Volume 44, Issue 7 (Jul 2006)

Issues

Comparability of point-of-care whole-blood electrolyte and substrate testing using a Stat Profile Critical Care Xpress analyzer and standard laboratory methods

Zlata Flegar-Meštrić / Sonja Perkov
Published Online: 2011-09-21 | DOI: https://doi.org/10.1515/CCLM.2006.148

Abstract

Background: Rapid technological progress in point-of-care testing allows the measurement of multiple analytes in whole-blood samples. The present study evaluated biosensor-based methods for the measurement of electrolytes and substrates in whole blood using a Stat Profile Critical Care Xpress (Nova Biomedical, Waltham, MA, USA) multiprofile analyzer and their comparability with standard laboratory methods. Because of the increased utilization of arterial blood samples in hospitalized patients and limited information on differences between arterial and venous blood for most routine laboratory tests, analytical differences caused by different sample types were evaluated.

Methods: Whole-blood arterial samples and venous serum samples were obtained from 70 random patients with a variety of diagnoses admitted to the intensive care unit. The Stat Profile Critical Care Xpress analyzer was used to obtain whole-blood electrolyte and substrate profiles. For comparison studies, plasma or serum samples were analyzed according to standard laboratory methods using an Olympus AU 600 analyzer (Olympus Mishima, Shizuoka, Japan).

Results: Imprecision, expressed as the coefficient of variation (CV%), was less than 5.7% for all analytes at both high and low concentrations, except for creatinine, with a CV of 13.8% for low and 9.5% for high concentrations. The inaccuracy of electrolyte and substrate measurements in whole blood using a Stat Profile Critical Care Xpress analyzer met the analytical quality specification required for near patient testing, with observed bias within the range −4.5% to 5.3%. Statistically significant correlation (p<0.05) was obtained between standard laboratory methods performed on arterial plasma or venous serum samples on an Olympus AU 600 analyzer and direct whole-blood measurements on the Stat Profile Critical Care Xpress point-of-care analyzer for all parameters tested, although slope and intercept values showed analytical differences for electrolyte measurement.

Conclusions: The Stat Profile Critical Care Xpress multiprofile point-of-care analyzer provides rapid and accurate direct whole-blood measurement with acceptable performance compared to standard laboratory methods. The results obtained for electrolytes and substrates in whole blood were comparable to those for standard laboratory methods using arterial plasma or venous serum samples.

Clin Chem Lab Med 2006;44:898–903.

Keywords: analytical quality specification; biosensor-based analysis; point-of-care testing

References

  • 1.

    Kost GJ. New whole blood analyzers and their impact on cardiac and critical care. Crit Rev Clin Lab Sci 1993; 30:153–202.CrossrefGoogle Scholar

  • 2.

    Kost GJ, Vu H-T, Inn M, DuPlantier R, Fleisher M, Kroll MH, et al. Multicenter study of whole-blood creatinine, total carbon dioxide content, and chemistry profiling for laboratory and point-of-care testing in critical care in the United States. Crit Care Med 2000; 28,7:2379–89.CrossrefGoogle Scholar

  • 3.

    Fraser CG. Optimum analytical performance for point-of-care testing. Clin Chim Acta 2001; 307:37–43.Google Scholar

  • 4.

    Bullock DG. Quality control and quality assurance in point-of-care testing. In: Price CP, St John A, Hicks JM, editors. Point-of-care testing, 2nd ed. Washington, DC: AACC Press, 2004:137–45.Google Scholar

  • 5.

    Hyltoft Petersen P, Fraser CG, Kallner A, Kenny D. Strategies to set global analytical quality specifications in laboratory medicine. Scand J Clin Lab Invest 1999; 59:475–585.Google Scholar

  • 6.

    Nova Biomedical. Stat Profile® Critical Care Express. Instructions for user manual. Waltham, MA: Nova Biomedical, 2004:B5–7.Google Scholar

  • 7.

    ISO 9001:2000. Quality management system – requirements. Geneva: International Organization for Standardization, 2000.Google Scholar

  • 8.

    Passing H, Bablok W. A new biometrical procedure for testing the quality of measurement from two different analytical methods. Part I. J Clin Chem Clin Biochem 1983; 21:709–20.Google Scholar

  • 9.

    Ricoś C, Alvarez V, Cava F, García-Lario JV, Hernández A, Jiménez CV, et al. Biological variation database, 2001 update. www.westgard.com/guest17.htm.Google Scholar

  • 10.

    Briedigkeit L, Müller-Plathe O, Schlebusch H, Ziems J. Recommendations of the German Working Group Medical Laboratory Testing (AML) on the introduction and quality assurance of procedures for point-of-care testing (POCT) in hospitals. Clin Chem Lab Med 1999; 37:919–25.CrossrefGoogle Scholar

  • 11.

    Kampa IS, Keffer P. The use of a whole-blood benchtop analyzer (Nova 16) in a cardiac stat intensive care unit. Clin Chem 1998; 44:884–5.Google Scholar

  • 12.

    National Committee for Clinical Laboratory Standards. Method comparison and bias estimation using patient samples; approved guideline EP9-A. Wayne, PA: NCCLS, 1995.Google Scholar

  • 13.

    Burnett RW, Covington AK, Fogh-Andersen N, Külpmann WR, Levenstam A, Maas AH, et al. Recommendations for measurement of and conventions for reporting sodium and potassium by ion-selective electrodes in undiluted serum, plasma or whole blood. Clin Chem Lab Med 2000; 38:1065–71.CrossrefGoogle Scholar

  • 14.

