Simultaneous Quantification of the New Psychoactive Substances 3-FMC, 3-FPM, 4-CEC, and 4-BMC in Human Blood using GC-MS

  • Abdulsallam Bakdash EMAIL logo
From the journal Open Chemistry

Abstract

A gas chromatography-mass spectrometry (GC-MS) method for simultaneous quantification of 3-fluoromethcathinone (3-FMC), (±)-3-fluorophenmetrazine (3-FPM), 4-chloroethcathinone (4-CEC) and 4-Bromomethcathinone (4-BMC) in human blood with (±)-methcathinone-D3 as internal standard has been developed and validated. Whole blood samples were treated with 10% trichloroacetic acid for protein precipitation before solid phase extraction. The method was selective, the calibration curves showed linearity for all substances with R2 ranging from 0.991 to 0.998 in the range 5-1.000 ng/mL. Analysis of blank samples showed no-sign of carryover. Precision and accuracy were acceptable with values less than 20% (RSD) and ± 20% (Bias). The limit of quantification (LOQ) for all substances was 5ng/mL. Intra-day and inter-day precision were 2.111.7% and 1.3 -10.2% respectively and accuracy biases were between -10.6-19.6% % (intra-day) and 11-12.1% (inter-day). The extraction efficiencies were 85.4, 82.8, 79.1 and 74.9% for 3-FMC, 3-FPM, 4-CEC and 4-BMC respectively.

A robust and reliable simultaneous quantification method using gas chromatography-mass spectrometry in selected ion monitoring mode (GC-MS-SIM) is reported.

1 Introduction

New psychoactive substances (NPS) are becoming increasingly widespread. They are frequently sold online as research chemicals or bath salts. In order to detect them comprehensive bioanalytical methods are needed [1]. Important examples of these substances are synthetic derivatives of cathinone [2].

Khat shrub (Catha edulis) leaves contain the psychostimulant cathinone which is a natural amphetamine. Various synthetic derivatives of cathinone are being used as ‘legal highs’. As a consequence, some countries have introduced legislation restricting them [3,4]. Drugs of abuse have been detected in both conventional biological matrices (urine and blood) and non-conventional matrices (e.g. saliva, hair, nails and sweat) [4, 5, 6]. Compared to blood, urine carries higher concentrations of compounds and has a longer detection period [1, 7, 8]. Urine is therefore a primary source used in forensic screening analysis.

The symptoms of acute drug toxicity correlate, however, with the concentration of substances in the blood and not in urine. Therefore blood remains the most important specimen in the interpretation of the results of drugs of abuse detection [9,10].

Synthetic cathinones can be much more potent and hazardous than natural cathinones, or even lethal in some cases [11]. The metabolism of 3-Bromomethcathinone (3-BMC) and 3-fluoromethcathinone (3-FMC) in rat urine and human liver microsomes was studied by Meyer et al. [12] using gas chromatography–mass spectrometry (GC– MS) and liquid chromatography–mass spectrometry (LC– MS). Based on the results, their metabolism is understood to involve N-demethylation, reduction of the keto group to the corresponding alcohol, hydroxylation of the aromatic system and combinations of these steps. Mardal et al. [13] studied the metabolites of 3-fluorophenmetrazine (3-FPM) using GC–MS and LC–MS. They found that it remains mainly unchanged and is excreted as N-oxide in human urine with microbial biotransformation, and as aryl hydroxylated metabolites in rat urine.

Immunoassay techniques (IA) are rapid and simple for handling samples and are therefore used extensively

for screening whole blood and serum in cases of drugs of abuse. However, these techniques have serious limitations relating to sensitivity and selectivity and can produce false positive as well as negative results [8, 9, 14, 15]. IA can also be used as a screening technique for the detection of synthetic cathinones in biological samples [16, 17].

Chromatographic techniques have been used to increase reliability and accuracy [1, 9, 14], and hyphenated chromatographic techniques have further improved reliability, particulary GC–MS [18, 19] and LC–MS [20, 21]. GC-MS enables the detection of a larger range of substances in a single analysis and is more cost efficient than LC-MS. It is therefore the primary analysis method used in forensics. However, there are limitations of GC-MS for blood analysis: they are not sensitive enough to identify the full range of compounds that are significant on the total ion chromatogram (TIC) [22], even in full scanning mode and with the use of specific search algorithms and mass spectral libraries. Selected–ion monitoring (SIM) mode has been used to improve sensitivity, but this, in turn, affects the detection spectrum.

The analytical techniques of first choice for the detection of synthetic cathinones are GC and LC coupled with different spectroscopic instruments. GC-MS is applicable to cathinone compounds producing a very simple mass spectrum [23, 24, 25, 26]. LC–MS is employed in the analysis of cathinones more frequently than GC–MS, mostly in the electrospray ionization mass spectrometric (ESI-MSn) mode, allowing the characterization of the respective protonated molecular ions using the product ion formation patterns [25, 27, 28, 29, 30].

Most of the methods reported for whole blood analysis focused on only a single analyte and their metabolites or on small analyte groups [31, 32, 33]. As the range of different analytes is frequently unknown before testing and samples are often of a limited size, simultaneous quantification of analytes of interest is preferable in forensic analyses as it is faster and more effective. Due to this multi-analyte methods are now more widely used than in previous years [34, 35, 36].

In this work, a simultaneous quantification method for the determination of 3-FMC, 3-FPM, 4-CEC, and 4-BMC in human blood was developed using GC-MS-SIM. The internal standard used was (±)-methcathinone-D3. The analytical procedure reported was validated in accordance with the guidelines laid down by the Scientific Working Group on Forensic Toxicology (SWGTOX) on validation of bioanalytical methods [37].

