin tests (%) 0.000 0.000 0.077 0.384 0.000 0.006 0.095 0.557 0.006 a
Urine sample rejection PRE-17 Not received urine samples/Total requests (%) 0.351 0.804 1.359 2.017 0.393 0.877 1.370 1.869 0.073
Sample quality PRE-18 Serum samples with haemolytic index ≥0.5 g/L/Creatinine tests (%) 0.852 1.861 3.399 5.400 0.758 1.503 2.859 4.988 −0.358 a

  1. aStatistical significance in median test (p<0.05).

Definition of a rejection

A rejection was considered when one or several results from a request cannot be delivered due to preanalytical errors (a test is not performed or reported, as the specimen does not meet laboratory acceptability criteria). The rejection may be global, invalidating all the tests requested, or it may affect a specific test, allowing the rest to be reported.

In laboratory information systems (LIS), tests are usually grouped into requests or orders. In some systems, the request must correspond to a single type of specimen/sample, although in Spain it is usual for LIS to allow each order to contain tests belonging to different specimens. For this reason, the programme requires that rejections are collected according to the sample in which the determination has been made, i.e., the rejected test refers to the type of specimen/container from which it originates. The four types included in the programme are the most common specimens in the laboratory: serum (as in Spanish routine laboratories it is more common to use serum than plasma for biochemistry tests, and usually with separating gel [24]), whole blood EDTA, plasma citrate, and random urine. Table 2 shows the calculation formula (numerator and denominator) for the different QIs.

Performance specification

All data collected from each participant´s submission (four times per year) were processed and the results of each laboratory are compared with respect to the p25, p50, p75, and p90 derived from the distribution of all results of the round. A p50 skew graph is additionally provided (Figure 1).

Figure 1: 
Example of the results reports for indicator number one (PRE-01) of the pre-analytical programme of SEQCML. Four box plots are shown, in which the result of the laboratory indicator is compared with: (1) global; the percentiles calculated globally for all the laboratories, (2) type of centre; the percentiles calculated for the same type of centre (hospital, hospital + primary care, or independent laboratory), (3) ownership; the percentiles calculated for the same type of laboratory ownership (private or public/subsidised), (4) number of request; percentiles calculated for laboratories with the same number of request (25,000–50,000/year, 50,000–150,000/year, 150,000–300,000/year, 300,000–600,000/year, or >600,000/year).

Figure 1:

Example of the results reports for indicator number one (PRE-01) of the pre-analytical programme of SEQCML. Four box plots are shown, in which the result of the laboratory indicator is compared with: (1) global; the percentiles calculated globally for all the laboratories, (2) type of centre; the percentiles calculated for the same type of centre (hospital, hospital + primary care, or independent laboratory), (3) ownership; the percentiles calculated for the same type of laboratory ownership (private or public/subsidised), (4) number of request; percentiles calculated for laboratories with the same number of request (25,000–50,000/year, 50,000–150,000/year, 150,000–300,000/year, 300,000–600,000/year, or >600,000/year).

PS of the programme, calculated for the first time using 2014–2017 data, were recalculated on a biennial basis, the percentiles mean of eight submissions are used for the establishment of PS as follows: p25 as the optimum specification, p50 as the desirable specification, p75 as the minimum specification, and also p90, because it is considered that voluntarily participating in an EQA programme and remaining in the central 90% of the distribution of results can be considered a reasonable performance, in a similar way as in Biochemistry programmes where there are no higher-level specifications. The results obtained were compared to the previously established period as median variation (median test), if significant improvement was noticed, new specifications were adopted.

Data analysis

IBM SPSS v21 software was used for all statistical analyses (median test). Statistical significance was set at 0.05.

Results

A total of 63 laboratories participated in the first period between 2014 and 2017, while 72 laboratories participated in the two-year period 2018–2019. PS for both periods are shown in Table 2. A decrease was seen in PRE-04: Total serum sample rejection, PRE-06: Haemolysed serum samples, and PRE-18: Serum samples with haemolytic index ≥0.5 g/L, alongside with an increase in PRE-08: Total whole-blood EDTA sample rejections and PRE-16: Haemolysed plasma coagulation citrate samples (Table 2 and Figure 2). The other QIs show high stability with an inter-period variability below 0.090. For summary data: mean, standard deviation, coefficient of variation, and number of submissions for each percentile (p25, p50, p75, and p90), see Supplementary Material, Table S1.

