Impact of Physical Parameters on Dosing Errors due to a Syringe Exchange in Multi-Infusion Therapy

  • Roland A. Snijder

    Roland Snijder obtained his PhD at Utrecht University in 2016 and a master’s degree in Biomedical Engineering at the University of Groningen with a specialization curriculum in the area of medical physics (medical instrumentation and imaging) in 2012. As a PhD candidate, Roland conducted research at the department of Medical Technology and Clinical Physics of University Medical Center Utrecht (UMC Utrecht) in cooperation with the department of Clinical Pharmacy. Roland’s research focused on investigating physical causes of dosing errors in multi-infusion systems. Specifically, the aim was to investigate the origin and characteristics of these dosing errors using in vitro measurement methods and theoretical modeling. Moreover, the clinical relevance of these dosing errors was investigated using pharmacological theory and a retrospective clinical study. The research was conducted within the framework of the European Metrology Research Programme (EMRP).

    EMAIL logo
    , Maurits K. Konings

    Maurits Konings, PhD, is a physicist working at the Department of Medical Technology and Clinical Physics of the University Medical Center Utrecht. His research interests are non-invasive monitoring of cardiac function, new methods for non-invasive neurostimulation and neuromonitoring, and multi-infusion. His is inventor of over 8 patents in the field of medical technology.

    , Agnes van den Hoogen

    Agnes van den Hoogen is a nurse scientist and teacher at the University Medical Centre of Utrecht (UMCU) – Wilhelmina Children’s Hospital in Utrecht, The Netherlands. She obtained her PhD at the University of Utrecht, medical faculty of the University Medical Centre Utrecht. Topic of her thesis: Infections in Neonatal Intensive: Care Prevalence, Prevention and Antibiotic use. Clinical Interest: Neonatal infections, Management of Intra Venous access in Neonates, Family Centred Care, Family Empowerment, Ethics, Epidemiology and Education

    and Annemoon M.D.E. Timmerman

    Annemoon Timmerman received her PhD in 2002 at Delft University of Technology. At 2005 she attained her registration as a medical physicist at Erasmus Medical Center. Since 2005 she has been working at the University Medical Center Utrecht as a medical physicist. Her research is applicative and focuses on the physical aspects of infusion therapy, especially on the safe use of Infusion Technology for multi-infusion applications. This field involves measurement and predictive modelling of catheter outflow drug concentrations but also practical strategies to improve accurate dosage of IV medication and education of infusion technology users. In hospital she is patient safety officer for medical technology and is in charge of the general medical physics residents training. Improving patient safety by training of medical personnel and finding solutions that work in clinical practice is what drives her every day to work.

Abstract

Introduction

Infusion therapy is challenging and dosing errors may occur due to physical phenomena related to the infusion hardware, despite the use of accurate syringe pumps. These errors typically occur after interventions, such as the exchange of a syringe. We aimed to characterize and quantify dosing errors due to a syringe exchange in relation to physical properties of infusion hardware.

Methods

An analytical simulation model was used to investigate dosing errors due to two different syringe exchange protocols (variations). Each protocol involved a fast syringe pump, containing a non-critical medication, and a slow syringe pump, containing a critical drug. The protocols were also reproduced in in vitro experiments to verify the simulation results. In addition, impact of syringe size, infusion set compliance, catheter diameter and the duration of the syringe exchange procedure on the quantity of the dosing errors was investigated.

Results

The syringe exchange of the slow pump resulted in an additional delay of up to 3536 seconds due to backflow. Syringe exchange of the fast pump resulted in an undiluted volume of critical drug (0.17 ml) accumulated in the infusion system, which may result in a dosing error rate of 2400 %. The quantity of the dosing errors are related to the syringe exchange duration; however, impact of infusion hardware properties is generally larger. Smaller syringes, catheters with larger diameters and less compliant infusion systems in general give rise to smaller dosing errors during a syringe exchange. If both lines are clamped, additional dosing errors can be prevented.

Conclusion

Infusion hardware has a substantial impact on the dosing errors during a syringe exchange. Clamping or blocking the infusion lines using, e. g. stopcocks, on all infusion lines during a syringe exchange is essential.

Introduction

In critical care, almost all critically ill patients require intravenous (IV) infusion therapy. However, infusion therapy is challenging and constitutes a high potential for adverse effects in critically ill patients (1, 2, 3, 4, 5, 6). There are several reasons for this. Many patients in critical care suffer from conditions that require the administration of critical drugs, such as, vasoactive drugs, sedatives, analgesia etc. For example, patients on the ICU are frequently diagnosed with hypotension and require the immediate administration of vasopressors. Because these drugs often have short half-lives, continuous administration is required. Moreover, these potent short- and rapid-acting vasoactive drugs have to be administered accurately as well as precisely. In order to meet these demands, syringe pumps are used. Another important difficulty is limited vascular access, the challenging nature of inserting a vascular access device (7) and the high risks of infection such as catheter-related sepsis (8, 9). To reduce the vascular access points to a minimum, multiple pumps are used to deliver their drugs to one central infusion set and catheter lumen. This principle is called multi-infusion (10) and is associated with significant flow rate variability and dosing errors, despite the use of syringe pumps (10, 11, 12). Even small dosing errors in vasoactive drugs may cause hemodynamic instability. This is especially true during clinical interventions (13). An important example of an intervention is a syringe exchange (also called syringe changeover), i. e. the renewal of an old, empty syringe with a full new one. Syringe exchanges are frequently conducted due to the limited capacity of syringes, which is typically 10 to 50 ml. From clinical practice and laboratory investigations it is known that a syringe exchange produces relatively large dosing errors and several techniques aiming to reduce these errors have been investigated (14, 15, 1617). However, to date the improvement achieved by technical innovations are limited. A recurring conclusion is that awareness and understanding of the physical mechanisms related to drug delivery by infusion may reduce errors (5, 18, 19). Argaud et al. (20) argued that operators may lack knowledge about syringe exchange mechanisms and therefore resort to poorly established empirical procedures. In a before and after intervention study they found that standardization of a syringe exchange procedure resulted in significantly less incidents (20).