2 Materials and Methods

2.1 Chemicals and Reagents

The drug standards were obtained as methanolic solutions, (1.0 mg/mL) of (±)-3-fluorophenmetrazine hydrochloride (3-FPM), 3-fluoromethcathinone hydrochloride (3-FMC), and 4-chloroethcathinone hydrochloride (4-CEC) purchased from Cerilliant (USA). A 1.0 mg standard 4-bromomethcathinone hydrochloride (4-BMC) (also known as Brephedrone) was obtained from Cayman Chemical (USA). (±)-Methcathinone-D3 hydrochloride solution 100 μg/mL in acetonitrile and Trifluoroacetic anhydride (TFAA), Hydrogen chloride ReagentPlus® ≥99% and HPLC water were obtained from Sigma Aldrich (USA).

The clean screen extraction column (UCT, size 200mg/3ml) was purchased from United Chemical Technologies (USA) consisting of octyl (C8) with benzyl sulfonic acid. Acetic acid 99.7%, ammonium hydroxide 28 - 30%, ethyl acetate, HPLC grade methanol, and HPLC grade acetone were acquired from Fisher Scientific (UK). GC grade dichloromethane and GC grade Pyridine were acquired from Riedel-DeHaën (Germany). Trichloroacetic Acid (TCA) was obtained from Biochemicals (USA). UHPLC grade Acetonitrile was acquired from Applichem Panreac (Germany) and Isopropyl alcohol (2-propanol) was acquired from Chem-Lab (Belgium).

2.2 Samples

Blank human blood samples were obtained from ten different volunteers and collected in 5 mL Vacutainer tubes containing sodium heparin. The samples tested negative for synthetic cathinones (general GC-MS full scan test). 1 mL aliquots from each blank pool were analyzed in full scan mode to confirm the absence of potential interference prior to use, as a calibration matrix and quality control (QC) procedure. Before analysis, specimens were stored at −20°C.

2.3 Preparation of Calibrators and QC Samples

3-FMC, 3-FPM, 4-BMC and 4-CEC (1.0 mg base/mL) stock solutions were prepared in methanol. A mixed calibration stock solution of 50,000 ng/mL was made by combining and diluting the methanol standards of the four analytes (1,0 mg/mL). Concentrations of 50, 250, 500, 1,000, 5,000 and 10,000 ng/mL were used for working solutions. To yield final concentrations of 5 – 1,000 ng/g, 100 μL aliquots of working solutions were added to 900 μL of blank blood. Final calibration concentrations for the method were as follows: 5, 25, 50, 100, 500, 1,000 ng/mL for each analyte.

Various ampules of 1.0 mg / mL methanol standards were used to prepare the QC working solutions used to make the calibration solutions; where available a different number of people were used to make these solutions. 15, 150, and 750 ng/mL working QC solutions were used to fortify blank whole blood samples. This yielded low, medium and high QC samples for each analyte.

100,000 ng/mL internal standard stock solution was prepared by diluting 1.0 mg/mL methanolic solution of the internal standard (±)-methcathinone-D3 in methanol. 25 μL of a 4,000 ng/mL working solution of the internal standard prepared in methanol was added to each calibration and quality control sample. This yielded 100 ng/mL internal standard concentrations. All stock solutions were stored at − 80°C before use and work solutions were prepared daily.

2.4 Instrumentation

GC analyses were performed by an Agilent 7890 GC coupled to an Agilent 5975A GC-MS Single Quad mass spectrometer (Agilent, Santa Clara, CA, USA), an ALS 7693 and a MassHunter GC/MS Acquisition (version B.07.04.2260, 28 October 2015) using an 5% phenyl methyl siloxane column (HP-5 MS UI; 30 m × 0.25 mm i.d.; 250 nm film thickness).

The GC conditions were as follows: splitless injection mode; injection port was set at 280°C; carrier gas, helium; flow rate, 1 ml/min; The GC oven temperature was initially increased from 80°C (hold 2 minutes) to 280°C at a rate of 15°C/min, giving a total run time of 15.333 minutes; The temperatures of interface, ion source and the quadrupole were 250, 230 and 110°C respectively. Ion monitoring (SIM) mode (70 eV, electron impact mode) was used for the detection.

The retention times and m/z values selected for the detection in the selected ion mode (SIM, three time windows) are shown in Table 1. The mass spectrum of targeted cathinone derivatives is shown in Figure 1.

Figure 1 EI Mass spectra, proposed structures and predominant fragmentation patterns of TFA derivatives of the target cathinones. 3-FPM indicates of (±)-3-Fluorophenmetrazine; 4-CEC, 4-Chloroethcathinone; 4-BMC, 4-Bromomethcathinone; 3-FMC, 3-Fluoromethcathinone using GC-MS, full sacn mode.
Figure 1

EI Mass spectra, proposed structures and predominant fragmentation patterns of TFA derivatives of the target cathinones. 3-FPM indicates of (±)-3-Fluorophenmetrazine; 4-CEC, 4-Chloroethcathinone; 4-BMC, 4-Bromomethcathinone; 3-FMC, 3-Fluoromethcathinone using GC-MS, full sacn mode.

Table 1

Retention times and m/z values chosen for the GC-MS-SIM analysis of 3-FMC, 3-FPM, 4-CEC, 4-BMC and the IS Methcathinone-D3 (three time windows).