Figure 2: 
Mean for annual values of the p50 for each of the quality indicators (QIs) from 2014 to 2019.

Figure 2:

Mean for annual values of the p50 for each of the quality indicators (QIs) from 2014 to 2019.

Table 3 shows the current PS from the programme compared to those established by the IFCC (with 59 participants laboratories) and the KIMMS (with 60 participants laboratories) programmes, as well as coefficient of variation for p50 along the whole period (2014–2019), being comparable with CV from the KIMMS programme. The total the specifications for those indicators that can be matched between programmes, despite the differences in the denominator, showed that the p50 of misidentified samples is lower in the SEQCML programme compared to KIMMS and IFCC. Nevertheless, the number of misidentified samples was very low in all three (≤0.100). Regarding the total number of samples not received, p50 were higher in KIMMS than those of IFCC and lower than those of SEQCML. However, taking into account the type of specimen, according to the SEQCML programme, the most frequently not received samples in Spanish laboratories are urine samples (0.877%), followed by plasma coagulation citrate samples (0.743%), whole-blood EDTA samples (0.266%), and serum samples (0.210%). The main cause of rejection of serum samples, haemolysis, showed very similar values to those of IFCC, and lower than in KIMMS. Insufficient and coagulated samples showed the same trend. For CVs, the values were higher in the SEQCML programme than in KIMMS, showing a greater variability in the laboratories participating in this programme.

Table 3:

Comparison of the performance specifications (PS) of the SEQCML programme with those of the QIs of the Working Group “Laboratory Errors and Patient Safety” (WG-LEPS) of the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) [13] and the Key Incident Management and Monitoring System (KIMMS) of the Royal College of Pathologists Australasia (RCPA) [14].

