Skip to content
BY-NC-ND 4.0 license Open Access Published by De Gruyter September 22, 2018

Biomechanics of a novel extra-articular implant for younger patients with knee osteoarthritis

  • Mehdi Saeidi , Maziar Ramezani , Piaras Kelly , Mohd Sabri Hussin and Thomas Neitzert

Abstract

This research aimed to study the efficacy of a novel implant for osteoarthritic knees. This implant is designed to eliminate excessive loads through the knee and to provide suitable conditions for possible tibiofemoral cartilage regeneration. The implant was designed for the medial side of the knee joint. Finite Element Analysis (FEA) was performed for an extended knee position of the knee joint. Contact pressure distributions on the medial and lateral compartments were investigated as well as stress distributions throughout the implant’s plates. Results with and without the implant were compared, and it was seen that the contact pressures on the surface of the distal femur were reduced by more than 90% after the introduction of the implant.

Published Online: 2018-09-22
Published in Print: 2018-09-01

© 2018 the author(s), published by Walter de Gruyter Berlin/Boston

This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.

Downloaded on 29.11.2023 from https://www.degruyter.com/document/doi/10.1515/cdbme-2018-0050/html