Extending laboratory automation to the wards: effect of an innovative pneumatic tube system on diagnostic samples and transport time

Juliane Suchsland
  • Institute of Clinical Chemistry and Laboratory Medicine, University Medicine Greifswald, Greifswald, Germany
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Theresa Winter
  • Institute of Clinical Chemistry and Laboratory Medicine, University Medicine Greifswald, Greifswald, Germany
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Anne Greiser
  • Institute of Clinical Chemistry and Laboratory Medicine, University Medicine Greifswald, Greifswald, Germany
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Thomas Streichert / Benjamin Otto / Julia Mayerle / Sören Runge / Anders Kallner / Matthias Nauck
  • Institute of Clinical Chemistry and Laboratory Medicine, University Medicine Greifswald, Greifswald, Germany
  • DZHK (German Centre for Cardiovascular Research), Partner Site Greifswald, University Medicine, Greifswald, Germany
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Astrid Petersmann
  • Corresponding author
  • Institute of Clinical Chemistry and Laboratory Medicine, University Medicine Greifswald, Ferdinand-Sauerbruch-Strasse, 17475 Greifswald, Germany
  • Email
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
Published Online: 2016-06-30 | DOI: https://doi.org/10.1515/cclm-2016-0380

Abstract

Background:

The innovative pneumatic tube system (iPTS) transports one sample at a time without the use of cartridges and allows rapid sending of samples directly into the bulk loader of a laboratory automation system (LAS). We investigated effects of the iPTS on samples and turn-around time (TAT).

Methods:

During transport, a mini data logger recorded the accelerations in three dimensions and reported them in arbitrary area under the curve (AUC) units. In addition representative quantities of clinical chemistry, hematology and coagulation were measured and compared in 20 blood sample pairs transported by iPTS and courier.

Results:

Samples transported by iPTS were brought to the laboratory (300 m) within 30 s without adverse effects on the samples. The information retrieved from the data logger showed a median AUC of 7 and 310 arbitrary units for courier and iPTS transport, respectively. This is considerably below the reported limit for noticeable hemolysis of 500 arbitrary units.

Conclusions:

iPTS reduces TAT by reducing the hands-on time and a fast transport. No differences in the measurement results were found for any of the investigated 36 analytes between courier and iPTS transport. Based on these findings the iPTS was cleared for clinical use in our hospital.

Keywords: laboratory automation; pneumatic tube system; pre-analytical phase; sample transport

References

  • 1.

    Armbruster DA, Overcash DR, Reyes J. Clinical chemistry laboratory automation in the 21st century – Amat Victoria curam (Victory loves careful preparation). Clin Biochem Rev 2014;35:143–53.Google Scholar

  • 2.

    Fernandes CM, Worster A, Eva K, Hill S, McCallum C. Pneumatic tube delivery system for blood samples reduces turnaround times without affecting sample quality. J Emerg Nurs 2006;32:139–43.Google Scholar

  • 3.

    Lippi G, Plebani M, Favaloro EJ. Technological advances in the hemostasis laboratory. Semin Thromb Hemost 2014;40:178–85.Google Scholar

  • 4.

    Al-Riyami AZ, Al-Khabori M, Al-Hadhrami RM, Al-Azwani IS, Davis HM, Al-Farsi KS, et al. The pneumatic tube system does not affect complete blood count results; a validation study at a tertiary care hospital. Int J Lab Hematol 2014;36:514–20.Google Scholar

  • 5.

    Gossez M, Poitevin-Later F, Demaret J, Jallades L, Venet F, Malcus C, et al. Effect of pneumatic tube transport on T lymphocyte subsets analysis. Cytometry B Clin Cytom 2015;88:371–4.Google Scholar

  • 6.

    Pragay DA, Edwards L, Toppin M, Palmer RR, Chilcote ME. Evaluation of an improved pneumatic-tube system suitable for transportation of blood specimens. Clin Chem 1974;20:57–60.Google Scholar

  • 7.

