Experience with post-market surveillance of in-vitro diagnostic medical devices for lay use in Germany

  • Rüdiger Siekmeier and Johannes Lütz

Abstract

Background: The European Directive 98/79/EC on in-vitro diagnostic medical devices (IVDs) regulates IVD marketing practices and post-market surveillance. IVD manufacturers have to inform the responsible Competent Authorities of any issues. In Germany, the Federal Institute for Drugs and Medical Devices (BfArM) is responsible for most IVDs, with a small subset of IVDs being within the responsibility of the Paul-Ehrlich-Institute (PEI).

Methods: All IVD notifications received by BfArM between 1999 and June 2006 were analysed in terms of the source of notification, underlying product defects and corrective actions.

Results: A total of 773 notifications were received, 566 related to IVDs for professional use and 207 related to over-the-counter (OTC) products for lay use. The latter included systems for blood glucose testing (analysers, tests and control materials; n=166) or coagulation testing (n=13) and pregnancy tests (n=25). Most reports came from manufacturers (n=115; 55.6%) and users (n=72; 34.8%) mainly via pharmacies and the Drug Commission of the German Pharmaceutical Association. Manufacturer investigations for all lay IVD cases reported revealed underlying product defects in 53 cases (25.6%). Product failure was excluded in 80 cases (38.6%), which included a large number of user errors (n=34). Many cases (n=74, 35.7%) could not be clarified because the test strips and/or analysers were not returned to the manufacturer for further investigation. In most cases, product defects identified by manufacturer investigations were related to the test strips and not to the analysers. Because of the high proportion of cases without proven product failure, corrective actions were performed only in a subset (n=54, 26.1%) of the cases reported for IVDs specified for lay use.

Conclusions: The results show that the governmental system for post-marketing surveillance of IVDs is an established tool to ensure product safety. The proportion of notifications for OTC products indicates that they should be the focus for action by the competent surveillance authorities.

Clin Chem Lab Med 2007;45:396–401.

:

Corresponding author: Priv.-Doz. Dr. R. Siekmeier, Bundesinstitut für Arzneimittel und Medizinprodukte (BfArM), Kurt-Georg-Kiesinger-Allee 3, 53119 Bonn, Germany

References

1. Directive 98/79/EC of the European Parliament and of the Council of 27 October 1998 on in vitro diagnostic medical devices. OJ L 331, 1998:1–37. http://europa.eu.int/eur-lex/en/index.html.Search in Google Scholar

2. Medizinproduktegesetz (MPG) in der Fassung der Bekanntmachung vom 7. August 2002. BGBl. I S. 3146. www.bfarm.de/de/Medizinprodukte/mp_recht/index.php.Search in Google Scholar

3. Medizinprodukte-Sicherheitsplanverordnung (MPSV) vom 24. Juni 2002. BGBl. I S. 2131. www.bfarm.de/de/Medizin produkte/mp_recht/index.php.Search in Google Scholar

4. Guidelines on a medical devices vigilance system. MEDDEV 2.12-1 rev. 4. http://europa.eu.int/comm/enterprise/medical_devices/guidelinesmed/baseguidelines.htm.Search in Google Scholar

5. Medical Devices Post Market Surveillance: National Competent Authority Report Exchange Criteria. GHTF/SG2/N20R10:2002. www.ghtf.org.Search in Google Scholar

6. Die Anforderungen der Medizinprodukte-Sicherheitsplanverordnung für in-vitro-Diagnostika im Rahmen des Medizinproduktegesetzes. MedizinProdukteRecht 4, 2002:120–133. www.vdgh.de.Search in Google Scholar

7. Siekmeier R, Will HG. German experience with the European Vigilance System for in vitro diagnostic medical devices 1999–2002 [abstract]. Clin Chem Lab Med2003;41:A87.Search in Google Scholar

8. Siekmeier R, Will HG. Safety of lay use products – German experience with the European vigilance system for in vitro diagnostic medical devices [abstract]. Clin Chem Lab Med2004;42:A151.Search in Google Scholar

Published Online: 2007-03-22
Published in Print: 2007-03-01

©2007 by Walter de Gruyter Berlin New York

Downloaded on 28.3.2024 from https://www.degruyter.com/document/doi/10.1515/CCLM.2007.058/html
Scroll to top button