CompoundTime window (min)Retention time (min)Values of m/z in SIM mode
Quantitation IonQualification Ions
3-FMC8.00 – 10.008.97154110, 123
Methcathinone-D38.00 – 10.009.14105113, 157
3-FPM10.00 – 10.7010.3216770, 98, 290
4-CEC10.70 – 12.5011.11168111, 140
4-BMC10.70 – 12.5011.27154110, 183, 185

2.5 Samples Preparation

25 mL of the 4,000 ng / mL internal standard working solution and a defined volume of the corresponding working calibration or QC fortifying solution (if applicable) were added to 1.0 mL of whole blood. 1 mL fortified blood samples of calibration and QC samples were treated with 4 mL of 10% trichloroacetic acid (TCAA) for protein precipitation before being introduced to Solid phase extraction (SPE) column cartridge (UCT, size 200mg/3ml). Samples were mixed, vortexed for 5 minutes and centrifuged at 6000 rpm for 10 minutes at 5°C. The supernatant was transferred to clean test tubes for SPE. Sequential addition of 2.5 mL water and 2.5 mL methanol were used to condition the SPE cartridges. Samples were allowed to flow under gravity after being loaded on the columns. They were then washed and rinsed by sequential addition of 2.5 mL 0.1M HCl, 2.5 mL water, 2.5 mL methanol and 2.5 mL water. They were then dried for 10 minutes under full vacuum. Using 4 ml of freshly prepared dichloromethane/isopropyl alcohol/ammonium hydroxide [80:20:2, v/v/v], the analytes were eluted under gravity and collected in glass tubes. A gentle stream of nitrogen at ≤ 40°C was used to evaporate the extracts to dryness. The residues were derivatized in capped tubes in 50 μL of ethyl acetate and 50 μL of TFAA by incubation for 30 minutes at 70°C in a heater block. Samples were then cooled to room temperature, evaporated to dryness under nitrogen at 50°C and reconstituted with 50 μL of ethyl acetate. For GC-MS / SIM analysis, 2 μL were injected.

Ethical approval: The conducted research is not related to either human or animal use.

3 Results

3.1 Method validation

The international guidelines on bioanalytical method validation [SWGFT] were the basis for performing the described analytical procedure. Stability (autosampler, bench-top and short-term freeze/thaw), precision (intraday and intermediate), selectivity, limits of detection (LOD) and quantitation (LOQ), carryover, linear range, bias, extraction and efficiency were studied.

3.2 Selectivity

10 pools from different sources of blank samples of blood were analyzed to study selectivity. Ions were monitored for each significant analyte and respective internal standard( IS). The pools were monitored and investigated for interferences at the relevant retention times. From each pool, two sets of samples (n = 10) were spiked with the same concentration (150 ng/mL) of all analytes of interest (Methcathinone-D33-FMC, 3-FPM, 4-BMC and 4-CEC). Ten negative and ten positive samples were submitted to the above methodologies. All analytes were identified successfully in all the spiked matrices. No interference of the matrices was observed at the retention times and at the m / z values of the monitored ions. This was shown by the analysis of the negative blank pools. Based on these results, the procedure described was considered selective for identifying the selected substances in human blood matrix. Total ion chromatograms for the target cathinone TFA derivatives are shown in Figure 2.

Figure 2 Total Ion Chromatograms of the given ions at 70, 98, 105, 110, 111, 113, 123, 140, 154, 157, 167, 168, 183, 185 and 290 indicating TFA derivatives of the target cathinones. 3-FPM indicates of (±)-3-Fluorophenmetrazine; 4-CEC, 4-Chloroethcathinone; 4-BMC, 4-Bromomethca thinone; 3-FMC, 3-Fluoromethcathinone, obtained after protein precipitation and SPE extraction (SPE column UCT, size 200mg/3ml) of Whole Blood using GC-MS (GC column HP-5ms, SIM mode).
Figure 2

Total Ion Chromatograms of the given ions at 70, 98, 105, 110, 111, 113, 123, 140, 154, 157, 167, 168, 183, 185 and 290 indicating TFA derivatives of the target cathinones. 3-FPM indicates of (±)-3-Fluorophenmetrazine; 4-CEC, 4-Chloroethcathinone; 4-BMC, 4-Bromomethca thinone; 3-FMC, 3-Fluoromethcathinone, obtained after protein precipitation and SPE extraction (SPE column UCT, size 200mg/3ml) of Whole Blood using GC-MS (GC column HP-5ms, SIM mode).

3.3 Linearity

The peak area ratio of each analyte and the IS were used to construct the calibration curves (n=6) by linear regression. In the whole blood, the linear range was 5 to 1,000 ng/mL (5, 25, 50, 100, 500, 1,000 ng/mL) for each substance. A blank sample with IS was added, and three separate concentration levels (15, 150, and 750 ng/mL) of QC samples (n=3) were then analysed. After the highest point of the calibration curve, a blank sample was analyzed and showed no peaks for the target analytes that the method is free of carryover. The LOQ was measured in SIM mode using blank blood fortified with all analytes at concentrations of 1, 2.5, 5, and 10 ng/mL. The S/N ratio was calculated from triplicate measurements. The lowest concentration at which the S/N ratio was equal or greater than three was considered as the LOD.

To evaluate the LOQ, the minimum concentration of the analytes measured with adequate precision (CV ≤ 20%) and bias (± 20%) was used. The acceptance criteria for calibration data have been accomplished (Table 2). Selected ion monitoring chromatograms for the analysis of whole blood blank and fortified LOQ samples are shown for all analytes in Figure 3.

Figure 3 Total ion chromatograms of TFA derivatives of cathinones at (A) limit of quantification (5 ng/mL) and (B) blank whole blood. 3-FPM indicates of (±)-3-Fluorophenmetrazine ; 4-CEC, 4-Chloroethcathinone; 4-BMC, 4-Bromomethcathinone; 3-FMC, 3-Fluoromethcathinone after protein precipitation and SPE extraction (SPE column UCT, size 200mg/3ml) of the Whole Blood GC-MS (GC column HP-5ms, SIM mode).
Figure 3

Total ion chromatograms of TFA derivatives of cathinones at (A) limit of quantification (5 ng/mL) and (B) blank whole blood. 3-FPM indicates of (±)-3-Fluorophenmetrazine ; 4-CEC, 4-Chloroethcathinone; 4-BMC, 4-Bromomethcathinone; 3-FMC, 3-Fluoromethcathinone after protein precipitation and SPE extraction (SPE column UCT, size 200mg/3ml) of the Whole Blood GC-MS (GC column HP-5ms, SIM mode).