Although laboratory studies have produced ample evidence that physical properties of the infusion hardware are related to dosing errors (21), the exact relationship between the syringe exchange duration, the infusion hardware and drug dosing errors remains elusive. Moreover, laboratory investigations are essentially limited to static infusion configurations, i. e. the properties of the hardware and therefore the physical parameters are either fixed or the variation is limited. Simulation studies may be able to establish this relationship better, which could help to provide a well-substantiated advice for hospital personnel.

Our objective is two-fold. First, to investigate the impact of physical parameters on dosing errors as a result of a syringe exchange in multi-infusion therapy using a realistic multi-infusion setup. Second, to investigate the impact on the dosing error of hypothetical situations due to a syringe exchange, in which the physical parameters of the infusion hardware, as well as syringe exchange duration time, are varied.

Theoretical simulation model

An analytical theoretical simulation model has been described previously (22). The model incorporates the most important (21) physical parameters necessary to simulate a low-flow multi-infusion system. The physical parameters used in this study are:

  1. Mechanical compliance (C) [volume per pressure]: which occurs due to the elasticity of components, especially the syringe.

  2. Flow resistance (R) (pressure per flow rate): which is due to the small diameters of some components, most notably the catheter.

Materials and methods

A typical clinical case

To start with, a typical NICU-case (Neonatal Intensive Care Unit) based on a patient treated on the NICU at the Wilhemina Children’s Hospital, Utrecht, the Netherlands (WKZ) is described. This case was investigated in order to acquire typical parameters of a challenging case of infusion therapy. Accordingly, the infusion hardware properties and flow rates used in this case are defined as the “typical clinical case” and used for the in vitro experiments and simulation explained later. Because the patient suffered from hypotension, IV-treatment with dopamine was started with a flow rate of 0.5 ml/h, defined as the “slow pump” (S) containing the “critical drug”. Multiple medications are typically co-administered with dopamine. For simplicity we defined one second pump containing the “non-critical drug” total parenteral nutrition (TPN) as the “fast pump” (F) with a flow rate of 12 ml/h. The pumps used were two Perfusor Bbraun syringe pumps (B.Braun, Melsungen, Germany) which were equipped with 50 ml-syringes B.Braun Omnifix (B.Braun, Melsungen, Germany). In our NICU, dopamine and TPN are typically administered through one lumen of a double lumen catheter. This leaves the other lumen open for drugs that are more compatible with temporary flow rate changes. This allows clinicians, for example, to administer a bolus. The co-administration of dopamine and TPN was achieved using a custom 173 cm central infusion set disposable (Impromediform GmbH, Lüdenscheid, Germany) (Figure 1). Both pumps were connected to the mixing point M2. Subsequently, the central infusion set was connected to the distal lumen of a 4 Fr double lumen catheter (Vygon, Ecouen, France). The catheter entered the patient trough the umbilical route.

Figure 1: Neonatal central infusion set disposable (173 cm), going upstream from mixing points M1 to M3.The Luerlock connector at M1 may be used for non-critical medication, not considered in this study. M2 is used for the pumps containing the non-critical and critical medication, which are co-administered through one line, respectively indicated as F and S. M3 is used for lipids, not considered in this study. VA indicates the outflow and the location where a vascular access device, e. g. the catheter, is typically connected.
Figure 1:

Neonatal central infusion set disposable (173 cm), going upstream from mixing points M1 to M3.

The Luerlock connector at M1 may be used for non-critical medication, not considered in this study. M2 is used for the pumps containing the non-critical and critical medication, which are co-administered through one line, respectively indicated as F and S. M3 is used for lipids, not considered in this study. VA indicates the outflow and the location where a vascular access device, e. g. the catheter, is typically connected.

Next, two protocols (A–B) are defined, each describing a different procedure of syringe exchange. These two situations are selected on the basis that they illustrate two main types of procedures that may occur in clinical practice and may produce dosing errors. In protocol A, syringe S is exchanged without clamping or stopping syringe F. In protocol B, syringe F is exchanged without clamping the infusion line S. See Figure 2 for a graphical representation of the protocols.

Figure 2: Schematic of a graph representing the set-point flow rates of two pumps F and S as a function of time, for protocols A–B (left).The Flow rates of the fast pump F and slow pump S are 12 and 0.5 ml/h, respectively (b). The syringe exchange is performed between toff and ton for a period of Texchange (0–120 seconds), according to the description of protocol A, in which the slow pump S is exchanged, and protocol B, in which the fast pump F is exchanged (c). Next, the delivery was continued, in this period dosing errors and delays in drug delivery were observed (d). The corresponding condition of the infusion setup is represented schematically (right). Compliances C1 and C2 resp. of the syringes F and S and resistances (impedances) resp. of the infusion lines F (R1) and S (R2) are indicated. Resistance of the catheter RL is indicated as well.
Figure 2:

Schematic of a graph representing the set-point flow rates of two pumps F and S as a function of time, for protocols A–B (left).

The Flow rates of the fast pump F and slow pump S are 12 and 0.5 ml/h, respectively (b). The syringe exchange is performed between toff and ton for a period of Texchange (0–120 seconds), according to the description of protocol A, in which the slow pump S is exchanged, and protocol B, in which the fast pump F is exchanged (c). Next, the delivery was continued, in this period dosing errors and delays in drug delivery were observed (d). The corresponding condition of the infusion setup is represented schematically (right). Compliances C1 and C2 resp. of the syringes F and S and resistances (impedances) resp. of the infusion lines F (R1) and S (R2) are indicated. Resistance of the catheter RL is indicated as well.

The duration of a syringe exchange procedure varied. The protocol for syringe replacement, if it exists at all, may differ per hospital. A variation of methodologies was observed to conduct the changeover. For example, if the infusion lines were also renewed, the exchange process took longer compared to the situation in which only the syringes were exchanged. From our own time measurements on the NICU it was found that a syringe exchange procedure, performed by an experienced nurse, takes approximately 30–120 seconds.

Mathematical modeling of a syringe change: Multiple causes and multiple effects

Input

The dosing errors are dependent on several physical parameters, which in turn are based on the physical properties of the infusion hardware. These physical parameters are used as input parameters for the model. In the following, each input parameter will be explained.

