Abstract
Electrical impedance tomography (EIT) is used to monitor the regional distribution of ventilation in a transversal plane of the thorax. In this manuscript we evaluate the impact of different quantities of electrodes used for current injection and voltage measurement on the reconstructed shape of the lungs. Results indicate that the shape of reconstructed impedance changes in the body depends on the number of electrodes. In this manuscript, we demonstrate that a higher number of electrodes do not necessarily increase the image quality. For the used stimulation pattern, utilizing neighboring electrodes for current injection and voltage measurement, we conclude that the shape of the lungs is best reconstructed if 16 electrodes are used.
1 Introduction
Electrical impedance tomography (EIT) is an emerging medical imaging technique, which is used in a clinical context to generate tomographic images of the thorax, depicting regional ventilation of the lungs. Currently, EIT is mainly used in the intensive care unit, where the regional information about lung ventilation is used to adjust ventilator parameters and thus prevent ventilator induced lung injuries [1]. Recently it has also been used in the examination of spontaneously breathing patients suffering from obstructive lung diseases such as chronic obstructive lung disease (COPD) [2], cystic fibrosis (CF) [3] or asthma.
In EIT an array of electrodes is attached around the circumference of the thorax. It is assumed that the impedance of the lungs increases during inspiration. Electrodes are used to inject small alternating currents and measure resulting voltages.
In clinical context difference-EIT is an established method to visualize changes in impedance, rather than absolute values of impedance. This modality is less sensitive to unknown electrode contact impedance and unknown thorax shape. A reconstruction algorithm is used to generate tomographic images of impedance variations from changes in measured electrode voltages. Usually one set of voltages is recorded at end-expiration, whereas another set of voltages is traced after inspiration. The underlying changes in impedance are reconstructed from the voltage measurements and displayed in real-time. Compared to structural images techniques, such as x-ray computed tomography (CT), the images of impedance change are severely blurred and structural information cannot be obtained. Thus, in many EIT applications only specific measures, e.g. the centre of ventilation (CoV) [4] or the global inhomogeneity index (GI) [5] are used.
However, the allocation of structural lung information (e.g. derived from CT) to functional images of the lungs (from EIT) is necessary for a comprehensive diagnosis. Therefore, EIT images should preserve the position and the shape of the lungs.
Typically the number of measured voltages is by orders of magnitude lower than the number of pixels depicting the reconstructed impedance changes. Thus, it seems obvious to increase the number of electrodes (and as a result the number of measured voltages) to increase image quality. The objective of this study is to compare different quantities of electrodes used for current injection and voltage measurement regarding their preservation of lung shape in reconstructed images.
2 Methods
2.1 Simulation of voltages
Calculations have been carried out with MATLAB 2015a (Mathworks, Natick, MA) and the EIDORS toolbox [6]. Meshes of the used finite element models (FEM) were generated using NETGEN [7].
Usually the EIT problem, i.e. reconstructing internal changes in impedance (or conductivity) from voltage measurements on the surface is solved on a discretized domain. For voltage simulations we used 3D FEM models. The height of the models was set to 1.5, the semi major axis of the elliptical base area was set to 1, the semi minor axis to 0.8. Two elliptical cylinders with a semi minor axis of 0.5 and a semi major axis of 0.9 were placed inside the virtual phantom to simulate the lungs. The centre positions of the “lung cylinders” were set to −0.45 and 0.45 on the x-axis. Circular electrodes were attached equidistantly around the circumference at a height of 0.75. An exemplarily illustration of the FEM model with 32 electrodes is depicted in Figure 1.

Simulation model with 32 circularly attached electrodes and elliptical cylinders to simulate the lungs.
The conductivity of FEM elements belonging to the “lung cylinders” was varied between σexp = 74 mS/m and σinsp =43 mS/m to simulate conductivity changes during breathing. Elements not belonging to the “lungs” were set to a constant value of σbkg = 480 mS/m. The number of electrodes was varied systematically in steps of four from n = 8 to n = 32. The “adjacent injection pattern” was used to simulate voltage vectors vexp and vinsp, where neighbouring electrodes are used for current injection or voltage measurement, respectively. Current injecting and measuring electrodes are changed in a rotating manner. Voltages are not measured on the electrodes used for current injection, such that v ∈
2.2 Image reconstruction
A one-step Gauss-Newton solver was used to reconstruct conductivity changes from simulated voltage changes.
where
To enable a comparison between reconstructed images with different quantities of electrodes, we chose the hyperparameter λ such that a noise amplification of 0.5, 0.75 or 1 was achieved. The noise amplification NF is defined to be
where
2.3 Evaluation of lung shape
To evaluate the shape of the reconstructed lungs we developed the shape mismatch measure (SMM) which is defined as
with
SMM values range from 1 to −1, where SMM = 1 indicates that the area of the reconstructed lung is equal to the area of the lung used in the simulation. Therefore,
3 Results
An exemplary illustration of reconstructed images of conductivity changes with different number of electrodes is depicted in Figure 2.