Table 2

Linearity results for 3-FMC, 3-FPM, 4-CEC, 4-BMC in whole blood.

AnalytesLinear Rang (ng/mL)LinearityLOD (ng/mL)LOQ (ng/mL)
SlopeInterceptr2
3-FMC5-1,0000.00224±0.00002-0.00513±0.001140.9972.55
3-FPM5-1,0000.00135±0.00008-0.00207±0.003860.9982.55
4-CEC5-1,0000.00288±0.00007-0.03654±0.003400.9912.55
4-BMC5-1,0000.00174±0.00001-0.01013±0.005640.9982.55

3.4 Precision and accuracy

Coefficient of variation (CV, %) was used to express precision. The intra-day precision and bias were determined by analyzing batches of five QC samples on the same day at three different concentration levels (15, 150 and 750 ng/mL) in whole blood. The intermediate precision and bias were evaluated by the analysis of five QC samples (15, 150 and 750 ng/mL) in whole blood over three consecutive days (n=27). The bias was calculated as [ (mean of measured concentrations – theoretical concentration)/theoretical concentration x 100] for each group of concentrations of QC samples. Acceptable values for CV ± 20% and a bias variability of ± 20% have been established. The obtained values were acceptable at all concentrations for intra-day precision (%RSD; 2.1-11.7%) and bias (-10.6-19.6%), as well as to the intermediate precision (%RSD; 3.7-9.6%) and bias (-11-12.1%) in the studied matrices (Tables 3 and 4).

Table 3

Intra-day precision and accuracy (n=5).

SubstanceSpiked Concentrations (ng/mL)
15150750
Concentration FoundRSDBiasConcentration FoundRSDBias (%)Concentration FoundRSDBias
(ng\mL) Mean ±SD(%)(%)(ng\mL) Mean ±SD(%)(ng\mL) Mean ±SD(%)(%)
3-FMC17.9±0.52.719.6161.5±9.35.7-5.6753±53.77.1-7.3
3-FPM16.3±1.37.9-8.7142.4±7.75.45.0787.8±57.47.6-5.0
4-CEC16.8±1.911.7-12143.4±7.75.34.4729.6±15.22.12.7
4-BMC16.6±1.59.0-10.6163.6±3.12.2-9.1769.8±30.63.9-2.6
Table 4

Inter-day precision and accuracy (n=15).

SubstanceSpiked Concentrations (ng/mL)
15150750
Concentration FoundRSDBiasConcentration FoundRSDBiasConcentration FoundRSDBias
(ng\mL) Mean ±SD(%)(%)(ng\mL) Mean ±SD(%)(%)(ng\mL) Mean ±SD(%)(%)
3-FMC16.2±1.38.06.1146.9±7.35.03.7768.1±42.65.56.9
3-FPM15.7±1.610.2-3.2153.3±5.63.7-4.2738.3±26.43.63.7
4-CEC16.8±1.16.5-11160.5±2.11.36.5763.6±22.93.0-10.8
4-BMC15.2±0.95.99.6156.8±6.24.07.1759.7±32.44.312.1

3.5 Recovery

The efficiency of extraction was measured by comparing two sets of six different QC samples at each concentration level (15, 150 and 750 ng / mL) in which the IS was added after extraction. Blank samples were fortified in set 1 with all analytes, while analytes were added to the QC samples after the extraction procedure (100 percent recovery) in set 2. The average peak range ratio was compared as follows: Set1/Set2 x100%. The extraction efficiencies for 3-FMC, 3-FPM, 4-CEC and 4-BMC were 85.4, 82.8, 79.1 and 74.9 respectively, as shown in Table 5.

Table 5

Extraction efficiency for 3-FMC, 3-FPM, 4-CEC and 4-BMC.

SubstanceExtraction efficiency % (mean values ± standard deviation) (n=6)
15 (ng/mL)150 (ng/mL)750 (ng/mL)
3-FMC82.2±5.785.3±3.888.6±2.1
3-FPM86.6±3.582.4±4.579.3±0.9
4-CEC79.1±6.381.6±0.576.5±3.2
4-BMC76.7±4.279.3±1.574.9±6.7

4 Discussion

The study aimed at developing a GS-MS-SIM procedure for simultaneous analysis as well as for precise and accurate quantification of four synthetic derivatives of cathinone in whole human blood. Mass spectral profiles of cathinones are very simple in positive mode GC–MS. In the mass spectra obtained in the EI-MS mode each main fragment ion detected for 4-CEC, 3-FMC, 4-BMC and 3-FPM was at m/z 168, 154, 154 and 167, respectively. Other less intense fragments present in the spectra were at m/z 140 and 111 for 4-CEC, m/z 110 and 123 for 4-FMC, m/z 182 and 110 for 4-BMC, and m/z 98 and 123 for 3-FPM. Possible structures of the fragmentation products derived from the parent structures of the four compounds are presented in Figure 1. The obtained spectra showed the presence of the oxonium and iminium ions and were in accordance with reported fragmentation pathways of other cathinone derivatives [38].

Analysis of cathinones in biological materials (human blood, urine and cellular fraction) which has been carried out using SPE, liquid-liquid extraction LLE and solid-phase microextraction headspace (HS-SPME) followed by GC-MS methods has been widely reported [39, 40, 41, 42, 43, 44, 45, 46, 47, 48, 49, 50]. LOD values using liquid-liquid extraction (LLE) were between 2-50 ng/mL [40, 41, 43]. In another study LODs of eighteen synthetic cathinones (flephedrone, methcathinone, ethcathinone, buphedrone, mephedrone, pentedrone, 4-MEC, 4-EMC, methedrone, methylone, ethylone, α-PVP, butylone, MPBP, pentylone, pyrovalerone, MDPBP, MDPV and naphyrone) ranged between 25-100 ng after mixed mode SPE extraction of the blood samples and GC-MS detection ion full scan [51]. Also the analytical recoveries for blood ranged between 47-73%. LOQ values were 2.5 ng/ml for some cathinones [4]. Linear range values were between 2-2000 ng/mL for LLE [40, 42, 43], 10-2000 ng/mL for SPE [46], and 1-200 ng/mL for SPME [44].

