Abstract
Objective
To compare the accuracy of cord blood lactate measurement using gas analyzer and portable devices in order to assess possibility of implementation of these devices in clinical practice.
Methods
We performed a prospective observational study using 30 umbilical cord samples which were obtained immediately after birth. Portable electrochemical devices Lactate Scout (SensLab GmbH, Leipzig, Germany) and StatStrip Lactate (NOVA Biomedical, Waltham, MA, USA) were used to determine lactate level. A gas analyzer ABL800 FLEX (Radiometer Medical ApS, Brønshøj-Husum, Denmark) was used as a reference. Base excess (BE), pH, partial oxygen (pO2) and carbon dioxide (pCO2) pressure, hemoglobin (ctHb) and bilirubin (ctBl) levels were measured.
Results
The mean umbilical cord blood lactate level determined by the gas analyzer was 5.85 ± 2.66 mmol/L (ranging from 1.4 mmol/L to 13.4 mmol/L). Lactate level estimated by Lactate Scout was 5.66 ± 2.65 mmol/L and did not significantly differ from the reference method level (P = 0.2547). The mean lactate level determined by StatStrip Lactate was significantly lower than by the gas analyzer – 4.81 ± 2.38 mmol/L (P < 0.0001). Umbilical cord blood pH, BE, pO2 and pCO2, ctHb and ctBl levels did not affect the accuracy of the lactate measurement in absolute units (mmol/L).
Conclusion
Umbilical cord blood lactate level measured by StatStrip Lactate was lower than estimated by the ABL800 FLEX gas analyzer. This shows the necessity to develop decision-making reference points separately for each device. Umbilical cord blood pH, BE, pO2 and pCO2, ctHb and ctBl levels did not affect the accuracy of measurements by electrochemical portable devices.
Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.
Research funding: State assignment “Modern Aspects of Diagnosis of the Fetal State in Births”.
Employment or leadership: None declared.
Honorarium: None declared.
Competing interests: The funding organization(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.
References
1. Perlman JM. Hyperthermia in the delivery: potential impact on neonatal mortality and morbidity. Clin Perinatol 2006;33:55–63.10.1016/j.clp.2005.11.002Search in Google Scholar PubMed
2. Perlman JM. Summary proceedings from the neurology group on hypoxic-ischemic encephalopathy. Pediatrics 2006;117:S28–33.10.1542/peds.2005-0620ESearch in Google Scholar PubMed
3. Orsonneau J-L, Fraissinet F, Sébille-Rivain V, Dudouet D, Bigot-Corbel E. Suitability of POC lactate methods for fetal and perinatal lactate testing: considerations for accuracy, specificity and decision making criteria. Clin Chem Lab Med 2013;51:397–404.10.1515/cclm-2012-0201Search in Google Scholar PubMed
4. Prikhod’ko A, Romanov Ay, Evgrafova A, Shuklina D. Detection of pH and blood lactate levels in the presenting part of the foetus for evaluation of its status during labour. Vopr Ginekol Akusherstva i Perinatol 2017;16:96–9.10.20953/1726-1678-2017-6-96-99Search in Google Scholar
5. East CE, Leader LR, Sheehan P, Henshall NE, Colditz PB, Lau R. Intrapartum fetal scalp lactate sampling for fetal assessment in the presence of a non-reassuring fetal heart rate trace. Cochrane database Syst Rev 2015:CD006174.10.1002/14651858.CD006174.pub3Search in Google Scholar PubMed
6. Ramanah R, Martin A, Clement M-C, Maillet R, Riethmuller D. Fetal scalp lactate microsampling for non-reassuring fetal status during labor: a prospective observational study. Fetal Diagn Ther 2010;27:14–9.10.1159/000262281Search in Google Scholar PubMed
7. Kruger K, Hallberg B, Blennow M, Kublickas M, Westgren M. Predictive value of fetal scalp blood lactate concentration and pH as markers of neurologic disability. Am J Obstet Gynecol 1999;181:1072–8.10.1016/S0002-9378(99)70083-9Search in Google Scholar PubMed
8. Nordström L. Fetal scalp blood measurements during labour-lactate or pH? Clin Biochem 2011;44:456–7.10.1016/j.clinbiochem.2011.03.007Search in Google Scholar PubMed
9. Tanner RK, Fuller KL, Ross MLR. Evaluation of three portable blood lactate analysers: Lactate Pro, Lactate Scout and Lactate Plus. Eur J Appl Physiol 2010;109:551–9.10.1007/s00421-010-1379-9Search in Google Scholar PubMed
10. Carbonne B, Nguyen A. Fetal scalp blood sampling for pH and lactate measurement during labour. J Gynecol Obstet Biol Reprod (Paris) 2008;37:S65–71.10.1016/j.jgyn.2007.11.012Search in Google Scholar PubMed
11. Heinis A, van Dillen J, Oosting J, Rhöse S, Vandenbussche F, Van Drongelen J. Clinical evaluation of Statstrip® Lactate for use in fetal scalp blood sampling. Acta Obstet Gynecol Scand 2017;96:334–41.10.1111/aogs.13078Search in Google Scholar PubMed
12. Wang M, Chua SC, Bouhadir L, Treadwell EL, Gibbs E, McGee TM. Point-of-care measurement of fetal blood lactate – Time to trust a new device. Aust N Z J Obstet Gynaecol 2018;58:72–8.10.1111/ajo.12671Search in Google Scholar PubMed
13. Iorizzo L, Klausen TW, Wiberg-Itzel E, Ovin F, Wiberg N. Use of Lactate ProTM2 for measurement of fetal scalp blood lactate during labor – proposing new cutoffs for normality, preacidemia and acidemia: a cross-sectional study. J Matern Fetal Neonatal Med 2019;32:1762–8.10.1080/14767058.2017.1416603Search in Google Scholar PubMed
14. Rørbye C, Perslev A, Nickelsen C. Lactate versus pH levels in fetal scalp blood during labor–using the Lactate Scout System. J Matern Fetal Neonatal Med 2016;29:1200–4.10.3109/14767058.2015.1045863Search in Google Scholar PubMed
15. Heinis AMF, Dinnissen J, Spaanderman MEA, Lotgering FK, Gunnewiek JMTK. Comparison of two point-of-care testing (POCT) devices for fetal lactate during labor. Clin Chem Lab Med 2011;50:89–93.10.1515/cclm.2011.734Search in Google Scholar
16. Tolan NV, Wockenfus AM, Koch CD, Crews BO, Dietzen DJ, Karon BS. Analytical performance of three whole blood point-of-care lactate devices compared to plasma lactate comparison methods and a flow-injection mass spectrometry method. Clin Biochem 2017;50:168–73.10.1016/j.clinbiochem.2016.11.004Search in Google Scholar PubMed
17. Colon-Franco JM, Lo SF, Tarima SS, Gourlay D, Drendel AL, Brook Lerner E. Validation of a hand-held point of care device for lactate in adult and pediatric patients using traditional and locally-smoothed median and maximum absolute difference curves. Clin Chim Acta 2017;468:145–9.10.1016/j.cca.2017.02.015Search in Google Scholar PubMed
18. Wiberg-Itzel E, Lipponer C, Norman M, Herbst A, Prebensen D, Hansson A, et al. Determination of pH or lactate in fetal scalp blood in management of intrapartum fetal distress: randomised controlled multicentre trial. Br Med J 2008;336:1284–7.10.1136/bmj.39553.406991.25Search in Google Scholar PubMed PubMed Central
©2020 Walter de Gruyter GmbH, Berlin/Boston