Practical recommendations for managing hemolyzed samples in clinical chemistry testing

Giuseppe LippiORCID iD: http://orcid.org/0000-0001-9523-9054 / Janne Cadamuro / Alexander von Meyer
  • Institute for Laboratory Medicine, Kliniken Nordoberpfalz AG and Klinikum St. Marien, Weiden and Amberg, Germany
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Ana-Maria Simundic /
Published Online: 2018-01-26 | DOI: https://doi.org/10.1515/cclm-2017-1104

Abstract

We suggest here a pragmatic approach for managing results of clinical chemistry testing in hemolyzed samples collected from adults/older children, attempting to balance the need to produce quality laboratory data with clinical urgency of releasing test results. Automatic measurement of the hemolysis index (H-index) in serum or plasma is highly advisable, whilst low-quality assessment of this test remains less good than a visual inspection. Regarding its practical use, when the H-index value does not generate an analytically significant bias, results can be released, whilst when the value is associated with analyte variation in a range between analytically and clinically significant bias (i.e. variation does not exceed the reference change value [RCV]), results of hemolysis-sensitive tests can be released in association with a comment describing the direction in which data are potentially altered, suggesting the need to collect another sample. When the H-index is associated with analyte variation exceeding clinically significant bias (i.e. variation exceeds the RCV), results of hemolysis-sensitive tests should be suppressed and replaced with a comment that biased results cannot be released because the sample is preanalytically compromised and advising the recollection of another sample. If H-index values reach an even higher critical cut-off (i.e. H-index corresponding to a cell-free hemoglobin concentration ≥10 g/L), all laboratory data may be unreliable and should hence be suppressed and replaced with a comment that all data cannot be released because the sample is grossly hemolyzed, also suggesting the recollection of another sample. Due to inaccuracy and imprecision, the use of corrective formulas for adjusting data of hemolysis-sensitive tests is discouraged.

Keywords: hemolysis; interference; quality; safety

Introduction

The preanalytical phase is an essential part of the total testing process along with both the analytical and postanalytical phases [1]. Solid evidence has been provided that the vast majority of diagnostic errors emerge from the many and still largely manually-intensive activities related to collection, handling and transportation of biological samples [2]. Among the various preanalytical problems, poor sample quality remains central [3]. More specifically, spuriously hemolyzed specimens are still the prevailing sources of unsuitable samples in clinical laboratories worldwide, with a frequency that is 5- to 10-fold higher than that of other conditions needing suppression of test results [4]. This is mainly due to the fact that breakdown of red blood cells (RBCs) and subsequent release of hemoglobin and other intracellular components in serum or plasma cause a kaleidoscope of biological, chemical and spectrophotometric interferences, thus finally often making test results unreliable [5].

Understandably, the practice of test results suppression in unsuitable samples is associated with many clinical (i.e. delayed diagnosis), economic (i.e. incremental costs needed for recollecting blood samples) and organizational (i.e. litigation with clinicians) consequences [6], [7]. In the past, several proposals on how to manage unsuitable specimens have been published [8], [9]. Nevertheless, an agreed harmonized strategy for managing unsuitable samples across different laboratories worldwide, is still lacking. The previously mentioned evidence, combined with the dramatic lack of harmonized strategies, have catalyzed a passionate debate about the most appropriate and reliable strategy for dealing with preanalytically altered laboratory test results, especially of those obtained in spuriously hemolyzed samples [10], [11], [12], [13].

Bilirubin and lipids are also long since known as interfering substances, by altering the color or turbidity of the sample [14], [15], and causing interference due to spectrophotometric mechanisms, volume depletion effects, partitioning of the sample and through physicochemical mechanisms (lipemia), or through spectrophotometric mechanism and chemical interference of bilirubin (icterus) [5]. Although hemolysis mostly occurs during blood sampling and transportation, lipemia and icterus are associated with patient status and characteristics, so that they cannot be simply solved by repeated blood sampling. Moreover, due to different mechanisms through which interference occurs, hemolysis, icterus and lipemia cannot be looked at and managed as being the same phenomena. Interference from lipemia and icterus will thus not be discussed in this article, and their practical management in the laboratory will be analyzed in another specific document.