By comparing the results of the previous studies using GC-MS SIM mode with the current study the method developed has acceptable selectivity (it is free of interferences), linearity (5-1,000) (plus free from carryover), LOD (2.5 ng/mL), precision and accuracy, and a very good recovery (75-88%). The method gave very low LOQ values (5 ng/ml) for the four analytes. These values are in good agreement with the previous studies. This method meets or exceeds the cut-off requirements set out in SWGTOX ‘s Standard Practices for Validation of Methods in Forensic Toxicology [37].

The various cathinone administration modes used by abusers cause the concentration of cathinones in blood samples to vary. In fatal cases, the concentrations of the cathinones in blood were 400 ng/ml [46], 560–3,300 [43], 272 [52], and 60–1,120 ng/ml [53] for methylone; 1.2–22 [54], 5.1 [55], and 5.5 lg/ ml [56] for mephedrone; 55.2 ng/ml for a-PBP [57]; 486 [44] and 654 ng/ml [58] for a-PVP; 180 ng/ml [59] for PV9; and 170 [41], 82 [60], 1,200 [74], 440 [46], 17–38 [61], and 700 ng/ml [62] for MDPV. In nonfatal cases, concentrations ranged from 250-910 ng/mL for bk-MBDB [63]. The linearity range (5-1,000 ng/mL) as well as LOD and LOQ values in the recent study are in good agreement with the previous studies and can cover the possible blood concentrations in both fatal and non-fatal cases associated with studied cathinones. The results above provide a definite indication that the use of mixed mode SPE and GC-MS SIM mode could lead to reliable results for genuine samples with the studied analytes.

5 Conclusion

A robust and reliable simultaneous quantification method of four synthetic derivatives of cathinone in whole human blood using GC-MS-SIM technique is reported.

  1. Conflict of interest: The author reports no conflicts of interest relevant to this article.

References

[1] Ambach L., Hernández Redondo A., König S., Angerer V., Schürch S., Weinmann W., Detection and quantification of 56 new psychoactive substances in whole blood and urine by LC– MS/MS, Bioanalysis., 2015, 7(9), 1119-36.10.4155/bio.15.48Search in Google Scholar PubMed

[2] Kuś P., Kusz J., Książek M., Pieprzyca E., Rojkiewicz M., Spectroscopic characterization and crystal structures of two cathinone derivatives: N-ethyl-2-amino-1-phenylpropan-1-one (ethcathinone) hydrochloride and N-ethyl-2-amino-1-(4-chlorophenyl) propan-1-one (4-CEC) hydrochloride, Forensic Toxicol., 2017, 35(1), 114-24.10.1007/s11419-016-0345-6Search in Google Scholar PubMed PubMed Central

[3] Kelly J.P., Cathinone derivatives: a review of their chemistry, pharmacology and toxicology, Drug Test. Anal., 2011, 3(7‐8), 439-53.10.1002/dta.313Search in Google Scholar PubMed

[4] Mohamed K.M., Bakdash A., Comparison of 3 Derivatization Methods for the Analysis of Amphetamine-Related Drugs in Oral Fluid by Gas Chromatography-Mass Spectrometry, Anal. chem. insights., 2017, 12, 1177390117727533.10.1177/1177390117727533Search in Google Scholar PubMed PubMed Central

[5] de la Torre R., Farré M., Navarro M., Pacifici R., Zuccaro P., Pichini S., Clinical pharmacokinetics of amfetamine and related substances, Clin. pharmacokinet., 2004, 43(3), 157-85.10.2165/00003088-200443030-00002Search in Google Scholar PubMed

[6] Villamor J.L., Bermejo A.M., Fernandez P., Tabernero M.J., A new GC-MS method for the determination of five amphetamines in human hair, J. Anal. Toxicol., 2005,, 29(2), 135-9.10.1093/jat/29.2.135Search in Google Scholar PubMed

[7] Villamor J.L., Bermejo A.M., Fernandez P., Tabernero M.J., GC‐ MS determination of amphetamines in human urine, Ana. lett., 2005, 38(5), 781-90.10.1081/AL-200051022Search in Google Scholar

[8] Lee Y., Lai K.K., Sadrzadeh S.H., Simultaneous detection of 19 drugs of abuse on dried urine spot by liquid chromatography– tandem mass spectrometry, Clin biochem., 2013, 46(12), 1118-24.10.1016/j.clinbiochem.2013.03.027Search in Google Scholar PubMed

[9] Gunnar T., Mykkänen S., Ariniemi K., Lillsunde P., Validated semiquantitative/quantitative screening of 51 drugs in whole blood as silylated derivatives by gas chromatography-selected ion monitoring mass spectrometry and gas chromatography electron capture detection, J. Chromatogr. B. Biomed. Sci. Appl., 2004, 806(2), 205-19.10.1016/j.jchromb.2004.04.005Search in Google Scholar PubMed

[10] Steuer A.E., Forss A.M., Dally A.M., Kraemer T., Method development and validation for simultaneous quantification of 15 drugs of abuse and prescription drugs and 7 of their metabolites in whole blood relevant in the context of driving under the influence of drugs––Usefulness of multi-analyte calibration, Forensic Sci. Int., 2014, 244, 92-101.10.1016/j.forsciint.2014.08.022Search in Google Scholar PubMed

[11] Baumann M.H., Awash in a sea of ‘bath salts’: implications for biomedical research and public health. Addiction, 2014, 109, 1577-1579.10.1111/add.12601Search in Google Scholar