The flow rates were chosen according to the case described earlier. Next, the syringe exchange was simulated for the pumps S and/or F, according to the protocols A–B. In principle, a syringe exchange results in a temporary and abrupt decrease in flow rate from the set point to 0 ml/h (Figure 2). The syringe exchange duration (Texchange) was varied according to the clinical case described earlier as well, it was decided to add an extra margin in order to observe possible theoretical effects. For the simulation, Texchange was varied between 0–240 seconds. Appendix I explains the eqs (1) and (2) that were used for the evaluation of protocols A and B, respectively.

Varying hardware parameters within realistic boundaries

In order to investigate the influence of the physical parameters, low and high extremes for each parameter within realistic boundaries in critical care were explored. The quantities of the physical parameters were either acquired from literature, measured or calculated for this study. Table 1 lists all the physical parameters related to infusion hardware. See appendix II for the methodologies used to acquire the hardware parameters.

  1. Mechanical compliance (C):

Table 1:

Infusion hardware input parameters, compliance and resistance, respectively related to syringe and filter compliance and catheter resistance. See Appendix II for values that were measured.

Compliance (ml/Pa)Resistance (Pa per ml/h)
10-ml syringe1 Fr Catheter
0.159 × 10−51949
50-ml syringe4 Fr Catheter and flowmeter*
1.5 × 10−5 (25)23.68
50-ml syringe (non-rigid, filter)*7 Fr Catheter
2.1 × 10−5 (25)0.058
in vitro situation (4 Fr catheter + flowmeter)* 23.68 + 1121 = 1145
  1. *Value used in the in vitro simulation based on the generic clinical case with a neonatal infusion set including a filter.

  2. Value calculated using the Poiseuille law.

Compliance is mostly related to syringes. Some syringes are more rigid, additionally, smaller syringes are known to be less compliant (21, 23). However, some of the compliance is also caused by filters, the infusion line and the pump itself (24, 25). We previously found 2.1 ml/bar (2.1×10−5 ml /pa) for a relatively compliant setup including a filter and a 50-ml B.Braun syringe (B.Braun, Melsungen, Germany), which approximately corresponds to the compliance of the typical clinical case and in vitro setup (explained later). The filter was estimated to have a compliance of about 0.5 ml/bar (0.5×10−5 ml/pa) (24). For a more rigid setup with a 50-ml syringe, 1.0–1.5 ml/bar (1.0–1.5×10−5 ml /pa) was found (24); as an intermediate value 1.5 ml/bar was therefore used. A very small compliance of 0.159 ml/bar (0.159×10−5 ml/pa) was measured for a 10-ml Plastipak syringe (Becton Dickinson, NJ, USA) syringe (Appendix II).

  1. Flow resistance (R):

Resistance was mostly related to the catheters. To acquire a full range of resistive quantitates, the resistances of three catheters were acquired. The catheters were a Vygon Premicath 1-Fr (Ecouen, France), Vygon double lumen (proximal lumen used) umbilical catheter 4-Fr (Ecouen, France) and 7-Fr catheters Argon Careflow Single Lumen (Plano, TX, USA). The resistances were calculated and measured by producing a flow, with the use of a non-compliant syringe and syringe pump, Chemyx Nexus 3000 (Stafford, TX, USA) and an in-line pressure gauge SensorTechnics CTE8002AY4N and CTE8005AY4N (Berlin, Germany) (appendix II). See Table 1 for an overview of the compliance and resistance values.

Output

Several end point were investigated following the syringe exchange after ton (Figure 2) as a function of the input parameters, defined in the previous section. In protocol A the quantity Tno medication was acquired. This is the time that the patient does not receive the critical medication due the syringe exchange, in addition to the unavoidable, potentially substantial Tdeadvolume (caused by the dead volume, see discussion) (22) and Texchange. In protocol B the quantity Vundiluted was acquired. This is the amount of drug solution unintendedly not diluted by the faster pump F. The simulation results based on the multi-infusion system used in the in vitro experiments (explained later) are defined as “in vitro”.

The end point quantities of the simulation model and their dependent parameters are summarized as follows:

  1. Protocol A: Tno medication (R,C) [s]: The duration in seconds, in addition to Tdeadvolume and Texchange, in which the patient does not receive the critical medication presented in seconds as a function of resistance (R) and compliance (C).

  2. Protocol B: Vundiluted (R,C,Texchange,) [ml]: The volume in milliliters of critical undiluted medication administered to the patient as a function of resistance (R), compliance (C) and the duration of the exchange procedure (Texchange)

The equations that were used, were evaluated in Mathematica 10.3 Wolfram (Champaign, IL, USA).

In vitro experiments

In order to validate the simulation results experimentally, the syringe exchange procedures were mimicked in vitro in a laboratory according to the two protocols A–B (Figure 2). Two different in vitro experiments were performed to acquire Tno medication and Vundiluted experimentally for protocols A and B, respectively. Hardware properties and flow rates of the typical clinical case were used.

Protocol A

In order to measure Tno medication, the cumulative flow from pumps S and F at the catheter tip was measured using a Coriolis flowmeter M12P (Bronkhorst, Ruurlo, The Netherlands). The cumulative flow rate was 12.5 ml/h in steady state prior to the syringe exchange, i. e. Utotal (t)=Ufast (t) + Uslow (t) [ml/h], where Utotal, Uslow, Ufast and t are the total flow rate, the flow rates of pumps S and F, and time, respectively. After the syringe exchange, when the syringe S is reattached, the flow rate Ufast of pump F (12 ml/h) remains unaltered. Consequently, any flow registered by the flow meter lower than 12 ml/h necessarily flows towards the newly attached syringe S. The time from the point where the flow rate is lower than 12 ml/h, after syringe S is reattached, until the flow rate of 12 ml/h is reached again is defined as Tto syringe. The total volume produced during Tto syringe (while the total flow rate was lower than 12 ml/h) was defined as backflow (Vbackflow) [ml]. Once the measured flow rate is larger than 12 ml/h again, pump S (with a set point of 0.5 ml/h), thus produces a flow rate larger than 0 ml/h and therefore pushed the volume Vbackflow of the non-critical medication (drug F), still inside line S, beyond the mixing point. From this tipping point, pump S produces a flow rate of 0–0.5 ml/h. Assuming that Ufast retains a flow rate of 12 ml the whole time, Uslow is obtained as: Uslow (t) = Utotal (t) − Ufast (t). The time required by Uslow to push Vbackflow beyond the mixing point after the tipping point is defined as Tto patient. Adding Tto syringe and Tto patient together results in Tno medication. Figure 3(a) illustrates this measurement principle.

