Failure Analysis of a Femoral Hip Stem Made of Stainless Steel After Short Time of Exposure

Żaneta Anna Mierzejewska
  • Faculty of Mechanical Engineering, Division of Materials and Biomedical Engineering, Bialystok University of Technology, ul. Wiejska 45c, 15-351 Białystok, Poland
  • Email
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Zbigniew Oksiuta
  • Faculty of Mechanical Engineering, Division of Materials and Biomedical Engineering, Bialystok University of Technology, ul. Wiejska 45c, 15-351 Białystok, Poland
  • Email
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
Published Online: 2014-12-20 | DOI: https://doi.org/10.2478/ama-2014-0026

Abstract

This paper describes a case study summarising the failure analysis of a stainless steel femoral stem, which failed prematurely within 36 months of exploitation in human body. In order to determine the mechanism of failure, a broken stem component were analyzed by means of macroscopic and microscopic obserwations and hardness measurements. Metallurgical obserwations revealed that the tested material does fulfill ASTM requirements. Scanning electron microscopy images revealed the presence of stress-induced cracking. The results of the hardness revealed significant nonuniformity from the surface towards the inner part of the stem. It is assumed that any discontinuity or defect on the fracture surface of the stem acted as preferential site for a crack nucleation and propagation by fatigue until the cross section of stem was not able to sustain a load generated by a patient.

Keywords: Hip Replacement; Endoprosthesis Stem Fracture; Isolated Fracture; Total Hip Arthroplasty Complications; Fatigue Fracture

References

  • 1. Brown T. (2012) Preventing Mechanical Failures - An Introduction to Failure Mode Identification, A Culture of Reliability, Vol. 3, 46-48.Google Scholar

  • 2. Charnley J. (1975), Fracture of femoral prostheses in total hip replacement : a clinical study. Clin Orthop, 1 1 1 :105-20.Google Scholar

  • 3. Delaunay, C. P., Kapandji, A. I. (1996), Primary total hip arthroplasty with the Karl Zweymuller firstgeneration cementless prosthesis: a 5 to 9 year retrospective study, J. Arthroplasty, 11, 643.652.Google Scholar

  • 4. Galante J. (1980), Current Concept Review Causes of Fractures of the Femoral Component in Total Hip Replacement, Journal of Bone and Joint Surgery, Vol. 62 (4), Vol. 62(4), 670-673.Google Scholar

  • 5. Griza S., Vieira dos Santos S., Ueki M. M., Bertoni F., Strohaecker T.R. (2013), Case study and analysis of a fatigue failure in a THA stem, Engineering Failure Analysis, 28, 166-175.CrossrefGoogle Scholar

  • 6. HernandezRodriguez M. A. L., OrtegaSaenz J. A., CotrerasHernandez G. R. (2010), Failure analysis of a total hip prosthesis implanted in active patient, Journal of the Mechanical Behavior of Biomedical Materials, Vol. 3 (8), 619-622.Web of ScienceGoogle Scholar

  • 7. Jarvi K., Kerry R. M. (2007), Case Report Segmental Stem Fracture of a Cemented Femoral Prosthesis, The Journal of Arthroplasty, Vol. 22(4), 612-616.PubMedWeb of ScienceCrossrefGoogle Scholar

  • 8. Kamachi M. U., Sridhar T. M., Eliaz N., Baldev R. A. J. (2003), Failures of stainless steel orthopaedic devices: causes and remedies, Corros Rev, 21, 231-67.Google Scholar

  • 9. Kishida Y.,Sugano N., Ohzono K., Sakai T., Nishii T., Yoshikawa H. (2002), Stem Fracture of the Cementless Spongy Metal Lu¨ beck Hip Prosthesis, The Journal of Arthroplasty, Vol. 17 (8), 1021-1027.CrossrefGoogle Scholar

  • 10. Lakstein D., Eliaz N., Levi O, Backstein D., Kosashvili Y., Safir O., Gross A. E. (2011), Fracture of Cementless Femoral Stems at the Mid-Stem Junction, J Bone Joint Surg Am., 93, 57-65.CrossrefPubMedGoogle Scholar

  • 11. Malchau H, Herberts P, Eisler T, Garellick G, Soderman P. (2002), The Swedish Total Hip Replacement Register, J Bone Joint Surg Am, 84-A, Suppl 2, 2-20.Google Scholar

  • 12. Martens M., Aernoudt E., De Meester P., Ducheyne P., Muller J. C., Delangh R., Kestelijn P. (1974), Factors in the Mechanical Failure of the Femoral Component in Total Hip Prosthesis, Acta Orthop Scandinavica, 45(5), 693-710.CrossrefGoogle Scholar

  • 13. Mollan R. A. B., Watters P. H., Steele R., McClelland C. J. (1984), Failure of the femoral component in the Howse total hip arthroplasty, C/in Orthop, 190, 142-7.Google Scholar

  • 14. Petch N. J. (1953), The cleavadge strength of policrystalls, JISi, 174, 25-28.Google Scholar

  • 15. Roffey P. (2012), Case study: Failure of a high nitrogen stainless steel femoral stem, Engineering Failure Analysis, Vol. 20, 173-182.Web of ScienceCrossrefGoogle Scholar

  • 16. Scheerlinck T, Druyts P, Casteleyn P. (2004), The use of primary total hip arthroplasty in university hospitals of the European Union, Acta Orthop Belg, 70(3), 231-9.PubMedGoogle Scholar

  • 17. Sen R. K., Mootha A. K., Saini R., Kumar V. (2009) Segmental Fracture Of A Cemented Femoral Stem - A Case Report And Review Of Litrature, The Internet Journal of Orthopedic Surgery, Vol. 13 (1), 13-15.Google Scholar

  • 18. Standard Specification for Wrought Nitrogen Strengthened 21 Chromium-10 Nickel-3 Manganese-2.5 Molybdenum Stainless Steel Alloy Bar for Surgical Implants, American Society for Testing and Materials, Materals, West Conshohocken, PA, ASTM F1586.Google Scholar

  • 19. Wilson L. F., Nolan J. F., Wood-Waddington M. B. (1992), Fracture of the femoral stem of the ring TCH hip prosthesis, J Bone Joint Surg Br, 74(5), 725-8.PubMedGoogle Scholar

About the article

Published Online: 2014-12-20

Published in Print: 2014-09-01


Citation Information: Acta Mechanica et Automatica, Volume 8, Issue 3, Pages 146–150, ISSN (Online) 2300-5319, DOI: https://doi.org/10.2478/ama-2014-0026.

Export Citation

© Bialystok University of Technology. This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License. BY-NC-ND 3.0

Citing Articles

Here you can find all Crossref-listed publications in which this article is cited. If you would like to receive automatic email messages as soon as this article is cited in other publications, simply activate the “Citation Alert” on the top of this page.

[1]
Zbigniew Oksiuta, Zbigniew Mazurkiewicz, Eliza Romanczuk, Marcin Andrzejewicz, and Agnieszka Krawczynska
Engineering Failure Analysis, 2017, Volume 72, Page 1

Comments (0)

Please log in or register to comment.
Log in