[12] Meyer M.R., Vollmar C., Schwaninger A.E., Wolf E., Maurer H.H., New cathinone‐derived designer drugs 3‐bromomethcathinone and 3‐fluoromethcathinone: studies on their metabolism in rat urine and human liver microsomes using GC–MS and LC–high‐ resolution MS and their detectability in urine. Journal of Mass Spectrometry, 2012, 47(2), 253-262.10.1002/jms.2960Search in Google Scholar

[13] Mardal M., Miserez B., Bade R., Portolés T., Bischoff M., Hernández F., Meyer M.R. 3-Fluorophenmetrazine, a fluorinated analogue of phenmetrazine: Studies on in vivo metabolism in rat and human, in vitro metabolism in human CYP isoenzymes and microbial biotransformation in Pseudomonas Putida and wastewater using GC and LC coupled to (HR)-MS techniques. Journal of pharmaceutical and biomedical analysis, 2016, 128, 485-495.10.1016/j.jpba.2016.06.011Search in Google Scholar

[14] Lillsunde P., Michelson L., Forsström T., Korte T., Schultz E., Ariniemi K., Portman M, Sihvonen ML, Seppälä T. Comprehensive drug screening in blood for detecting abused drugs or drugs potentially hazardous for traffic safety, Forensic Sci. Int., 1996,77(3), 191-210.10.1016/0379-0738(95)01862-XSearch in Google Scholar

[15] Hino Y., Ojanperä I., Rasanen I., Vuori E., Performance of immunoassays in screening for opiates, cannabinoids and amphetamines in post-mortem blood. Forensic Sci. Int., 2003, 131(2-3), 148-55.10.1016/S0379-0738(02)00430-9Search in Google Scholar

[16] Ellefsen K.N., Anizan S., Castaneto M.S., Desrosiers N.A., Martin L.T.M., Klette C.K.L., Huestis M.A., Validation of the only commercially available immunoassay for synthetic cathinones in urine: Randox Drugs of Abuse V Biochip Array technology. Drug Test Anal., 2014, 6, 728–738.10.1002/dta.1633Search in Google Scholar

[17] Swortwood M.J., Hearn W.L., DeCaprio A.P., Cross-reactivity of designer drugs, including cathinone derivatives, in commercial enzyme-linked immunosorbent assays. Drug Test Anal., 2014, 6, 716–727.10.1002/dta.1489Search in Google Scholar

[18] Polettini A., Groppi A., Vignali C., Montagna M., Fully-automated systematic toxicological analysis of drugs, poisons, and metabolites in whole blood, urine, and plasma by gas chromatography–full scan mass spectrometry. J. Chromatogr. B. Biomed. Sci. Appl., 1998, 713(1), 265-79.10.1016/S0378-4347(98)00062-0Search in Google Scholar

[19] Staerk U., Külpmann W.R., High-temperature solid-phase microextraction procedure for the detection of drugs by gas chromatography–mass spectrometry, J. Chromatogr. B. Biomed. Sci. Appl., 2000, 745(2), 399-411.10.1016/S0378-4347(00)00312-1Search in Google Scholar

[20] Weinmann W., Wiedemann A., Eppinger B., Renz M., Svoboda M., Screening for drugs in serum by electrospray ionization/collision-induced dissociation and library searching, J. AM. Soc. Mass. Spectr., 1999, 10(10), 1028-37.10.1016/S1044-0305(99)00070-7Search in Google Scholar

[21] Bogusz M.J., Liquid chromatography–mass spectrometry as a routine method in forensic sciences: a proof of maturity, J. Chromatogr. B. Biomed. Sci. Appl., 2000, 748(1), 3-19.10.1016/S0378-4347(00)00461-8Search in Google Scholar

[22] Aebi B., Bernhard W., Advances in the use of mass spectral libraries for forensic toxicology, J. Anal. Toxicol., 2002, 26(3), 149-56.10.1093/jat/26.3.149Search in Google Scholar PubMed

[23] Kudo K., Usumoto Y., Usui K., Hayashida M., Kurisaki E., Saka K,, Tsuji A,, Ikeda N., Rapid and simultaneous extraction of acidic and basic drugs from human whole blood for reliable semi-quantitative NAGINATA drug screening by GC–MS. Forensic Toxicol., 2013, 32, 97–104.10.1007/s11419-013-0215-4Search in Google Scholar

[24] Namera A, Kawamura M, Nakamoto A, Saito T, Nagao M., Comprehensive review of the detection methods for synthetic cannabinoids and cathinones. Forensic Toxicol. 2015, 33(2), 175–194.10.1007/s11419-015-0270-0Search in Google Scholar PubMed PubMed Central

[25] Zuba D., Identification of cathinones and other active components of “legal highs” by mass spectrometric methods. Trends Anal Chem., 2012, 32, 15–30.10.1016/j.trac.2011.09.009Search in Google Scholar

[26] Kohyama E., Chikumoto T., Kitaichi K., Horiuchi T., Ito T. Differentiation of the isomers of N-alkylated cathinones by GC-EI-MS-MS and LC-PDA, Anal. Sci., 2016, 32, 831–837.10.2116/analsci.32.831Search in Google Scholar PubMed

[27] Rojek S., Kłys M., Strona M., Maciów M., Kula K. “Legal Highs”– toxicity in the clinical and medico-legal aspect as exemplified by suicide with bk-MDMA administration, Forensic Sci. Int., 2012, 22, e1–e6.10.1016/j.forsciint.2012.04.034Search in Google Scholar PubMed

[28] Ibáñez M., Sancho J.V., Bijlsma L., van Nuijs A.L.N., Covaci A., Hernández F. Comprehensive analytical strategies based on high-resolution time-of-flight mass spectrometry to identify new psychoactive substances, Trends Anal Chem., 2014, 57, 107–117.10.1016/j.trac.2014.02.009Search in Google Scholar