Figure 3: Illustration of the acquisition of Tno medication in protocol A (a).After reattaching syringe S, flow rates lower than pump F (12 ml/h) occur because a certain volume is not flowing towards the flow meter and therefore flows towards syringe S into line S. This volume (Vbackflow), is accumulated in a period Tto syringe. After pump S has been restarted and delivers a flow rate > 0 ml/h (indicated as Uslow in the illustration), the same volume Vbackflow flows through the mixing point in a period Tto patient. The time Tno medication results from adding the periods Tto syringe and Tto patient. Illustration of the acquisition of Vundiluted (error) in protocol B (b). After syringe F is exchanged, a volume Vundiluted from pump S (containing a dye in the experiment) is accumulated beyond the mixing point. At the outflow point, a spectrometer continuously measures the dye concentration while a flow meter measures the cumulative flow rate. When the accumulated volume from pump S flows out of the outflow point this is registered by the experimental setup as a deviation in the dosing rate. This dosing rate is proportional to the steady state flow rate of Uslow (produced by pump S), which allows Vundiluted (error) to be calculated. Vundiluted (error) is acquired by subtracting Vsteady state from the volume measured during the measurement period (Vdelivered). Steady state was defined as a fluctuation of less than ±10 % for 120 seconds.
Figure 3:

Illustration of the acquisition of Tno medication in protocol A (a).

After reattaching syringe S, flow rates lower than pump F (12 ml/h) occur because a certain volume is not flowing towards the flow meter and therefore flows towards syringe S into line S. This volume (Vbackflow), is accumulated in a period Tto syringe. After pump S has been restarted and delivers a flow rate > 0 ml/h (indicated as Uslow in the illustration), the same volume Vbackflow flows through the mixing point in a period Tto patient. The time Tno medication results from adding the periods Tto syringe and Tto patient. Illustration of the acquisition of Vundiluted (error) in protocol B (b). After syringe F is exchanged, a volume Vundiluted from pump S (containing a dye in the experiment) is accumulated beyond the mixing point. At the outflow point, a spectrometer continuously measures the dye concentration while a flow meter measures the cumulative flow rate. When the accumulated volume from pump S flows out of the outflow point this is registered by the experimental setup as a deviation in the dosing rate. This dosing rate is proportional to the steady state flow rate of Uslow (produced by pump S), which allows Vundiluted (error) to be calculated. Vundiluted (error) is acquired by subtracting Vsteady state from the volume measured during the measurement period (Vdelivered). Steady state was defined as a fluctuation of less than ±10 % for 120 seconds.

Similar to the simulation model output, the results are presented as Tno medication in seconds.

Protocol B

Instead of a critical and non-critical medication for pumps S and F, e. g. TPN or NaCl and Dopamine or Norepinephrine, we respectively used the dyes Indigo Carmine (20 mg/l) and Tartrazine (20 mg/l). The relevant physical properties of the dye solutions were assumed not to differ significantly from the actual drug solutions. To measure the dye concentration at the catheter tip, absorption spectrophotometry was used. This method was demonstrated before (11). The dye concentrations at the catheter tip were measured using a QE65000 spectrometer (Ocean Optics, Dunedin, FL, USA) within the visible light spectrum (250–2500 nm) at wavelengths of 610 nm and 425 nm for Indigo Carmine and Tartrazine, respectively. The cumulative flow rate was again measured using the Bronkhorst M12P flowmeter. Because both concentration and flow rate were measured, it was possible to acquire the value of Vundiluted (error), which is defined as follows:

Vundilutederror=VdeliveredVsteadystate

Where Vdelivered is the volume of drug S, delivered after the initiation of the syringe exchange procedure until steady state was reached again. Vsteady state is the volume delivered in the nominal situation by pump S, i. e. without dosing errors. In this case, steady state was determined as a period of 120 seconds, after the syringe exchange, in which the dosing rate fluctuated less than ±10 %. Vundiluted (error) was used to remove systematic errors and obtain only the dosing error relative to steady state. See Figure 3(b). The results of the simulation model were corrected accordingly: Vundiluted (error)=Vundiluted – (Uslow×Texchange)=Vundiluted – (0.5 ml/h×Texchange)

In this experiment, Vundiluted (error) in milliliters, is presented as a function of the duration of the syringe exchange procedure (Texchange). The experiment was performed with a Texchange of 30, 60 and 120 seconds.

The data were continuously recorded with a sample time of 1 second. The measurements were repeated three and five times (n=3 and 5) for protocols A and B, respectively.

Data analysis

The measurements were analyzed with Matlab 2014a (Mathworks, Natick, MA USA). All measurement results are presented as mean ± SD.

Results

Protocol A: Syringe exchange of the slow pump

Figure 4 shows an overview of the simulated results of Tno medication. Tno medication was between 0.008 and 3536 seconds according to the simulation model. It can be seen that Tno medication is virtually zero for the 7-Fr catheter for all of the syringes and almost one hour for the 1-Fr catheter and the 50-ml syringe with the non-rigid setup. Moreover, if smaller catheters were used, the relative impact of the syringe size/infusion set compliance, on Tno medication increased. The typical clinical case that was reproduced in the in vitro experiment resulted in a Tno medication of 1623 ± 122 seconds.

Figure 4: Extra delay time (Tno medication) caused by the requirement to push the backflow volume (Vbackflow) beyond the mixing point as a function of infusion hardware (1–7 Fr, 10–50 ml).These are the results from the computer simulation. Also indicated are the results for Tno medication in the in vitro situation, both measured and simulated.
Figure 4:

Extra delay time (Tno medication) caused by the requirement to push the backflow volume (Vbackflow) beyond the mixing point as a function of infusion hardware (1–7 Fr, 10–50 ml).

These are the results from the computer simulation. Also indicated are the results for Tno medication in the in vitro situation, both measured and simulated.

