Automated analysis of pleural fluid total and differential leukocyte counts with the Sysmex XE-2100

Robert de Jonge
  • Department of Clinical Chemistry, Erasmus MC, Rotterdam, The Netherlands
/ Rob Brouwer
  • Department of Clinical Chemistry, Erasmus MC, Rotterdam, The Netherlands
/ Marcel van Rijn
  • Department of Clinical Chemistry, Erasmus MC, Rotterdam, The Netherlands
/ Bernadette A.C. van Acker
  • Department of Clinical Chemistry, Erasmus MC, Rotterdam, The Netherlands
/ Henk J.A.M. Otten
  • Department of Pulmonary Medicine, Erasmus MC, Rotterdam, The Netherlands
/ Jan Lindemans
  • Department of Clinical Chemistry, Erasmus MC, Rotterdam, The Netherlands
Published Online: 2006-11-07 | DOI: https://doi.org/10.1515/CCLM.2006.242

Abstract

Background: Determination of leukocyte (WBC) counts in pleural fluid is routinely performed by microscopic examination. In this study, we evaluated the performance of automated (differential) WBC counting in comparison with manual counting.

Methods: Pleural fluid samples (n=45) were obtained from patients undergoing diagnostic thoracocentesis. The manual total WBC count was determined after Samson staining in a Fuchs-Rosenthal hemocytometer; microscopic differential counts were performed on May-Grünwald Giemsa-stained cytospin slides. The Sysmex XE-2100 hematology analyzer was used for automated (differential) WBC counting. The functional detection limit was determined by serial dilution of continuous ambulatory peritoneal dialysis (CAPD) fluid and replicate measurements of each dilution.

Results: The automated WBC count (×106/L) was highly correlated with that of the microscopic reference method (r2=0.95; WBC-analyzer=0.97×WBC-reference method+16; n=45). Good agreement was also observed for the absolute lymphocyte count (r2=0.92; WBC-analyzer=0.99×WBC-reference method+32; n=36), neutrophil count (r2=0.94; WBC-analyzer=0.91×WBC-reference method+6; n=35), and monocyte count (r2=0.73; WBC-analyzer=0.83×WBC-reference method+6; n=38). The functional detection limit for WBCs was calculated at 50×106/L (coefficient of variation 20%).

Conclusions: With some limitations, total and differential WBC counts in pleural fluid can be reliably determined using the Sysmex XE-2100 instrument.

Clin Chem Lab Med 2006;44:1367–71.

Keywords: cell counts; differential count; pleural fluid; Sysmex XE-2100

References

  • 1.

    Balfour-Lynn IM, Abrahamson E, Cohen G, Hartley J, King S, Parikh D, et al. BTS guidelines for the management of pleural infection in children. Thorax 2005; 60(Suppl 1):i1–21. [Crossref]

  • 2.

    Light RW, Erozan YS, Ball WC, Jr. Cells in pleural fluid. Their value in differential diagnosis. Arch Intern Med 1973; 132:854–60. [Crossref]

  • 3.

    Maskell NA, Butland RJ. BTS guidelines for the investigation of a unilateral pleural effusion in adults. Thorax 2003; 58(Suppl 2):ii8–17. [Crossref]

  • 4.

    Salinas M, Rosas J, Iborra J, Manero H, Pascual E. Comparison of manual and automated cell counts in EDTA preserved synovial fluids. Storage has little influence on the results. Ann Rheum Dis 1997; 56:622–6. [Crossref]

  • 5.

    Aune MW, Sandberg S. Automated counting of white and red blood cells in the cerebrospinal fluid. Clin Lab Haematol 2000; 22:203–10. [Crossref]

  • 6.

    Aulesa C, Mainar I, Prieto M, Cobos N, Galimany R. Use of the Advia 120 hematology analyzer in the differential cytologic analysis of biological fluids (cerebrospinal, peritoneal, pleural, pericardial, synovial, and others). Lab Hematol 2003; 9:214–24.

  • 7.

    de Jonge R, Brouwer R, Smit M, de Frankrijker-Merkestijn M, Dolhain RJ, Hazes JM, et al. Automated counting of white blood cells in synovial fluid. Rheumatology (Oxford) 2004; 43:170–3. [Crossref]

  • 8.

    Ziebig R, Lun A, Sinha P. Leukocyte counts in cerebrospinal fluid with the automated hematology analyzer CellDyn 3500 and the urine flow cytometer UF-100. Clin Chem 2000; 46:242–7.

  • 9.

    Hoffmann JJ, Janssen WC. Automated counting of cells in cerebrospinal fluid using the CellDyn-4000 haematology analyser. Clin Chem Lab Med 2002; 40:1168–73.

