Problems of Comparing Blood Glucose Molality and Molarity Determined with an Omni, an EML 105 and an Ebio Analyser

Rainer Haeckel / Petra Hänecke
Published Online: 2005-06-01 | DOI: https://doi.org/10.1515/CCLM.2003.145

Abstract

The comparability between glucose concentrations measured in various sample systems is still a matter of debate. Decision limits are usually determined in venous plasma and then converted to either blood or to the aqueous compartment (activity). The conversion factors recommended have not yet been generally accepted.

In the present study, glucose concentrations were determined in blood and plasma with an Ebio analyzer (molarity) and in the aqueous compartment with both an EML 105 and an Omni (molality). All analytical results were referred to the same aqueous standard solution. The Ebio results agreed with reference method values in control materials. Concentrations determined in the various sample systems from patients (molarity) correlated well with the molality values measured either with the EML or the Omni. However, the mean values of the EML were not consistent with those derived theoretically by considering the different water content. With the Omni, only molality values in whole blood appeared plausible, but not in plasma, although the two sample systems should provide identical molality values. The EML results were almost identical in whole blood and plasma.

Theoretically, glucose molality would be the ideal diagnostic quantity. However, no diagnostic advantage of molality determined in whole blood with the Omni vs. molarity values could be detected in a group of 40 non-diabetic and 27 diabetic subjects.

About the article

Published Online: 2005-06-01

Published in Print: 2003-07-21


Citation Information: Clinical Chemistry and Laboratory Medicine, Volume 41, Issue 7, Pages 950–957, ISSN (Print) 1434-6621, DOI: https://doi.org/10.1515/CCLM.2003.145.

Export Citation

Citing Articles

Here you can find all Crossref-listed publications in which this article is cited. If you would like to receive automatic email messages as soon as this article is cited in other publications, simply activate the “Citation Alert” on the top of this page.

[1]
Karen L. Gerber and Kathleen P. Freeman
Veterinary Clinical Pathology, 2016, Volume 45, Number 1, Page 10

Comments (0)

Please log in or register to comment.
Log in