Protocol B: Syringe exchange of the fast pump

Figure 5 shows all the results for Vundiluted for each of the listed catheters and syringes/infusion sets. Vundiluted was between 1.1×10−6 and 0.17 ml after a syringe exchange duration (Texchange) of 120 seconds.

Figure 5: Results of the computer simulation model for Vundiluted, the excess volume accumulated beyond the mixing point as a function of Texchange, the duration of the syringe exchange.Note the logarithmic scale for the Y-axis. NR=non-rigid.
Figure 5:

Results of the computer simulation model for Vundiluted, the excess volume accumulated beyond the mixing point as a function of Texchange, the duration of the syringe exchange.

Note the logarithmic scale for the Y-axis. NR=non-rigid.

The results of the in vitro experiment, which reproduced the typical clinical case, are shown in Figure 6. The simulation results, using input parameters corresponding to the in vitro experiment are also illustrated in Figure 6.

Figure 6: Comparison of Vundiluted (error) acquired from the in vitro experiment and the simulation model results according to the input parameters based on the in vitro experiment.
Figure 6:

Comparison of Vundiluted (error) acquired from the in vitro experiment and the simulation model results according to the input parameters based on the in vitro experiment.

Discussion

The results show that each mechanism results in counter-intuitive dosing errors, which are strongly related to the mechanical and physical properties of the infusion hardware used. It was shown that if the slow pump (S) is exchanged, a significant additional delay of up to almost one hour may occur until the desired delivery of the critical medication has been reestablished. Next, it was shown that if the fast pump (F) is exchanged, a significant amount of undiluted critical drug was administered, which causes an overdose. If both infusion lines are clamped, none of the physical mechanisms described in this study will produce dosing errors or delays during a syringe exchange. The results therefore reinforce the importance of clamping or otherwise blocking of all infusion lines during a syringe exchange. The results also show that smaller syringes and catheters with larger diameters reduce the quantity of dosing errors. Catheter diameter, however, is limited by the size of the vasculature. For example, in preterm neonates these may be very small. Smaller syringes, also have disadvantages. First of all, smaller syringe deplete faster than larger syringes and are therefore required to be replaced more often. Second, smaller syringes produce much more pressure when a similar force is introduced at the plunger in comparison to larger syringes. Among other reasons, this may pose a serious danger in case a vascular access device is inserted erroneously, such as an accidental subcutaneous position.

It is known that a syringe exchange causes errors in drugs administration. Backflow has also been studied extensively (10, 13, 26). The impact of compliance on backflow was investigated in a simulation study before (27). In that study, backflow was caused by an occlusion in the infusion set, beyond the mixing point. Due to the compliance, a volume of fluid was allowed to accumulate. This volume which was bigger for a 50-ml syringe compared to a 10 ml-syringe. These results are thus in agreement with our study (27).

In our study a set of realistic clinical situations were assessed and simulated for a range of infusion components. We have provided a generalized overview of all the important factors that impact dosing errors during a syringe exchange. These findings can be used by hospital personnel to substantiate the decisions for a syringe exchange protocol as well as the decisions for using certain infusion hardware.

Physical interpretation and clinical considerations

Refer to Appendix I for a comprehensive explanation of the physical effects involved in the simulated mechanisms.

Protocol (A) Whereas clinicians and nurses will be aware of the effects of Texchange, and hence will try to minimize the duration of the syringe exchange procedure in order to minimize Texchange and its effects, they might be unaware of the delay caused by Tno medication.

From the equations generated by the model (Appendix I), the mechanism behind this volume can be understood and is explained as follows. The slow syringe S was exchanged, however, no pressure build-up has been realized inside the new syringe of the slow pump S. Consequently, after the new syringe is connected to the infusion set, the direction of the least pressure will be towards this newly attached syringe. The storage of this volume is allowed because the syringe is somewhat compliant, i. e. some parts of the syringe are compressible, most notably the plunger. This, in turn, allows a volume of fluid (Vbackflow) to be pushed back from the infusion set, beyond the mixing point, into the infusion line S during the compression of the syringe connected to the line S. This volume of fluid consists of the “non-critical drug” originating from infusion line F, instead of the “critical drug” from the infusion line S (Appendix I). Therefore it will take some time, Tno medication, before the critical medication reaches the mixing point again, and hence the total time that no critical drug S is fed into the mixing point amounts to Texchange+Tno medication.

As the results show, Tno medication is dependent on the hardware properties of the infusion system. Specifically, the amount of the backflow allowed is strongly dependent on the compliance of syringe S (C2). This is because the compressibility of the syringe allows the fluid from the fast syringe, fluid F, to flow back in the direction of the slow syringe, syringe S (Appendix I). Therefore, smaller syringes, which are typically much less compliant, will result in less potential backflow. Moreover, the resistance of the catheter (RL) has a substantial impact on Tno medication, if the resistance of the catheter is low, the affinity for the fluid to flow back into syringe S is lower. These results are immediately evident from Figure 4. In fact, the results show that in the in vitro experiment, the problems are only clinically relevant with catheters smaller than 4 Fr. Additionally, Tno medication is dependent on the ratio of the flow rates between the fast and the slower pump (Appendix I). In the presented typical clinical case of this simulation, the flow rates of pump S and F were 0.5 and 12 ml/h. Consequently, it takes considerably longer to push the non-critical drug F out of the infusion line S, than the time that was required to push the drug into line S.

Clinicians should also consider the dangers if the syringe is outside of the pump. In this case the plunger is not fixated, which may allow free flow to occur (28). Moreover, during the experiments it was empirically observed that the volume of drug F, introduced in line S due to backflow, might be even larger in this case as the plunger is simply allowed to be pushed backwards, out of the syringe. The extent of this is dependent on the pressure coming from pump F and the friction of the plunger in the syringe, which also may have a significant impact (29) and may differ per syringe type.