  • 10.

    Aune MW, Becker JL, Brugnara C, Canfield W, Dorfman DM, Fiehn W, et al. Automated flow cytometric analysis of blood cells in cerebrospinal fluid: analytic performance. Am J Clin Pathol 2004; 121:690–700. [Crossref]

  • 11.

    Kresie L, Benavides D, Bollinger P, Walters J, Pierson D, Richmond T, et al. Performance evaluation of the application of body fluids on the Sysmex XE-2100 series automated hematology analyzer. Lab Hematol 2005; 11:24–30. [Crossref]

  • 12.

    Angeloni S, Nicolini G, Merli M, Nicolao F, Pinto G, Aronne T, et al. Validation of automated blood cell counter for the determination of polymorphonuclear cell count in the ascitic fluid of cirrhotic patients with or without spontaneous bacterial peritonitis. Am J Gastroenterol 2003; 98:1844–8. [Crossref]

  • 13.

    Conner BD, Lee YC, Branca P, Rogers JT, Rodriguez RM, Light RW. Variations in pleural fluid WBC count and differential counts with different sample containers and different methods. Chest 2003; 123:1181–7. [Crossref]

  • 14.

    Wyler D, Marty W, Bar W. Correlation between the postmortem cell content of cerebrospinal fluid and time of death. Int J Legal Med 1994; 106:194–9. [Crossref]

  • 15.

    Spencer CA, Takeuchi M, Kazarosyan M, MacKenzie F, Beckett GJ, Wilkinson E. Interlaboratory/intermethod differences in functional sensitivity of immunometric assays of thyrotropin (TSH) and impact on reliability of measurement of subnormal concentrations of TSH. Clin Chem 1995; 41:367–74.

About the article

Corresponding author: Dr. Robert de Jonge, Department of Clinical Chemistry (L-137), Erasmus MC, POB 2040, 3000 CA Rotterdam, The Netherlands


Received: 2006-05-11

Accepted: 2006-08-29

Published Online: 2006-11-07

Published in Print: 2006-11-01



Citation Information: Clinical Chemistry and Laboratory Medicine (CCLM), ISSN (Online) 1437-4331, ISSN (Print) 1434-6621, DOI: https://doi.org/10.1515/CCLM.2006.242. Export Citation

Citing Articles

Here you can find all Crossref-listed publications in which this article is cited. If you would like to receive automatic email messages as soon as this article is cited in other publications, simply activate the “Citation Alert” on the top of this page.

[2]
Y.-U. Cho, H.-S. Chi, S. H. Park, S. Jang, Y.-J. Kim, and C.-J. Park
International Journal of Laboratory Hematology, 2015, Volume 37, Number 3, Page 346
[3]
T. M. Lehto, P. Leskinen, P. Hedberg, and T. E. Vaskivuo
International Journal of Laboratory Hematology, 2014, Volume 36, Number 2, Page 114
[5]
Sabrina Buoro, Rosanna Gustinetti, Paola Dominoni, Maria Grazia Alessio, Patrizia Filisetti, Maria Grazia Lucà, Stefano Fagiuoli, Cosimo Ottomano, and Giuseppe Lippi
Clinical Biochemistry, 2012, Volume 45, Number 15, Page 1263
[6]
G. Ho, A. Keutgens, A. Huwart, F. Tassin, and A. Gothot
Immuno-analyse & Biologie Spécialisée, 2011, Volume 26, Number 5-6, Page 256
[7]
Michel Henry Mendler, Arun Agarwal, Mateen Trimzi, Erika Madrigal, Matthew Tsushima, Elliot Joo, Marisela Santiago, Elissa Flores, Gabriel David, Adrienne Workman, and Bruce Runyon
Journal of Hepatology, 2010, Volume 53, Number 3, Page 477
[8]
Robert de Jonge, Rob Brouwer, Marieke T. de Graaf, Ronald L. Luitwieler, Cherina Fleming, Magda de Frankrijker-Merkestijn, Peter A.E. Sillevis Smitt, Joke G. Boonstra, and Jan Lindemans
Clinical Chemistry and Laboratory Medicine, 2010, Volume 48, Number 5
[9]
A. PARIS, T. NHAN, E. CORNET, J.-P. PEROL, M. MALET, and X. TROUSSARD
International Journal of Laboratory Hematology, 2010, Volume 32, Number 5, Page 539
[10]
Sook Won Ryu and In Bum Suh
The Korean Journal of Laboratory Medicine, 2007, Volume 27, Number 6, Page 394