Therefore, the aim of this article is to provide recommendations for managing the results of clinical chemistry testing obtained in hemolyzed samples collected from adults/older children. Our proposal is a pragmatic approach, tentatively balancing the need to preserve quality of testing with the clinical urgency of releasing test results, especially when they are for patients with critical or life-threatening disorders. Moreover, the discussion on practical management of unsuitable samples in other areas of diagnostic testing (i.e. immunochemistry, coagulation, hematology, molecular biology and point-of-care testing) is outside of the scope of this article because the quality of published evidence about interference is still low beyond clinical chemistry. Moreover, type of samples, analytical techniques and clinical implications are considerably different.

Our proposal outlined in the rest of this document, includes recommendations on:

  1. how to systematically assess serum indices,

  2. how to define hemolysis index (H-index) cut-offs for flagging, alarming or suppressing tests results

  3. reporting flagged or alarming tests results

  4. suppressing hemolysis-sensitive test results

  5. suppressing all tests results

  6. correcting data for the H-index

  7. including H-index data in the laboratory report

1. Systematic assessment of the serum indices (H-index)

Recommendations:

  • 1.1

    Visual assessment and management of hemolysis level, based on individual, non-standardized subjective opinion, is strongly discouraged and should be replaced with automated detection and management systems.

  • 1.2

    Systematic automated measurement of the H-index in all serum and plasma samples and implementation of automated algorithm-based decision making rules is a strongly recommended approach for managing unsuitable specimens.

  • 1.3

    In facilities where systematic automated measurement is unavailable, visual assessment of hemolysis should be done by comparing coloration of the sample with a color chart (e.g. a photo with serum or plasma sample containing different cell-free hemoglobin values).

  • 1.4

    Laboratories where systematic automated measurement is unavailable are urged to implement automated H-index measurement for patient safety reasons.

The assessment of sample quality before testing is a mainstay of either routine or urgent laboratory diagnostics, as also endorsed by international organizations such as the International Organization for Standardization (ISO) 15189:2012 [16], [17], [18] and the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) [19], [20]. Although consensus has been reached that the importance of this activity is unquestionable, sample quality has for a long time been assessed by visual inspection of serum or plasma, with the aim of identifying hemolysis, turbidity or icterus. Nevertheless, several lines of evidence now attest that visual sample inspection has too many drawbacks (i.e. inaccuracy, imprecision, highly dependent upon the operator) [21], and may also jeopardize patient safety [22]. It has also been shown that not only visual detection, but also the decision on whether or not to report the test results from hemolyzed samples, may vary substantially between individuals when based on individual judgment [23], [24], [25]. To minimize errors associated with such practice, visual assessment of hemolysis level and management of hemolyzed samples based on individual, non-standardized subjective opinion is thus strongly discouraged, should no longer be performed and shall be replaced with automated detection and management systems [26]. In fact, automatic assessment of serum or plasma indices in preanalytical and clinical chemistry platforms not only provides a much more precise, accurate and reproducible estimation of potential interference, but is also easy and inexpensive. Moreover, automated systems do not affect the turnaround time and are the only means for identification of unsuitable samples and direct data transfer to the laboratory information system (LIS) for laboratories using total laboratory automation [27]. Therefore, systematic automated measurement of the H-index in all serum and plasma samples and implementation of automated algorithm-based decision making rules have meaningless impact on laboratory budgets and are strongly recommended for managing unsuitable specimens. In facilities where the H-index is unavailable, we still support visual assessment of hemolysis by comparing sample hue with a color chart in order to maintain some sort of standardization. However, we urge these laboratories to implement automated H-index measurement for patient safety reasons.

2. Defining H-index cut-offs for flagging, alarming or suppressing tests results

Recommendations:

  • 2.1

    Test-specific cut-offs of the H-index above which an analytically significant bias may occur can be either adopted from manufacturers’ assay sheets (when available), or can be locally calculated.

  • 2.2

    Analytically significant bias (i.e. analytically acceptable imprecision) for each test should be derived from available data in accordance with the consensus recommendations of the 1st Strategic Conference of the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM), and its proposed hierarchy for establishing analytical specifications.

  • 2.3

    If quality specifications can be estimated from biological variability data, those may be either obtained from biological variation databases or from the available literature.