As mentioned, besides delays caused by Texchange and Tno medication there is also Tdeadvolume, which is caused by the dead volume of the infusion set. Therefore, the (near) absence of drug S will occur only after Tdeadvolume, the time required for the fluid to travel through the dead volume from the mixing point to the catheter tip into the patient’s blood stream. Tdeadvolume is dependent on the physical parameters: quantity of the dead volume, the total flow rate (Utotal), the laminar flow profile as described by Poiseuille and (predominantly radial) diffusion (22, 30, 31, 32, 33). In the in vitro experiment, the dead volume of the infusion set was approximately 0.4 ml, resulting in a Tdeadvolume of at least 58 seconds. However, due to the physical parameters it will take longer to reestablish the intended dosing rate. Nevertheless, Tno medication was substantially larger than Tdeadvolume in the simulation of the in vitro experiment with 2034 seconds. Contrary, Tno medication is smaller in the cases of the 4 and 7-Fr catheter. However, the additional delay of Tno medication may still amount to almost a minute, which may be unwanted in the case of, for example, a patient with severe hypotension. This mechanism therefore shows that delays in multi-infusion systems are caused by a multitude of physical effects.

During the experiments, it was empirically observed that the drug F will enter infusion line S. In the case of this laminar flow drug solutions closer to the walls of the infusion line move more slowly in the axial direction, towards the patient and remain more unaffected by the flow of drug F. This fundamental principle is explained by Poiseuille (22). Therefore the Tno medication simulated and measured is a worst case scenario, in practice some critical medication may still be delivered.

Protocol (B) From the simulation results, the errors resulting from the syringe exchange of the fast syringe can be explained as follows. During the exchange of the fast pump F, the slow pump S is stopped, but not clamped. As a result, due to compliance effects, the critical drug solution from the pump S, not diluted by the fast pump F, oozes into the common line beyond the mixing point. The volume of Vundiluted depends on the Texchange and many other factors, such as the initial flow (before Texchange) of the pump F, the mechanical compliances C1 and C2 of syringes F and S, respectively, and the resistances R1, R2, and RL of lines F, S, and the catheter, respectively (see Figure 2 and Appendix I) (22).

The effects of this procedure may be even more counter-intuitive as protocol A. Consequently, the probability that an adverse effect will occur may also be greater. From the results it is evident that the quantity of Vundiluted is strongly dependent on the resistance and compliance. Again, with catheter-diameters larger than 4-Fr, the effects are minimal and smaller syringes also minimize the amount of Vundiluted. However, even in a situation with zero compliance and resistance, a quantity Vundiluted will be produced if pump F is not stopped, although this amount is substantially smaller than the amount that results from the infusion system’s compliance. The clinical relevance of Vundiluted is dependent on the time required to deliver this volume to the patient, which in turn is dependent on the current flow rate. For example, if the largest measured Vundiluted of 0.17 ml is administered with 0.5 ml/h it is, in fact, administered with the intended dosing rate in 1224 seconds. However, in the typical clinical case, Vundiluted was administered with a flow rate of 12.5 ml/h in 49 seconds, which means that this quantity was administered with an excess dosing rate of 2400 %. It must be stated that such short dosing errors are typically clinically relevant for drugs with short onset times, e. g. noradrenaline and dopamine. In addition, a high concentration of the critical drug S, typically used for patients who cannot handle fluids, also increases the probability for an adverse event.

Limitations and future research

The spectrometer was not used in protocol A because a higher accuracy could be achieved by using only the flowmeter. However, it should be noted that while the flow rate of pump S is increased, the flow rate of pump F could be decreased slightly, due to compliance and resistance effects, therefore impacting Tno medication as measured in the in vitro experiment. Nevertheless, because the flow rate from pump F is much higher, this effect is negligible and using only the flow meter allows the acquisition of Tno medication relatively accurately.

A discrepancy could be observed between the experimental and simulated values. All in vitro results were systematically underestimating the simulated results, which suggest an overestimation of the compliance of this particular infusion setup, as described in the typical clinical case. Although the input values that we measured were typically quite accurate with relatively small confidence intervals and R2 > 0.98 (see Appendix II), repeated measurements of, e. g. multiple catheters/syringes of the same type, may give an additional error to the input values of the model. The output values of the model have a linear sensitivity to errors in the input parameters (R and C). In the case errors in the input parameters would be added, the simulated output may be approximated more closely by the in vitro experiment. Furthermore, the experimental in vitro results in protocol A did follow the expected exponential curve. We did not have the instruments available to measure the physical properties of the infusion setup in the typical clinical case in more detail, which is a likely cause for the discrepancy. Furthermore, only the results of the in vitro experiments using the 50 ml syringe and the non-rigid set, corresponding to a neonatal situation, were presented. The signal to noise ratio prevented us from obtaining usable results for a 10 ml syringe, which indicated that the measured quantities were too low. Obtaining usable experimental results for the smaller syringes would further validate the results of the theoretical simulation model. It is recommended to conduct these experiments with better instruments in future research.

In clinical practice, TPN or NaCl may be administered using a volumetric or gravity driven pump. These pumps do not have compliance in a similar fashion as syringes. Volumetric pumps are typically connected through specific infusion lines with a compliant part inside the actuator cartridge, which enables the commonly used peristaltic mechanism to operate, while the rest of the infusion line is less compliant. Although typical infusion lines were shown to be compliant (25), the impact of the infusion lines used in volumetric pumps on the physical mechanism described in this study was not investigated. Nevertheless, since in protocol A, Tno medication is mostly related to the compliance of syringe S (C2, see Figure 2) there will probably not be any substantial differences in the results. Likewise, in protocol B, the compliance of syringe S is also the most important parameter, in combination with the resistance of the catheter. Another noteworthy point is the flow rate of pump F (12 ml/h). This was chosen to observe the effects studied in this paper with sufficient confidence in the in vitro experiments. The flow rate is somewhat high as the flow rate in the NICU is preferred to be kept as low as possible. The formula for Tno medication in Appendix I can be used to calculate other flow rate combinations. In clinical practice this flow rate may also be produced by multiple pumps instead of just one syringe pump. However, in all these situations, the overall results of this study will be maintained.

Conclusion

It was shown that a syringe exchange produces dosing errors as well as substantial delay times in certain cases. This may have severe implications for patient safety. The quantity of the dosing error is dependent on the exchange duration and the physical properties of the infusion hardware used, most notably the syringes and the catheter. Smaller syringes, catheters with larger inner-diameters and less compliant infusion systems in general, result in smaller dosing errors during a syringe exchange. It is recommended to consider these factors when selecting infusion hardware. By clamping all infusion lines during a syringe exchange, most of the unwanted results can be prevented or mitigated. In fact, this is more important than performing the syringe exchange procedure quickly since the extra delay due to compliance effects can be unexpectedly large, as has been shown in this study. It is therefore advised to clamp or otherwise block all infusion lines that are co-administering drugs through the same infusion set and catheter lumen as the syringe that is exchanged.