SEQCML (2018–2019) IFCC (2018) KIMMS (2015–2018)
QI p25 (CI 95%) p50 (CI 95%) p75 (CI 95%) CV (%)a QI p25 (CI 95%) p50 (CI 95%) p75 (CI 95%) QI Average CV (%)
PRE-01 Total rejections/Total requests (%) 1.393 2.182 3.215 8.6 NE NE
(1.279–1.506) (2.102–2.261) (3.005–3.426)
PRE-02 Unlabelled samples/Total requests (%) 0.000 0.014 0.049 59.2 NE Unlabelled 0.084 10
(0.000–0.000) (0.010–0.018) (0.043–0.055)
PRE-03 Misidentified samples/Total requests (%) 0.000 0.009 0.020 33.8 Pre-MisR Percentage of: Number of misidentified requests/Total number of requests 0.010 0.025 0.070 Discrepancy of ID 0.100 19
(0.000–0.001) (0.006–0.011) (0.018–0.023) (0.000–0.010) (0.020–0.030) (0.054–0.100)
PRE-04 Total serum sample rejections/Creatinine tests (%) 0.418 0.853 1.737 16.9 NE NE
(0.395–0.441) (0.785–0.920) (1.523–1.951)
PRE-05 Not received serum samples/Creatinine tests (%) 0.104 0.210 0.350 12.9 Pre-NotRec Percentage of: Number of samples not received/Total number of samples 0.090 0.190 0.889 Samples not collected 0.317 6
(0.094–0.114) (0.199–0.221) (0.325–0.375) (0.060–0.100) (0.140–0.295) (0.602–1.470)
PRE-06 Haemolysed serum samples/Creatinine tests (%) 0.146 0.483 1.397 21.2 Pre-HemR Percentage of: Number of samples rejected due to haemolysis/Total number of checked samples for haemolysis 0.049 0.435 0.882 Haemolysed sample 0.770 7
(0.103–0.189) (0.029–0.041) (1.186–1.607) (0–0.165) (0.300–0.500) (0.587–1.410)
PRE-07 Insufficient serum samples/Creatinine tests (%) 0.003 0.035 0.111 26.0 Pre-InsV Percentage of: Number of samples with insufficient volume/Total number of samples 0.010 0.030 0.110 Insufficient sample 0.165 9
(0.001–0.005) (0.029–0.041) (0.099–0.123) (0.006–0.011) (0.022–0.040) (0.079–0.130)
PRE-08 Total whole-blood EDTA sample rejections/Complete blood counts (%) 0.317 0.519 0.769 12.1 NE NE
(0.298–0.336) (0.488–0.549) (0.741–0.797)
PRE-09 Not received whole-blood EDTA samples/Complete blood counts (%) 0.167 0.266 0.401 7.9 Pre-NotRec Percentage of: Number of samples not received/Total number of samples 0.090 0.190 0.889 Samples not collected 0.317 6
(0.155–0.178) (0.256–0.257) (0.392–0.410) (0.060–0.100) (0.140–0.295) (0.602–1.470)
PRE-10 Insufficient whole-blood EDTA samples/Complete blood counts (%) 0.005 0.022 0.062 31.1 Pre-InsV Percentage of: Number of samples with insufficient sample/Total number of samples 0.010 0.030 0.110 Insufficient sample 0.165 9
(0.003–0.007) (0.018–0.026) (0.057–0.067) (0.006–0.011) (0.022–0.040) (0.079–0.130)
PRE-11 Clotted whole-blood EDTA samples/Complete blood counts (%) 0.087 0.162 0.288 14.1 Pre-clot Percentage of: Number of samples clotted/Total number of samples with an anticoagulant checked for clots 0.080 0.237 0.402 Clotted samples 0.193 6
(0.079–0.094) (0.148–0.175) (0.272–0.304) (0.058–0.111) (0.200–0.270) (0.350–0.525)
PRE-12 Total plasma coagulation citrate imple rejections/Prothrombin tests (%) 0.937 1.887 3.298 10.9 NE NE
(0.842–1.032) (1.775–1.999) (2.973–3.623)
PRE-13 Not received plasma coagulation citrate samples/Prothrombin time (%) 0.388 0.743 1.468 15.1 Pre-NotRec Percentage of: Number of samples not received/Total number of samples 0.090 0.190 0.889 Samples not collected 0.317 6
(0.342–0.433) (0.681–0.805) (1.348–1–588) (0.060–0.100) (0.140–0.295) (0.602–1.470)
PRE-14 Insufficient plasma coagulation citrate samples/Prothrombin time (%) 0.150 0.465 1.230 24.7 Pre-SaAnt Percentage of: Number of samples with inappropriate volume-anticoagulant ratio/Total number of samples with anticoagulant 0.070 0.343 0.770 Incorrect fill of samples 0.089 21
(0.129–0.172) (0.430–0.500) (1.108–1.353) (0.040–0.114) (0.220–0.420) (0.690–0.980)
PRE-15 Clotted plasma coagulation citrate samples/Prothrombin time (%) 0.042 0.204 0.474 36.8 Pre-clot Percentage of: Number of samples clotted/Total number of samples with an anticoagulant checked for clots 0.080 0.237 0.402 Clotted samples 0.193 6
(0.022–0.062) (0.175–0.232) (0.441–0.506) (0.058–0.111) (0.200–0.270) (0.350–0.525)
PRE-16 Haemolysed plasma coagulation citrate samples/Prothrombin time (%) 0.000 0.006 0.095 247.2 Pre-HemR Percentage of: Number of samples rejected due to haemolysis/Total number of checked samples for haemolysis global 0.049 0.435 0.882 Haemolysed samples 0.770 7
(0.000–0.000) (0.000–0.011) (0.069–0.121) (0.000–0.165) (0.300–0.500) (0.587–1.410)
PRE-17 Not received urine samples/Total requests (%) 0.393 0.877 1.370 9.0 Pre-NotRec Percentage of: Number of samples not received/Total number of samples 0.090 0.190 0.889 Samples not collected 0.317 6
(0.337–0.449) (0.830–0.923) (1.325–1.415) (0.060–0.100) (0.140–0.295) (0.602–1.470)
PRE-18 Serum samples with haemolytic index ≥0.5 g/L/total serum samples where the haemolytic index is performed (%) 0.758 1.503 2.859 15.3 Pre-HemI Percentage of: Number of samples with free haemoglobin (hb) >0.5 g/L detected by automated haemolytic index/Total number of checked samples for haemolysis 0.690 1.810 3.230 NE
(0.686–0.831) (1.408–1.598) (2.460–3.259) (0.460–0.970) (1.490–2.230) (2.595–3.800)