[29] Jankovics P., Váradi A., Tölgyesi L., Lohner S., Németh-Palotás J., Kőszegi-Szalai H. Identification and characterization of the new designer drug 4′-methylethcathinone (4-MEC) and elaboration of a novel liquid chromatography-tandem mass spectrometry (LC-MS/MS) screening method for seven different methcathinone analogs, Forensic Sci. Int., 2011, 210, 213–220.10.1016/j.forsciint.2011.03.019Search in Google Scholar PubMed

[30] Fornal E., Identification of substituted cathinones: 3,4-methylenedioxy derivatives by high performance liquid chromatography-quadrupole time of flight mass spectrometry, J. Pharm. Biomed. Anal., 2013, 81-82, 13-9.10.1016/j.jpba.2013.03.016Search in Google Scholar PubMed

[31] Maurer H.H., Mass spectrometric approaches in impaired driving toxicology, Anal. Bioanal. Chem., 2009, 393(1), 97.10.1007/s00216-008-2338-8Search in Google Scholar PubMed

[32] Kraemer T., Paul L.D., Bioanalytical procedures for determination of drugs of abuse in blood, Anal. Bioanal. Chem., 2007, 388(7), 1415-35.10.1007/s00216-007-1271-6Search in Google Scholar PubMed

[33] Fernández M.D., Wille S.M., Samyn N., Wood M., López-Rivadulla M., De Boeck G., High-Throughput Analysis of Amphetamines in Blood and Urine with Online Solid-Phase Extraction-Liquid Chromatography—Tandem Mass Spectrometry, J. Anal. Toxicol., 2009, 33(9), 578-87.10.1093/jat/33.9.578Search in Google Scholar PubMed

[34] Musshoff F., Trafkowski J., Kuepper U., Madea B., An automated and fully validated LC‐MS/MS procedure for the simultaneous determination of 11 opioids used in palliative care, with 5 of their metabolites, J. mass spectrom., 2006, 41(5), 633-40.10.1002/jms.1021Search in Google Scholar PubMed

[35] Bjørk M.K., Nielsen M.K., Markussen L.Ø., Klinke H.B., Linnet K., Determination of 19 drugs of abuse and metabolites in whole blood by high-performance liquid chromatography– tandem mass spectrometry, Anal. Bioanal. Chem., 2010, 396(7), 2393-401.10.1007/s00216-009-3268-9Search in Google Scholar PubMed

[36] Simonsen K.W., Hermansson S., Steentoft A., Linnet K., A validated method for simultaneous screening and quantification of twenty-three benzodiazepines and metabolites plus zopiclone and zaleplone in whole blood by liquid-liquid extraction and ultra-performance liquid chromatography-tandem mass spectrometry, J. Anal. Toxicol., 2010, 34(6), 332-41.10.1093/jat/34.6.332Search in Google Scholar PubMed

[37] Scientific Working Group for Forensic Toxicology., Scientific Working Group for Forensic Toxicology (SWGTOX) standard practices for method validation in forensic toxicology, J. Anal. Toxicol., 2013, 37, 452-74.10.1093/jat/bkt054Search in Google Scholar PubMed

[38] Matsuta S., Katagi M., Nishioka H., Kamata H., Sasaki K., Shima N., et al., Structural characterization of cathinone-type designer drugs by EI mass spectrometry (in Japanese with English abstract). Jpn. J. Forensic Sci. Tech., 2014, 19, 77–89.10.3408/jafst.19.77Search in Google Scholar

[39] Meyer M.R., Du P., Schuster F., Maurer H.H., Studies on the metabolism of the a-pyrrolidinophenone designer drug methylenedioxy-pyrovalerone (MDPV) in rat and human urine and human liver microsomes using GC–MS and LC– highresolution MS and its detectability in urine by GC–MS, J. Mass Spectrom., 2010, 45, 1426–1442.10.1002/jms.1859Search in Google Scholar PubMed

[40] Strano-Rossi S., Cadwallader A.B., de la Torre X., Botre` F., Toxicological determination and in vitro metabolism of the designer drug methylenedioxypyrovalerone (MDPV) by gas chromatography/mass spectrometry and liquid chromatography/ quadrupole time-of-flight mass spectrometry, Rapid Commun. Mass Spectrom., 2010, 24, 2706–2714.10.1002/rcm.4692Search in Google Scholar PubMed

[41] Spiller H.A., Ryan M.L., Weston R.G., Jansen J., Clinical experience with and analytical confirmation of ‘‘bath salts’’ and ‘‘legal highs’’ (synthetic cathinones) in the US, Clin. Toxicol., 2011, 49,499–505.10.3109/15563650.2011.590812Search in Google Scholar PubMed

[42] Ojanpera¨ I.A., Heikman P.K., Rasanen I.J., Urine analysis of 3,4-methylenedioxypyrovalerone in opioid-dependent patients by gas chromatography–mass spectrometry, Ther. Drug. Monit., 2011, 33, 257–263.10.1097/FTD.0b013e318208b693Search in Google Scholar PubMed

[43] Pearson J.M., Hargraves T.L., Hair L.S., Massucci C.J., Frazee C.C., Garg U., Pietak B.R., Three fatal intoxications due to methylone, J. Anal. Toxicol., 2012, 36, 444–451.10.1093/jat/bks043Search in Google Scholar PubMed

[44] Saito T., Namera A., Osawa M., Aoki H., Inokuchi S., SPME–GC– MS analysis of a-pyrrolidinovaleorophenone in blood in a fatal poisoning case, Forensic Toxicol., 2013, 31, 328–332.10.1007/s11419-013-0183-8Search in Google Scholar

