Skip to content
BY 4.0 license Open Access Published by De Gruyter October 9, 2021

Self-Assembling Flexible 3D-MEAs for Cortical Implants

  • Lena Hegel EMAIL logo , Andrea Kauth , Karsten Seidl and Sven Ingebrandt

Abstract

Flexible Multi Electrode Arrays (MEAs) for neural interfacing reduce the mechanical mismatch between the soft brain tissue and the electrode arrays allowing accurate signal recordings and neural stimulation while reducing inflammatory responses. Many standard manufacturing processes of MEAs are designed for planar structures and the production of three-dimensional structures is challenging. In the present study, shaft structures with one to two circular gold microelectrodes (10 - 20 μm), each on a base polyimide (PI) substrate, were investigated. We describe a fabrication method, with which shafts made from bi-layer PI flip into the third dimension, which is a first step towards spontaneous assembly of electrodes in flexible 3D MEAs for neuroelectronic applications. A lift-up of the shafts was achieved by the contraction of a second PI layer and a steady nitrogen flow during polycondensation. This shrinking PI was structured in pits with a width of 5 - 600 μm. We achieved liftup angles of up to 42 degrees. The shaft structures can be hardened and later be used for neural implantation experiments.

Published Online: 2021-10-09
Published in Print: 2021-10-01

© 2021 The Author(s), published by Walter de Gruyter GmbH, Berlin/Boston

This work is licensed under the Creative Commons Attribution 4.0 International License.

Downloaded on 3.6.2023 from https://www.degruyter.com/document/doi/10.1515/cdbme-2021-2091/html
Scroll to top button