Systematic analysis about residual chloroform removal from PCL films

  • Michael Teske EMAIL logo , Daniela Arbeiter , Konstanze Schober , Thomas Eickner and Niels Grabow

Abstract

For fabrication of polymeric implants and drug carriers volatile organic solvents, such as chloroform, are widely used. In order to remove solvents for patient safety several processes have to be carried out to avoid changes in the thermal characteristics of the polymers, as well as in the drug release profile and mechanical properties. We analyzed films of the polymer poly(ε−caprolactone) (PCL), a common commercial medical grade polymer which is widely used in implants and drug carriers. Considering the influence on the morphology and thermal properties we screened different post treatment processes. Acceptable chloroform contents according to recommendations of the International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use (ICH) without changes in morphology and crystallinity were achieved via a simple annealing process for 24 h at 40°C and 40 mbar.

Published Online: 2018-09-22
Published in Print: 2018-09-01

© 2018 the author(s), published by Walter de Gruyter Berlin/Boston

This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.