  1. CI 95%, 95% confidence intervals; CV, coefficient of variation; NE, no equivalent indicator. aCoefficient of variation calculated from p50 results from 2014 to 2019.

A significant decrease in p50 value of total serum sample rejections indicator was observed between periods (Figure 3A), associated with a decrease in the p50 value of rejections due to haemolysis indicator (Figure 3B) (Table 2). The PS was modified to bring it in line with the state of the art of the participants. This indicator was similar to the IFCC Samples rejected due to haemolysis. The 95% confidence interval for the median of this indicator in 2018 was 0.06–0.87 [13], although in this case the indicator included all specimens and used as denominator all samples where haemolysis had been measured. In the SEQCML programme, haemolysed serum (median 0.5%) and citrate (0.006%) samples are reported separately. With respect to the KIMMS programme, the Haemolysed sample indicator presents a mean over four years period (2015–2018) of 0.770% with respect to total requests [14]. In both cases, the results are similar to those observed in the SEQCML programme (Table 3).

Figure 3: 
Values from 2014 to 2019 for each of the submissions of the p25, p50, p75, and p90.
(A) PRE-04, (B) PRE-06, (C) PRE-18.

Figure 3:

Values from 2014 to 2019 for each of the submissions of the p25, p50, p75, and p90.

(A) PRE-04, (B) PRE-06, (C) PRE-18.

The haemolysed serum samples indicator includes rejections, i.e., tests in which at least one result has been removed from the report on the assumption that the result was significantly interfered. Therefore, the prevalence of samples rejected due to haemolysis (mean 0.483%) was lower than the prevalence of samples with a haemolytic index ≥0.5 g/L (1.503%), indicative of sample deterioration (Figure 3C). Serum samples with haemolytic index ≥0.5 g/L indicator were higher than the number of rejections due to haemolysis because low-grade haemolysis can cause interferences that may not be considered significant in the biochemical test results. As with the haemolysis rejection indicator, a significant tendency to decrease was seen and therefore the specifications were modified. This indicator was similar to the IFCC Samples with free haemoglobin >0.5 g/L (automated detection), the 95% confidence interval for the median of this indicator in 2018 was 0.67–2.76 [13], using the number of samples in which haemolysis had been measured as the denominator (Table 3). With respect to the KIMMS programme, there no coincident indicator was found.

The number of rejections due to haemolysis in the citrate tube was much lower than in the serum tube (0.006% vs. 0.500%) but an significant increase in the indicator was seen between periods. Total whole blood EDTA sample rejections also showed a significant increase between periods, triggering a change in PS for both indicators. In the case of the whole blood EDTA sample, the main reason for preanalytical rejection was the coagulated sample.

The rest of the indicators showed a very stable evolution with a downward trend. The general indicators stood out, remaining at very low values. PRE-03 misidentified samples/total requests (%), which could be considered a sentinel indicator, had very low values (0.009). This indicator is similar to the percentage of: number of misidentified requests/total number of requests of the IFCC (0.025%) and the discrepancy of ID in KIMMS (0.010%). However, the number of unlabelled samples (PRE-02 unlabelled samples/total requests (%)) increased from 0.009 to 0.014%, this indicator is similar to the KIMMS Unlabelled, which has a higher PS of 0.84%, while no equivalent indicator from the IFCC was found.

