Skip to content
BY-NC-ND 4.0 license Open Access Published by De Gruyter September 7, 2017

Development of a novel pediatric forearm fracture treatment: simulation, prototype and evaluation

  • Hala El-Shaffey EMAIL logo , Stephan Klein and Franziska Hainer

Abstract

A treatment gap exists for moderately displaced pediatric distal forearm fractures (DFF), which carry a potential risk of improper reduction when treated conservatively. Currently, no standard method for the correction of the initial reduction of pediatric DFF is available. A novel closed reduction method has been proposed to correct and maintain a successful reduction throughout the healing process. This study further develops this novel cast system and evaluates its mechanical behavior through the design, construction, and testing of a biomechanical CAD model and pediatric forearm phantom. Initial results support the novel system as an improved treatment for pediatric fractures. Both the CAD model and tissue phantom indicate the ability of the system to manipulate a 15° fracture, with 0-1° of residual angulation.

Published Online: 2017-09-07

©2017 Hala El-Shaffey et al., published by De Gruyter

This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.

Downloaded on 1.12.2023 from https://www.degruyter.com/document/doi/10.1515/cdbme-2017-0182/html
Scroll to top button