Comparative investigation of beam quality index Q on flattening filter and flattening filter free 6 and 15 MV photon beams of an medical linear accelerator

  • Michael Schaks EMAIL logo , Johannes Zähle , Martin Janich , Patrick Hübsch , Nico Homonnay and Reinhard Gerlach

Abstract

An Elekta Synergy LINAC was used to investigate the beam quality index Q and the beam quality correction factor kQ, which is used for absolute dosimetry, in off-axis photon fields. It was found that the beam quality index Q of the photon energies with flattening filter decreases with increasing distance to the central axis, at 6 MV by 4.27% and at 15 MV by 3.98% inside a range of 15 cm off-axis. The beam quality index for flattening filter free photon fields also decreases with increasing distance to the central axis. In this case Q changed only by 1.01% inside the above range.