Discussion

There is a general interest in QIs related to the extra-analytical phases, but only a limited number of clinical laboratories collect regular and comprehensive data. The increasing number of participants in the Spanish programme could be considered a reflection of the growing concern in these phases of the TTP, although the number of participants is still far from those enrolled in the programmes of the analytical phase of the SEQCML. To overcome the well-known “quality indicator paradox” [25], the WG-LEPS developed a Model of Quality Indicators (MQI) based on a list of consensually defined QIs (available on www.ifcc-mqi.com) [3, 17, 26] and the collection of data received by clinical laboratories led to the definition of PS in the preanalytical phase [25]. In 2017, the SEQCML calculated the PS with the results of the past four years collected from the Spanish laboratories, and then recalculated in 2018–2019. The RCPA calculated its own PS based on the results of 2015–2018 period, unlike the other programmes, the results are presented as mean rather than median.

One important drawback of the preanalytical phase quality assurance programmes is the difficulty to compare results due to the different QI formulas applied in their calculations. In this sense, a call for harmonization is needed. According to the results of the SEQCML programme, QIs calculated with respect to the most frequently measured test in each type of sample gives the most comparable results among Spanish laboratories. Besides, since the number and type of errors in the emergency department (ED) or Intensive Care Units are different and more heterogeneous from that occurred in routine samples [27, 28], the SEQCML programme includes only data from routine samples to avoid a possible confounding factor in result interpretation. However, some of the indicators are comparable among programmes and show that the PS are very similar for the IFCC and more discrepant with the KIMMS. CVs also show more variation, which could be due to a longer period of years. The SEQCML programme reflects the stability of the indicators and is therefore useful to see the evolution of each laboratory and whether data is being collected properly.

The SEQCML programme aims to calculate simple indicators and enables intercomparison by providing an objective reference among different laboratories, establishing PS based on the state of the art. This allows the identification of critical points in the entire preanalytical phase, from the request to the storage and preparation of the sample before analysis, and it is also possible to evaluate the trend over time (positive or negative), since even if the specifications are met, there may be evidence of a deterioration. Stability and/or improvement in the overall indicators (total rejections, unlabelled, and misidentified samples) could reflect the effectiveness of the programme.

The haemolysed sample is by far the main reason for rejection of serum samples. Errors in collection or handling can cause haemolysis of the sample, usually detected in the analytical phase. In fact, in recent years a great effort has been made to improve the process of blood collection, detection of the haemolysed samples, and how to establish measures that help reduce their incidence. The significant decrease observed in serum sample may indicate an improvement in preanalytical practices in participating laboratories and is considered very positive. The minimum specification of the SEQCML programme suggests that laboratories should keep their number of haemolysis rejections below 1.4% of the total serum samples. This indicator is not far from the 0.9% of the IFCC programme and implies, in any case, that many laboratories reject tests from one out of every 100 samples due to haemolysis.

The percentage of serum samples with haemolytic index ≥0.5 g/L is used as an overall QI of the preanalytical process as it indicates a deterioration of the sample during one of the preanalytical processes (collection, transport, or preparation). The improvement in this indicator could also be considered indicative of awareness of the importance of the preanalytical phase in the laboratory. Also, in this case, results of SEQCML programme and the IFCC one suggest that laboratories should maintain at least a percentage of haemolysed samples (haemolytic index ≥0.5 g/L) below 3%.

Considering the p75 (minimum specifications) and the type of sample, coagulation citrate samples QIs have the least desirable results and laboratories should take measures to reduce the number of insufficient, clotted, and haemolysed samples. On the other hand, the worsening of the results could reflect the technology improvement of the haemostasis analysers, capable of detecting more preanalytical parameters, in the same way, that happened years ago with serum samples and biochemistry analysers (haemolytic, icterus, and turbidity index). Whatever the reason, is the responsibility of the laboratories to improve QIs. According to the results, the number of clotted whole-blood EDTA samples should also be reduced.

The results between programmes show very similar values in terms of specifications, the CVs are higher compared to KIMMS, but this could be due to the way they are calculated, as in the case of the SEQCML programme possible outliers are not excluded.