    Burnett RW, Covington AK, Fogh-Andersen N, Külpmann WR, Levenstam A, Maas AH, et al. Use of ion-selective electrodes for blood-electrolyte analysis. Recommendation for nomenclature, definitions and conventions. Clin Chem Lab Med 2000; 38:363–70.CrossrefGoogle Scholar

  • 15.

    Lolekha PH, Vanavanan S, Teerakarnjana N, Chaichanajarernkul U. Reference ranges of electrolyte and anion gap on the Beckman E4A, Beckman Synchron CX5, Nova CRT, and Nova Stat Profile Ultra. Clin Chim Acta 2001; 307:87–93.Google Scholar

  • 16.

    Jagarinec N, Flegar-Meštrić Z, Šurina B, Vrhovski-Hebrang D, Preden-Kereković V. Pediatric reference intervals for 34 biochemical analytes in urban schoolchildren and adolescents. Clin Chem Lab Med 1998; 36:327–37.Google Scholar

  • 17.

    Sasse E, Bartell C, Durl M, Patsches M. Chemistry reference intervals for serum and plasma for the Beckman Synchron CX7 system utilizing the NCCLS C28-P guideline. Clin Chem 1995; 41:S199.Google Scholar

  • 18.

    Radiometer Medical. The blood gas handbook. Brønshøj: Radiometer Medical A/S, 2000.Google Scholar

  • 19.

    Steffes MW. Measurement of circulating glucose concentrations: the time is now for consistency among methods and types of samples. Clin Chem 2005; 51:1569–70.CrossrefGoogle Scholar

  • 20.

    Fogh-Anderson N, D'Orazio P. Proposal for standardizing direct-reading biosensors for blood glucose. Clin Chem 1998; 44:655–9.Google Scholar

  • 21.

    D'Orazio P, Burnett RW, Fogh-Andersen N, Jacobs E, Kuwa K, Külpmann WR, et al. Approved IFCC recommendation on reporting results for blood glucose (abbreviated). Clin Chem 2005; 51:1573–6.CrossrefGoogle Scholar

  • 22.

    Sacks DB, Bruns DE, Goldstein DE, Maclaren NK, McDonald JM, Parrott M. Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin Chem 2002; 48:436–72.Web of ScienceGoogle Scholar

  • 23.

    Doumas BT, Hause LL, Simuncak DM, Breitenfeld D. Differences between values for plasma and serum in tests performed in the Ektachem 700 XR analyzer, and evaluation of “plasma separator tube (PTS)”. Clin Chem 1989; 35:151–3.Google Scholar

  • 24.

    Ciuti R, Rinaldi G. Serum and plasma compared for use in 19 common chemical tests performed in the Hitachi 737 analyzer [letter]. Clin Chem 1989; 35:1562–3.Google Scholar

About the article

Corresponding author: Zlata Flegar-Meštrić, I. Zajca 19, 10 000 Zagreb, Croatia Phone/Fax: +385-1-2431397,


Received: 2005-12-06

Accepted: 2006-03-22

Published Online: 2011-09-21

Published in Print: 2006-07-01


Citation Information: Clinical Chemistry and Laboratory Medicine (CCLM), ISSN (Online) 1437-4331, ISSN (Print) 1434-6621, DOI: https://doi.org/10.1515/CCLM.2006.148.

Export Citation

©2006 by Walter de Gruyter Berlin New York. Copyright Clearance Center

Citing Articles

Here you can find all Crossref-listed publications in which this article is cited. If you would like to receive automatic email messages as soon as this article is cited in other publications, simply activate the “Citation Alert” on the top of this page.

[1]
Mauria A. O'Brien, Maureen A. McMichael, Kevin Le Boedec, and George Lees
Journal of Veterinary Emergency and Critical Care, 2014, Volume 24, Number 3, Page 291
[2]
N. I. Stacy, C. J. Innis, and J. A. Hernandez
Conservation Physiology, 2013, Volume 1, Number 1, Page cot003
[3]
Hyo-Sik Kim, Hyoungchun Park, Sungpil Chung, and Jeong-Ho Kim
Laboratory Medicine Online, 2011, Volume 1, Number 4, Page 179
[4]
Yeong Uk Jang, Su Nam Kim, Hye Jung Cho, Yong Han Sun, So Yeon Shim, Dong Woo Son, and Pil Whan Park
Journal of the Korean Society of Neonatology, 2011, Volume 18, Number 2, Page 301
[5]
Mahathi R. Adla, Julio A. Gonzalez-Paoli, and Stephen I. Rifkin
Southern Medical Journal, 2009, Volume 102, Number 8, Page 867
[6]
Paolo Calzavacca, Augustine Tee, Elisa Licari, and Rinaldo Bellomo
Renal Failure, 2010, Volume 32, Number 1, Page 27
[7]
Hyunwoong Park, Dae-Hyun Ko, Jin Q Kim, and Sang Hoon Song
The Korean Journal of Laboratory Medicine, 2009, Volume 29, Number 5, Page 430
[8]
Anunaya Jain, Imron Subhan, and Mahesh Joshi
International Journal of Emergency Medicine, 2009, Volume 2, Number 2, Page 117

Comments (0)

Please log in or register to comment.
Log in