About the authors

Roland A. Snijder

Roland Snijder obtained his PhD at Utrecht University in 2016 and a master’s degree in Biomedical Engineering at the University of Groningen with a specialization curriculum in the area of medical physics (medical instrumentation and imaging) in 2012. As a PhD candidate, Roland conducted research at the department of Medical Technology and Clinical Physics of University Medical Center Utrecht (UMC Utrecht) in cooperation with the department of Clinical Pharmacy. Roland’s research focused on investigating physical causes of dosing errors in multi-infusion systems. Specifically, the aim was to investigate the origin and characteristics of these dosing errors using in vitro measurement methods and theoretical modeling. Moreover, the clinical relevance of these dosing errors was investigated using pharmacological theory and a retrospective clinical study. The research was conducted within the framework of the European Metrology Research Programme (EMRP).

Maurits K. Konings

Maurits Konings, PhD, is a physicist working at the Department of Medical Technology and Clinical Physics of the University Medical Center Utrecht. His research interests are non-invasive monitoring of cardiac function, new methods for non-invasive neurostimulation and neuromonitoring, and multi-infusion. His is inventor of over 8 patents in the field of medical technology.

Agnes van den Hoogen

Agnes van den Hoogen is a nurse scientist and teacher at the University Medical Centre of Utrecht (UMCU) – Wilhelmina Children’s Hospital in Utrecht, The Netherlands. She obtained her PhD at the University of Utrecht, medical faculty of the University Medical Centre Utrecht. Topic of her thesis: Infections in Neonatal Intensive: Care Prevalence, Prevention and Antibiotic use. Clinical Interest: Neonatal infections, Management of Intra Venous access in Neonates, Family Centred Care, Family Empowerment, Ethics, Epidemiology and Education

Annemoon M.D.E. Timmerman

Annemoon Timmerman received her PhD in 2002 at Delft University of Technology. At 2005 she attained her registration as a medical physicist at Erasmus Medical Center. Since 2005 she has been working at the University Medical Center Utrecht as a medical physicist. Her research is applicative and focuses on the physical aspects of infusion therapy, especially on the safe use of Infusion Technology for multi-infusion applications. This field involves measurement and predictive modelling of catheter outflow drug concentrations but also practical strategies to improve accurate dosage of IV medication and education of infusion technology users. In hospital she is patient safety officer for medical technology and is in charge of the general medical physics residents training. Improving patient safety by training of medical personnel and finding solutions that work in clinical practice is what drives her every day to work.

Acknowledgement

The authors would like to thank I. Varga for her contributions to the experimental work.

  1. Conflict of interest statement: Authors state no conflict of interest. All authors have read the journal’s Publication ethics and publication malpractice statement available at the journal’s website and hereby confirm that they comply with all its parts applicable to the present scientific work.

References

1. Westbrook JI, Rob MI, Woods A, Parry D. Errors in the administration of intravenous medications in hospital and the role of correct procedures and nurse experience. BMJ Qual Saf 2011;20:1027–1034.10.1136/bmjqs-2011-000089Search in Google Scholar PubMed PubMed Central

2. Kaushal R. Medication errors and adverse drug events in pediatric inpatients. Jama 2001;285:2114.10.1001/jama.285.16.2114Search in Google Scholar PubMed

3. Thomas AN, Panchagnula U. Medication-related patient safety incidents in critical care: a review of reports to the UK National Patient Safety Agency. Anaesthesia 2008;63:726–733.10.1111/j.1365-2044.2008.05485.xSearch in Google Scholar PubMed

4. Lovich MA, Wakim MG, Wei A, Parker MJ, Maslov MY, Pezone MJ et al. Drug infusion system manifold dead-volume impacts the delivery response time to changes in infused medication doses in vitro and also in vivo in anesthetized swine. Anesth Analg 2013;117:1313–1318.10.1213/ANE.0b013e3182a76f3bSearch in Google Scholar PubMed

5. Sherwin CT, Medlicott NJ, Reith DM, Broadbent RS. Intravenous drug delivery in neonates: lessons learnt. Arch Dis Child 2014;99:590–594.10.1136/archdischild-2013-304887Search in Google Scholar PubMed

6. Husch M, Sullivan C, Rooney D, Barnard C, Fotis M, Clarke J et al. Insights from the sharp end of intravenous medication errors: Implications for infusion pump technology. Qual Saf Health Care 2005;14:80–86.10.1136/qshc.2004.011957Search in Google Scholar PubMed PubMed Central

7. Moureau N, Lamperti M, Kelly LJ, Dawson R, Elbarbary M, Van Boxtel AJH et al. Evidence-based consensus on the insertion of central venous access devices: definition of minimal requirements for training. Br J Anaesth 2013;110:347–356.10.1093/bja/aes499Search in Google Scholar PubMed

8. Fletcher S. Catheter-related bloodstream infection. Contin Educ Anaesthesia, Crit Care Pain 2005;5:49–51.10.1093/bjaceaccp/mki011Search in Google Scholar

9. Legemaat MM, Jongerden IP, van Rens RMFPT, Zielman M. van den Hoogen A. Effect of a vascular access team on central line-associated bloodstream infections in infants admitted to a neonatal intensive care unit: a systematic review. Int J Nurs Stud 2015;52:1003–1010.10.1016/j.ijnurstu.2014.11.010Search in Google Scholar PubMed

10. Décaudin B, Dewulf S, Lannoy D, Simon N, Secq A, Barthélémy C et al. Impact of multiaccess infusion devices on in vitro drug delivery during multi-infusion therapy. Anesth Analg 2009;109:1147–1155.10.1213/ane.0b013e3181ae06e3Search in Google Scholar PubMed

11. Timmerman AM, Snijder RA, Lucas P, Lagerweij MC, Radermacher JH, Konings MK. How physical infusion system parameters cause clinically relevant dose deviations after setpoint changes. Biomed Tech 2015;60:365–376.10.1515/bmt-2014-0139Search in Google Scholar PubMed