In this study, the results obtained for QI between 2014 and 2017 were used to calculate PS and then compared with those in 2018–2019, to identify processes for which improvement should be a priority. In the next year, a recalculation of the data would be performed, and in the case of better results from the period 2020 to 2021, the specifications may be changed. Nonetheless, the better results generally seen in 2018–2019 reflect the awareness of clinical laboratories in preanalytical phase and the efforts to improve it. In the future, due to the rapid change in diagnostic tests and type of samples in the clinical laboratories (for instance samples for free DNA tests), the validity of the current QIs would need to be revised, and new ones could be included. In this sense, the Extra-analytical Quality Commission of the SEQCML has included a new QI in the 2021 programme: number of contaminated by infusion serum samples/Creatinine tests (%). In the meantime, the improvement of LIS tools will warrant a simplification in the collection and management of laboratory data, this together with the increase of participants in the EQA scheme, which could also do possible to include ED results.

One of the main limitations of this study is the short number of results, due to the number of participant laboratories (72 in 2018–2019 period), not comparable with the huge volume of data used in the case of analytical phase programmes (around 600 laboratories in 2019 in SERUM EQA Programme, the SEQCML programme with the highest number of participants), and the not exclusion of outliers’ results. Consequently, the robustness of the PS calculated might be compromised. Another disadvantage of this type of programme, different form the analytical phase schemes, is that in the case of being out of specifications, the measures that must be taken to solve the problem and improve the results of QI, could not be clear. An analysis of the risk associated with each situation should be carried out, this analysis would be useful to prioritize the intervention points. Once the corresponding actions were applied, a comparison of the new QI obtained with the proposed PS would allow verifying the effectiveness of the implemented measures.

In conclusion, preanalytical QIs and PS help identify the existence of errors and assess the effectiveness of implementing corrective actions, but by themselves they do not control anything or identify the causes of non-conformities. It is a shared responsibility between the scientific societies and the individual laboratories to monitor and implement the necessary measures. The SEQCML preanalytical programme offers a reliable formula to calculate the QIs, and the PS obtained are a helpful tool for benchmarking, as well as to identify processes whose improvement should be a priority for clinical laboratories.


Corresponding author: Andrea Caballero, PhD, Extra-analytical Quality Commission of the Spanish Society of Laboratory Medicine (SEQCML). Department of Clinical Biochemistry, Echevarne Laboratory, Avda. Can Bellet, 61-65 - 08174, Sant Cugat del Vallés, Spain, Phone: +34 672 828 350, E-mail:

Acknowledgments

The design and maintenance of the Quality Assurance Schemes of the Spanish Society of Laboratory Medicine (SEQCML) is carried out by the External Quality Programs Committee, composed by: Ma Carmen Perich Alsina (President), Beatriz Boned Juliani, Sandra Bullich Marín, Marià Cortés Rius, Pilar Fernández Calle, Rubén Gómez Rioja, Ma Antonia Llopis Díaz, Cecília Martínez Bru, Francisco Ramón Bauzá, Ángel Salas García, Margarida Simon Palmada, and Montserrat Ventura Alemany. And also, by the SEQCML Technical Coordinators: Berta Piqueras Andrés and Mariona Panadès Turró.

  1. Article Note: A translation of this article can be found here: https://doi.org/10.1515/almed-2022-0036.

  2. Research funding: None declared.

  3. Author contribution: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  4. Competing interests: Authors state no conflict of interest.

  5. Informed consent: Not applicable.

  6. Ethical approval: Not applicable.

References

1. Plebani, M, Astion, ML, Barth, JH, Chen, W, De Oliveira Galoro, CA, Escuer, MI, et al.. Harmonization of quality indicators in laboratory medicine. A preliminary consensus. Clin Chem Lab Med 2014;52:951–8. https://doi.org/10.1515/cclm-2014-0142.Search in Google Scholar PubMed

2. Lundberg, G. Managing the patient-focused laboratory, Reinhold, VN, editor. Cranbury: Medical Economics Co; 1975.Search in Google Scholar

3. Plebani, M, Sciacovelli, L, Aita, A, Pelloso, M, Chiozza, ML. Performance criteria and quality indicators for the pre-analytical phase. Clin Chem Lab Med 2015;53:943–8. https://doi.org/10.1515/cclm-2014-1124.Search in Google Scholar PubMed