Reconstruction of lung shapes with different numbers of electrodes. Left: eight electrodes; Right: 28 electrodes.
The more electrodes are used for voltage measurement, the more pronounced is the extension of the reconstructed lung shape in anterior-posterior direction (y-direction), whereas the reconstructions with only few electrodes leads to compressed lungs regarding the extension in this direction (see Figure 2). Additionally, the contours of the lungs are less blurred if more electrodes are used.
However, more electrodes do not necessarily improve the image quality, as measured with the above described SMM value. In all reconstructions the lowest SMM value was obtained for n = 8. Highest SMM values are obtained for n = 16. Figure 3 reveals that SMM values decrease if more than 16 electrodes are used for current injection and voltage measurement.

SMM values for different quantities of electrodes. Blue circle: NF = 0.5; Red X-marks: NF = 0.75; Yellow Diamond: NF = 1.
4 Discussion
In this paper, we demonstrated with a simulation study that the number of electrodes used for current injection and voltage measurement influences the shape of the reconstructed images of conductivity changes. Thereby more electrodes do not necessarily improve the image quality. For the used model the reconstructions made with 16 electrodes best preserved the shape of the lungs.
In this simulation study we used the “adjacent stimulation pattern”, which is the most often used pattern for current injection and voltage measurement. Several other patterns have been developed. The effect of the quantity of electrodes needs to be examined with other patterns in future work. Additionally, simulations were carried out on a simple ellipse shaped model and electrodes were spaced equidistantly. Further research should try to answer the question how many electrodes are necessary and where they should be placed to best depict the lung shapes of individual patients? Lung shapes might be obtained from CT data, which would lead to a patient specific EIT configuration, where quantity and placement of electrodes are determined individually. This could enable an easier correlation and allocation of anomalies in the EIT images with morphological imaging techniques and thus improve the interpretability of EIT images.
Author’s Statement
Research funding: This work was partially supported by the Federal Ministry of Education and Research (BMBF) under grant no. 03FH038I3 (MOSES). Conflict of interest: Authors state no conflict of interest. Material and Methods: Informed consent: Informed consent has been obtained from all individuals included in this study. Ethical approval: The research related to human use complies with all the relevant national regulations, institutional policies and was performed in accordance with the tenets of the Helsinki Declaration, and has been approved by the authors’ institutional review board or equivalent committee.
References
[1] Frerichs I. Electrical impedance tomography (EIT) in applications related to lung and ventilation: a review of experimental and clinical activities. Physiol Meas. 2000;21:R1–21.10.1088/0967-3334/21/2/201Search in Google Scholar PubMed
[2] Vogt B, Pulletz S, Elke G, Zhao Z, Zabel P, Weiler N, et al. Spatial and temporal heterogeneity of regional lung ventilation determined by electrical impedance tomography during pulmonary function testing. J Appl Physiol. 2012;113:1154–61.10.1152/japplphysiol.01630.2011Search in Google Scholar PubMed
[3] Zhao Z, Fischer R, Frerichs I, Muller-Lisse U, Moller K. Regional ventilation in cystic fibrosis measured by electrical impedance tomography. J Cyst Fibros. 2012;11:412–8.10.1016/j.jcf.2012.03.011Search in Google Scholar PubMed
[4] Karsten J, Luepschen H, Grossherr M, Bruch HP, Leonhardt S, Gehring H, et al. Effect of PEEP on regional ventilation during laparoscopic surgery monitored by electrical impedance tomography. Acta Anaesthesiol Scand. 2011;55:878–86.10.1111/j.1399-6576.2011.02467.xSearch in Google Scholar PubMed
[5] Zhao Z, Möller K, Steinmann D, Frerichs I, Guttmann J. Evaluation of an electrical impedance tomography-based global inhomogeneity index for pulmonary ventilation distribution. Intensive Care Med. 2009;35:1900–6.10.1007/s00134-009-1589-ySearch in Google Scholar PubMed
[6] Adler A, Lionheart WR. Uses and abuses of EIDORS: an extensible software base for EIT. Physiol Meas. 2006;27:S25–42.10.1088/0967-3334/27/5/S03Search in Google Scholar PubMed
[7] Schöberl J. NETGEN An advancing front 2D/3D-mesh generator based on abstract rules. Comput Vis Sci. 1997;1:41–52.10.1007/s007910050004Search in Google Scholar
[8] Javaherian A, Movafeghi A, Faghihi R, Yahaghi E. An exhaustive criterion for estimating quality of images in electrical impedance tomography with application to clinical imaging. J Vis Commun Image Represent. 2013;24:773–85.10.1016/j.jvcir.2013.05.003Search in Google Scholar
[9] Adler A, Borsic A, Polydorides N, Lionheart WR. Simple FEMs aren’t as good as we thought: experiences developing EIDORS v3. 3. Hannover, NH, USA: Proc. Conf. EIT; 2008.Search in Google Scholar
[10] Adler A, Guardo R, Berthiaume Y. Impedance imaging of lung ventilation: do we need to account for chest expansion? IEEE Trans Biomed Eng. 1996;43:414–20.10.1109/IEMBS.1994.411917Search in Google Scholar