12. Lovich MA, Doles J, Peterfreund RA. The impact of carrier flow rate and infusion set dead-volume on the dynamics of intravenous drug delivery. Anesth Analg 2005;100:1048–1055.10.1213/01.ANE.0000146942.51020.88Search in Google Scholar PubMed

13. Peterfreund RA, Philip JH. Critical parameters in drug delivery by intravenous infusion. Expert Opin Drug Deliv2013;10:1095–1108.10.1517/17425247.2013.785519Search in Google Scholar PubMed

14. Arino M, Barrington JP, Morrison AL, Gillies D. Management of the changeover of inotrope infusions in children. Intensive Crit Care Nurs 2004;20:275–280.10.1016/j.iccn.2004.06.003Search in Google Scholar PubMed

15. Cour M, Hernu R, Bénet T, Robert JM, Regad D, Chabert B et al. Benefits of smart pumps for automated changeovers of vasoactive drug infusion pumps: a quasi-experimental study. Br J Anaesth 2013;111:818–824.10.1093/bja/aet199Search in Google Scholar PubMed

16. De Barbieri I, Frigo AC, Zampieron A. Quick change versus double pump while changing the infusion of inotropes: an experimental study. Nurs Crit Care 2009;14:200–206.10.1111/j.1478-5153.2009.00333.xSearch in Google Scholar PubMed

17. Genay S, Décaudin B, Lédé S, Feutry F, Barthélémy C, Lebuffe G et al. In vitro comparison of two changeover methods for vasoactive drug infusion pumps: Quick-change versus automated relay. Biomed Eng /Biomed Tech 2015;60:377–380.10.1515/bmt-2014-0140Search in Google Scholar PubMed

18. Maiguy-Foinard A, Genay S, Lannoy D, Barthélémy C, Lebuffe G, Debaene B et al. Criteria for choosing an intravenous infusion line intended for multidrug infusion in anaesthesia and intensive care units. Anaesth Crit Care Pain Med 2017;36:53–63.10.1016/j.accpm.2016.02.007Search in Google Scholar PubMed

19. Snijder RA, Knape JTA, Egberts TCG, Timmerman AMDE. Hypertensive crisis during norepinephrine syringe exchange. A A Case Reports 2017;8:178–181.10.1213/XAA.0000000000000458Search in Google Scholar PubMed

20. Argaud L, Cour M, Martin O, Saint-Denis M, Ferry T, Goyatton A et al. Changeovers of vasoactive drug infusion pumps: impact of a quality improvement program. Crit Care 2007;11:R133.10.1186/cc6209Search in Google Scholar PubMed PubMed Central

21. Snijder RA, Konings MK, Lucas P, Egberts TC, Timmerman AD. Flow variability and its physical causes in infusion technology: a systematic review of in vitro measurement and modeling studies. Biomed Tech 2015;60:277–300.10.1515/bmt-2014-0148Search in Google Scholar PubMed

22. Konings MK, Snijder RA, Radermacher JH, Timmerman AMDE. Analytical method for calculation of deviations from intended dosages during multi-infusion. Biomed Eng Online 2017;16:18.10.1186/s12938-016-0309-4Search in Google Scholar PubMed PubMed Central

23. Neff SB, Neff TA, Gerber S, Weiss M. Flow rate, syringe size and architecture are critical to start-up performance of syringe pumps. Eur J Anaesthesiol 2007;24:602.10.1017/S0265021507000026Search in Google Scholar PubMed

24. Batista E, Almeida N, Furtado A, Filipe E, Souse L, Rui M et al. Assessment of drug delivery devices. Biomed Eng/Biomed Tech 2015;60:347–357.10.1515/bmt-2014-0138Search in Google Scholar PubMed

25. Weiss M, Banziger O, Neff T, Fanconi S. Influence of infusion line compliance on drug delivery rate during acute line loop formation. Intensive Care Med 2000;26:776–779.10.1007/s001340051246Search in Google Scholar PubMed

26. Ellger B, Kiski D, Diem E, van den Heuvel I, Freise H, Van Aken H et al. Non-return valves do not prevent backflow and bacterial contamination of intravenous infusions. J Hosp Infect 2011;78:31–35.10.1016/j.jhin.2010.12.015Search in Google Scholar PubMed

27. Murphy RS, Wilcox SJ. The link between intravenous multiple pump flow errors and infusion system mechanical compliance. Anesth Analg 2010;110:1297–1302.10.1213/ANE.0b013e3181d56758Search in Google Scholar PubMed

28. Cohen MR, Davis NM. Free flow associated with electronic infusion devices: an underestimated danger. Hosp Pharm 1992;27:384–390.Search in Google Scholar

29. Schmidt N, Saez C, Seri I, Maturana A. Impact of syringe size on the performance of infusion pumps at low flow rates. Pediatrcrit Care Med 2010;11:282–286.10.1097/PCC.0b013e3181c31848Search in Google Scholar PubMed

30. Taylor G. Dispersion of soluble matter in solvent flowing slowly through a tube. Proc R Soc A Math Phys Eng Sci 1953;219:186–203.Search in Google Scholar

31. Gurmen F. Models for nonideal reactors (University of Michigan) [Internet]. 2007 [cited January 1, 2016]. Available from: http://www.umich.edu/~elements/fogler&gurmen/html/course/lectures/fourteen/index.htmSearch in Google Scholar

32. Hutton P, Thornberry EA. Factors affecting delivery of drug through extension tubing. Br J Anaesth 1986;58:1141–1148.10.1093/bja/58.10.1141Search in Google Scholar PubMed

33. Parker MJ, Lovich MA, Tsao AC, Wei AE, Wakim MG, Maslov MY et al. Computer control of drug delivery by continuous intravenous infusion. Anesthesiology 2015;122:647–658.10.1097/ALN.0000000000000519Search in Google Scholar PubMed


Supplemental Material

The online version of this article offers supplementary material (DOI: https://doi.org/10.1515/pthp-2017-0002).


Received: 2017-1-10
Revised: 2017-4-13
Accepted: 2017-4-24
Published Online: 2017-5-16
Published in Print: 2017-8-28

© 2017 Walter de Gruyter GmbH, Berlin/Boston