4. Plebani, M. Errors in clinical laboratories or errors in laboratory medicine? Clin Chem Lab Med 2006;44:750–9. https://doi.org/10.1515/CCLM.2006.123.Search in Google Scholar PubMed

5. Plebani, M, Chiozza, ML, Sciacovelli, L. Towards harmonization of quality indicators in laboratory medicine. Clin Chem Lab Med 2013;51:187–95. https://doi.org/10.1515/cclm-2012-0582.Search in Google Scholar PubMed

6. Marzana Sanz, I, Ibarz Escuer, M, Llopis Diaz, MA, Barba Meseguer, N, Alsina Kirchner, MJ, Martínez Espartosa, D, et al.. Recomendaciones para el diseño e implementación de un programa de aseguramiento de la calidad de la fase preanalítica. Rev Del Lab Clínico 2019;12:e54–65. https://doi.org/10.1016/j.labcli.2019.01.003.Search in Google Scholar

7. Llopis, MA, Bauça, JM, Barba, N, Álvarez, V, Ventura, M, Ibarz, M, et al.. Spanish preanalytical quality monitoring program (SEQC), an overview of 12 years’ experience. Clin Chem Lab Med 2017;55:530–8. https://doi.org/10.1515/cclm-2016-0382.Search in Google Scholar PubMed

8. Laboratorios clínicos. Requisitos particulares para la calidad y la competencia (ISO 15189:2012). UNE-EN ISO 15189:2012.Search in Google Scholar

9. Sciacovelli, L, Lippi, G, Sumarac, Z, West, J, Garcia, I. Quality indicators in laboratory medicine: the status of the progress of IFCC working group “Laboratory Errors and Patient Safety” project. Clin Chem Lab Med 2017;55:348–57. https://doi.org/10.1515/cclm-2016-0929.Search in Google Scholar PubMed

10. Sciacovelli, L, Aita, A, Plebani, M. Extra-analytical quality indicators and laboratory performances. Clin Biochem 2017;50:632–7. https://doi.org/10.1016/j.clinbiochem.2017.03.020.Search in Google Scholar PubMed

11. Lippi, G, Becan-McBride, K, Behúlová, D, Bowen, RA, Church, S, Delanghe, J, et al.. Preanalytical quality improvement: in quality we trust. Clin Chem Lab Med 2013;51:229–41. https://doi.org/10.1515/cclm-2012-0597.Search in Google Scholar PubMed

12. Kristensen, G, Aakre, KM, Kristoffersen, AH, Sandberg, S. How to conduct external quality assessment schemes for the pre-analytical phase? Biochem Med 2014;24:114–22. https://doi.org/10.11613/bm.2014.013.Search in Google Scholar

13. Sciacovelli, L, Lippi, G, Sumarac, Z, Castro, IGDP, Ivanov, A, De Guire, V, et al.. Pre-analytical quality indicators in laboratory medicine: performance of laboratories participating in the IFCC working group “laboratory errors and patient safety” project. Clin Chim Acta 2019;497:35–40. https://doi.org/10.1016/j.cca.2019.07.007.Search in Google Scholar PubMed

14. Gay, S, Badrick, T. Changes in error rates in the Australian key incident monitoring and management system program. Biochem Med 2020;30:1–8. https://doi.org/10.11613/BM.2020.020704.Search in Google Scholar PubMed PubMed Central

15. Llopis, MA, Trujillo, G, Isabel, M, Tarre, E, Ruiz, R, Kirchner, MJA, et al.. Quality indicators and specifications for key analytical-extranalytical processes in the clinical laboratory. Five years’ experience using the Six Sigma concept. Clin Chem Lab Med 2011;49:463–70. https://doi.org/10.1515/CCLM.2011.067.Search in Google Scholar PubMed

16. Panteghini, M, Sandberg, S. Defining analytical performance specifications 15 years after the Stockholm conference. Clin Chem Lab Med 2015;53:829–32. https://doi.org/10.1515/cclm-2015-0303.Search in Google Scholar PubMed

