Jump to ContentJump to Main Navigation
Show Summary Details
More options …

Biomedical Engineering / Biomedizinische Technik

Joint Journal of the German Society for Biomedical Engineering in VDE and the Austrian and Swiss Societies for Biomedical Engineering and the German Society of Biomaterials

Editor-in-Chief: Dössel, Olaf

Editorial Board: Augat, Peter / Habibović, Pamela / Haueisen, Jens / Jahnen-Dechent, Wilhelm / Jockenhoevel, Stefan / Knaup-Gregori, Petra / Lenarz, Thomas / Leonhardt, Steffen / Plank, Gernot / Radermacher, Klaus M. / Schkommodau, Erik / Stieglitz, Thomas / Boenick, Ulrich / Jaramaz, Branislav / Kraft, Marc / Lenthe, Harry / Lo, Benny / Mainardi, Luca / Micera, Silvestro / Penzel, Thomas / Robitzki, Andrea A. / Schaeffter, Tobias / Snedeker, Jess G. / Sörnmo, Leif / Sugano, Nobuhiko / Werner, Jürgen /

6 Issues per year


IMPACT FACTOR 2016: 0.915
5-year IMPACT FACTOR: 1.263

CiteScore 2017: 0.48

SCImago Journal Rank (SJR) 2017: 0.202
Source Normalized Impact per Paper (SNIP) 2017: 0.356

Online
ISSN
1862-278X
See all formats and pricing
More options …
Volume 57, Issue 3

Issues

Volume 57 (2012)

Biomechanical analysis of anterior stabilization techniques for different partial and total vertebral corpectomy defect models

Benjamin Ulmar
  • Chirurgische Klinik und Poliklinik, Abteilung Unfall- und Wiederherstellungschirurgie, Universitätsmedizin Rostock, Rostock, Germany
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Alexander Disch
  • Center for Musculoskeletal Surgery, Musculoskeletal Tumor Surgery Section Charité-University Medicine Berlin, Berlin, Germany
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Stefanie Erhart / Werner Schmoelz
Published Online: 2012-05-24 | DOI: https://doi.org/10.1515/bmt-2011-0107

Abstract

Introduction: Restoration of the anterior spinal profile and regular load-bearing are the main goals in treating anterior spinal defects in cases of fracture. In large vertebral defects, the use of vertebral body replacements (VBRs) is common. For isolated anterior operations, VBRs are usually combined with antero-lateral angular stable or polyaxial plates. In preclinical biomechanical testing of spinal fracture stabilizations, complete corpectomies are often applied for fracture simulation. Complete corpectomies are reproducible and easily standardized but present a maximum instability not indicated for isolated anterior instrumentation. Therefore, this study investigates the stabilizing effect of VBRs and anterior plating systems on three different vertebral body defect models.

Methods: Using 24 thoracolumbar spines, two different partial (PC 1+2) and one total corpectomy defect (CC) models were simulated and stabilized using different combinations of two VBRs and three different anterior plating systems. After anterior stabilization, range of motion (RoM) in the three main movement planes was evaluated.

Results: RoM was significantly reduced in all three motion planes for both stabilized partial corpectomy defect models compared to the stabilized complete corpectomy model and the intact specimens. The stabilized complete corpectomy defect model did not reduce RoM in flexion/extension and axial rotation compared to the intact state.

Conclusion: Both partial corpectomy models are suitable for biomechanical testing of isolated anterior stabilization. A complete corpectomy model is not suitable to test isolated anterior instrumentation, as this simulates a rotationally unstable fracture, where isolated anterior instrumentation is not indicated in clinical practice.

Keywords: anterior plating; biomechanics; in situ distractible cage; partial and complete corpectomy; vertebral body replacement; vertebral fracture treatment

About the article

Corresponding author: Dr. Werner Schmoelz, Department of Trauma Surgery, Medical University Innsbruck, Anichstrasse 35, A-6020 Innsbruck, Austria Phone: +43-512-504 22413 Fax: +43-512-504 25743


Received: 2012-04-25

Accepted: 2012-05-10

Published Online: 2012-05-24

Published in Print: 2012-06-01


Citation Information: Biomedizinische Technik/Biomedical Engineering, Volume 57, Issue 3, Pages 149–155, ISSN (Online) 1862-278X, ISSN (Print) 0013-5585, DOI: https://doi.org/10.1515/bmt-2011-0107.

Export Citation

©2012 by Walter de Gruyter Berlin Boston.Get Permission

Citing Articles

Here you can find all Crossref-listed publications in which this article is cited. If you would like to receive automatic email messages as soon as this article is cited in other publications, simply activate the “Citation Alert” on the top of this page.

[1]
Jun Dong, Meng Lu, Baobao Liang, Xu Zhai, Jie Qin, and Xijing He