[45] Namera A., Urabe S., Saito T., Torikoshi-Hatano A., Shiraishi H., Arima Y., et al., A fatal case of 3,4-methylenedioxypyrovalerone poisoning: coexistence of a-pyrrolidinobutiophenone and a-pyrrolidinovalerophenone in blood and/or hair, Forensic Toxicol., 2013, 31, 338–34310.1007/s11419-013-0192-7Search in Google Scholar

[46] Wyman J.F., Lavins E.S., Engelhart D., Armstrong E.J., Snell K.D., Boggs P.D., et al., Postmortem tissue distribution of MDPV following lethal intoxication by bath salts, J. Anal. Toxicol., 2013, 37, 182–185.10.1093/jat/bkt001Search in Google Scholar PubMed

[47] Wright T.H., Cline-Parhamovich K., Lajoie D., Parsons L., Dunn M., Ferslew K.E., Deaths involving methylenedioxypyrovalerone (MDPV) in Upper East Tennessee, J. Forensic Sci., 2013, 58,1558–1562.10.1111/1556-4029.12260Search in Google Scholar PubMed

[48] Shima N., Katagi M., Kamata H., Matsuta S., Nakanishi K., Zaitsu K., et al., Urinary excretion and metabolism of the newly encountered designer drug 3,4-dimethylmethcathinone in humans, Forensic Toxicol., 2013, 31, 101–112.10.1007/s11419-012-0172-3Search in Google Scholar

[49] Uralets V., Rana S., Morgan S., Ross W., Testing for designer stimulants: metabolic profiles of 16 synthetic cathinones excreted free in human urine. J. Anal. Toxicol., 2014, 38, 233–241.10.1093/jat/bku021Search in Google Scholar PubMed

[50] Shima N., Katagi M., Kamata H., Matsuta S., Sasaki K., Kamata T., et al., Metabolism of the newly encountered designer drug a-pyrrolidinovalerophenone in humans: identification and quantitation of urinary metabolites, Forensic Toxicol., 2014, 32, 59–67.10.1007/s11419-013-0202-9Search in Google Scholar

[51] Kerrigan S., Improved detection of synthetic cathinones in forensic toxicology samples: thermal degradation and analytical considerations, Sam Houston State University, 2015.Search in Google Scholar

[52] Kovács K., Tóth A.R., Kereszty E.M., A new designer drug: methylone related death, Orv. Hetil., 2012, 153, 271–276.10.1556/OH.2012.29310Search in Google Scholar PubMed

[53] Cawrse B.M., Levine B., Jufer R.A., Fowler D.R., Vorce S.P., Dickson A.J., et al., Distribution of methylone in four postmortem cases, J. Anal. Toxicol., 2012, 36, 434–439.10.1093/jat/bks046Search in Google Scholar PubMed

[54] Torrance H., Cooper G., The detection of mephedrone (4-methylmethcathinone) in 4 fatalities in Scotland, Forensic Sci. Int., 2010, 202, e62–e63.10.1016/j.forsciint.2010.07.014Search in Google Scholar PubMed

[55] Lusthof K.J., Oosting R., Maes A., Verschraagen M., Dijkhuizen A., Sprong A.G., A case of extreme agitation and death after the use of mephedrone in The Netherlands, Forensic Sci. Int., 2011, 206, e93–e95.10.1016/j.forsciint.2010.12.014Search in Google Scholar PubMed

[56] Adamowicz P., Tokarczyk B., Stanaszek R., Slopianka M., Fatal mephedrone intoxication—a case report, J. Anal. Toxicol., 2013, 37, 37–42.10.1093/jat/bks085Search in Google Scholar PubMed

[57] Wurita A., Hasegawa K., Minakata K., Gonmori K., Nozawa H., Yamagishi I., et al., . Postmortem distribution of α-pyrrolidinobutiophenone in body fluids and solid tissues of a human cadaver, Leg. Med., 2014, 16, 241–246.10.1016/j.legalmed.2014.05.001Search in Google Scholar PubMed

[58] Hasegawa K., Suzuki O., Wurita A., Minakata K., Yamagishi I., Nozawa H., et al., Postmortem distribution of α-pyrrolidinovalerophenone and its metabolite in body fluids and solid tissues in a fatal poisoning case measured by LC–MS–MS with the standard addition method, Forensic Toxicol., 2014, 32, 225–234.10.1007/s11419-014-0227-8Search in Google Scholar

[59] Hasegawa K., Wurita A., Minakata K., Gonmori K., Nozawa H., Yamagishi I., et al., Identification and quantitation of a new cathinone designer drug PV9 in an “aroma liquid” product, antemortem whole blood and urine specimens, and a postmortem whole blood specimen in a fatal poisoning case, Forensic Toxicol., 2014, 32, 243–250.10.1007/s11419-014-0230-0Search in Google Scholar

[60] Murray B.L., Murphy C.M., Beuhler M.C., Death following recreational use of designer drug “bath salts” containing 3,4-methylenedioxypyrovalerone (MDPV), J. Med. Toxicol., 2012, 8, 69–75.10.1007/s13181-011-0196-9Search in Google Scholar PubMed PubMed Central

[61] Adamowicz P., Gil D., Skulska A., Tokarczyk B., Analysis of MDPV in blood—determination and interpretation, J. Anal. Toxicol., 2013, 37, 308–312.10.1093/jat/bkt025Search in Google Scholar PubMed

[62] Kesha K., Boggs C.L., Ripple M.G., Allan C.H., Levine B., Jufer-Phipps R., et al., Methylenedioxypyrovalerone (“bath salts”), related death: case report and review of the literature, J. Forensic Sci., 2013, 58, 1654–1659.10.1111/1556-4029.12202Search in Google Scholar PubMed

[63] Pinorini-Godly M. T., Sporkert F., Schläpfer M., Rusconi M., Conti A., Intoxication with butylone and phenazepam among young cocaine users, Toxichem. Krimtech, 2010, 77(3), 226-227.Search in Google Scholar

Received: 2019-01-21
Accepted: 2019-05-23