17. Plebani, M, Task, E, Speci, P. Performance specifications for the extra-analytical phases of laboratory testing : why and how. Clin Biochem 2017;50:550–4. https://doi.org/10.1016/j.clinbiochem.2017.02.002.Search in Google Scholar PubMed

18. Panteghini, M, Ceriotti, F, Jones, G, Oosterhuis, W, Plebani, M. Strategies to define performance specifications in laboratory medicine : 3 years on from the Milan Strategic Conference. Clin Chem Lab Med 2017;55:1849–56. https://doi.org/10.1515/cclm-2017-0772.Search in Google Scholar PubMed

19. Fraser, CG, Petersen, PH, Ricosl, C. Proposals for setting generally applicable quality goals solely based on biology. Ann Clin Biochem 1997;34:8–12. https://doi.org/10.1177/000456329703400103.Search in Google Scholar PubMed

20. Sciacovelli, L, Panteghini, M, Lippi, G, Sumarac, Z, Cadamuro, J, Alex, C, et al.. Defining a roadmap for harmonizing quality indicators in laboratory medicine : a consensus statement on behalf of the IFCC working group “Laboratory Error and Patient Safety” and EFLM task and finish group “performance specifications for the extra-analytical phases”. Clin Chem Lab Med 2017;55:1478–88. https://doi.org/10.1515/cclm-2017-0412.Search in Google Scholar PubMed

21. Alsina, MJ, Álvarez, V, Barba, N, Bullich, S, Cortés, M, Escoda, I, et al.. Revisión de los resultados del programa de evaluacion externa de la calidad preanalitica (resumen 2001-2005). Quim Clin 2006;26:325–31.Search in Google Scholar

22. Alsina, M, Virtudes, A, Barba, N, Bullich, S, Cortés, M, Escoda, I, et al.. Preanalytical quality control program – an overview of results (2001–2005 summary). Clin Chem Lab Med 2008;46:849–54. https://doi.org/10.1515/CCLM.2008.168.Search in Google Scholar PubMed

23. SEQCML . Programas de garantía externa de la calidad. Available from: http://www.seqc.es/es/programas-garantia-calidad.Search in Google Scholar

24. Plebani, M, Banfi, G, Bernardini, S, Bondanini, F, Conti, L, Dorizzi, R, et al.. Serum or plasma? An old question looking for new answers. Clin Chem Lab Med 2019;58:178–87. https://doi.org/10.1515/cclm-2019-0719.Search in Google Scholar PubMed

25. Plebani, M. The quality indicator paradox. Clin Chem Lab Med 2016;54:1119–22. https://doi.org/10.1515/cclm-2015-1080.Search in Google Scholar PubMed

26. Sciacovelli, L, Kane, MO, Skaik, A, Caciagli, P, Rin, GD, Ivanov, A, et al.. Quality indicators in laboratory medicine: from theory to practice preliminary data from the IFCC Working Group Project “Laboratory Errors and Patient Safety”. Clin Chem Lab Med 2011;49:835–44. https://doi.org/10.1515/cclm.2011.128.Search in Google Scholar PubMed

27. Lippi, G, Caola, I, Cervellin, G, Milanesi, B, Morandini, M, Giavarina, D. Error rates during blood collection in emergency departments and outpatient clinics: results of a prospective multicenter study. Clin Chim Acta 2015;445:91–2. https://doi.org/10.1016/j.cca.2015.03.022.Search in Google Scholar PubMed

28. Simundic, A, Nikolac, N, Vukasovic, I, Vrkic, N. The prevalence of preanalytical errors in a Croatian ISO 15189 accredited laboratory. Clin Chem Lab Med 2010;48:1009–14. https://doi.org/10.1515/CCLM.2010.221.Search in Google Scholar PubMed


Supplementary Material

The online version of this article offers supplementary material (https://doi.org/10.1515/almed-2021-0097).


Received: 2021-12-23
Accepted: 2022-03-18
Published Online: 2022-05-18

© 2022 Andrea Caballero et al., published by De Gruyter, Berlin/Boston

This work is licensed under the Creative Commons Attribution 4.0 International License.