    Ellis G. An episode of increased hemolysis due to a defective pneumatic air tube delivery system. Clin Biochem 2009;42:1265–9.Google Scholar

  • 8.

    Strubi-Vuillaume I, Carlier V, Obeuf C, Vasseur F, Maury J-C, Maboudou P, et al. Gentle blood aspiration and tube cushioning reduce pneumatic tube system interference in lactate dehydrogenase assays. Ann Clin Biochem 2016;53(Pt 2):295–7.Google Scholar

  • 9.

    Medical laboratories – Requirements for quality and competence (ISO 15189:2012, Corrected version 2014-08-15); German version EN ISO 15189:2012.Google Scholar

  • 10.

    Revision of the “Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations – Rili-BAEK” (unauthorized translation). J Lab Med 2015;39:26–69.Google Scholar

  • 11.

    Streichert T, Otto B, Schnabel C, Nordholt G, Haddad M, Maric M, et al. Determination of hemolysis thresholds by the use of data loggers in pneumatic tube systems. Clin Chem 2011;57:1390–7.Google Scholar

  • 12.

    Mensel B, Wenzel U, Roser M, Ludemann J, Nauck M. Considerably reduced centrifugation time without increased hemolysis: evaluation of the new BD Vacutainer SSTTMII Advance. Clin Chem 2007;53:794–5.Google Scholar

  • 13.

    Goswami B, Singh B, Chawla R, Gupta VK, Mallika V. Turn around time (TAT) as a benchmark of laboratory performance. Indian J Clin Biochem 2010;25:376–9.Google Scholar

  • 14.

    Wenham PR, Hanson T, Ashby JP. Interference in spectrophotometric analysis of cerebrospinal fluid by haemolysis induced by transport through a pneumatic tube system. Ann Clin Biochem 2001;38(Pt 4):371–5.Google Scholar

  • 15.

    Kratz A, Salem RO, van Cott EM. Effects of a pneumatic tube system on routine and novel hematology and coagulation parameters in healthy volunteers. Arch Pathol Lab Med 2007;131:293–6.Google Scholar

  • 16.

    Evliyaoglu O, Toprak G, Tekin A, Basarali MK, Kilinc C, Colpan L. Effect of pneumatic tube delivery system rate and distance on hemolysis of blood specimens. J Clin Lab Anal 2012;26:66–9.Google Scholar

  • 17.

    Hansen Rohrpost. Available at: URL:http://www.hansen-rohrpost.de/rohrmaterial/pvc-rohrpostfahrrohr/. Accessed: 20 Mar 2015.

About the article

Corresponding author: Dr. med. Dipl. Biol. Astrid Petersmann, Institute of Clinical Chemistry and Laboratory Medicine, University Medicine Greifswald, Ferdinand-Sauerbruch-Str., 17475 Greifswald, Germany, Phone: +49-3834-86-5670

aPresent address: Eppendorf AG, Hamburg, Germany.


Received: 2016-05-01

Accepted: 2016-05-30

Published Online: 2016-06-30

Published in Print: 2017-02-01


Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.

Research funding: None declared.

Employment or leadership: None declared.

Honorarium: None declared.

Competing interests: The funding organization(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.


Citation Information: Clinical Chemistry and Laboratory Medicine (CCLM), Volume 55, Issue 2, Pages 225–230, ISSN (Online) 1437-4331, ISSN (Print) 1434-6621, DOI: https://doi.org/10.1515/cclm-2016-0380.

Export Citation

©2017 Walter de Gruyter GmbH, Berlin/Boston.Get Permission

Citing Articles

Here you can find all Crossref-listed publications in which this article is cited. If you would like to receive automatic email messages as soon as this article is cited in other publications, simply activate the “Citation Alert” on the top of this page.

[1]
Laura Heireman, Pieter Van Geel, Lorenz Musger, Evelien Heylen, Wim Uyttenbroeck, and Boris Mahieu
Clinical Biochemistry, 2017

Comments (0)

Please log in or register to comment.
Log in