  • 2.4

    Clinically significant bias should be expressed as a reference change value (RCV).

Most manufacturers include in their assay sheets a specific documentation about the potential interference of cell-free hemoglobin on clinical chemistry tests, also providing instrument-specific cut-offs above which tests results are thought to be analytically biased. Although these thresholds have not been calculated according to the concept of “clinically significant” bias, they are still valid for establishing when the analytical quality specifications are no longer met.

When test-specific cut-offs of the H-index above which an analytically significant bias may occur cannot be adopted from manufacturers’ assay sheets because they are either not available or are not applicable, a laboratory shall define its own cut-off for the analytically significant bias for each test. Specifications for significant analytical bias (i.e. analytically acceptable imprecision) should be derived from available data in accordance with the consensus recommendations of the 1st Strategic Conference of the EFLM, and its proposed hierarchy for establishing analytical specifications, which entails outcome studies, biological variation and state of the art [28], [29]. If quality specifications can be estimated from biological variability data, these may be either obtained from reliable sources, such as in the database developed by Carmen Ricos and her coworkers (and hosted on Westgard’s website) or from the European Biological Variation Study (EuBIVAS) database, as well as from other scientific publications [30], [31], [32].

We recommend that clinically significant bias is expressed as RCV. The RCV which is also known as critical difference, is an objective tool for assessing the significance of differences in serial laboratory data obtained from the same subject. Originally proposed by Harris and Yasaka [33] in 1983, the RCV was then largely endorsed by Fraser [34] in the following years. The RCV is based on a specific formula, incorporating both within-subject biological variation (CVI) and analytical variation (CVA). Terms and symbols used to define the components of biological variation and other related aspects were recently proposed by Simundic et al. and endorsed by the EFLM Working Group for Biological Variation (WG-BV) [35], [36], as shown in Figure 1A.

Calculation formula of the reference change value (A) and the analytical imprecision (B) of laboratory tests.
Figure 1:

Calculation formula of the reference change value (A) and the analytical imprecision (B) of laboratory tests.

The RCV has become commonplace and almost irreplaceable in laboratory medicine, when establishing whether or not a certain variation of a measurable analyte can be considered “significantly” different from either the homeostatic point or from a previous value [37]. The application of RCV is a straightforward concept for identifying H-index cut-offs above which hemolysis-sensitive tests are plagued by “clinically significant” bias. Notably, this approach was developed and validated more than 10 years ago by Vermeer et al. [38], and has since been employed in many other subsequent investigations [39], [40], [41], [42].

Importantly, whilst CVI is conventionally considered to be rather constant in healthy individuals, the CVA can vary widely using different analytical techniques, methods and reagents [34]. This is why each laboratory should preferably calculate and use its own RCV, on a local basis, with the same instrumentation and reagents used for measuring the H-index. Data collected with internal quality controls (IQA) or external quality assessment (EQA) programs can be reliable sources for this information. In these cases, as multiple levels of quality control materials are conventionally measured, the value of CVA used for the RCV can be calculated as the mean CVA obtained from the different levels of quality control materials tested (Figure 1B).

For laboratories seeking to perform local experiments for defining interference cut-offs, the whole procedure must be carried out with accurate procedures (e.g. calculating the bias in serum or plasma samples containing scalar amounts of cell-free hemoglobin), and results should then be stored until there is a change in reagents or instrumentation. The advantages and limitations of the various techniques for preparing hemolyzed specimens and performing interference studies have been previously discussed elsewhere [43], [44].

3. Reporting flagged or alarming tests results

Recommendations:

  • 3.1

    Test results measured in samples where H-index values are associated with a bias ranging between analytically and clinically significant cut-offs should be reported to the requesting clinicians.

  • 3.2

    If test result is reported from a hemolyzed sample (irrespective of being within the reference range or not), the laboratory report should be accompanied by a comment stating: “Value possibly decreased/increased by hemolysis. Consider recollecting another sample”.

  • 3.3

    If test result is reported from a hemolyzed sample (irrespective of being within reference range or not), a comment can be either placed immediately below the numerical value, or added as a note at the end of the laboratory report.

  • 3.4

    Comments should always provide a clear indication of the direction in which the test result is potentially biased (i.e. increased or decreased).

Test results measured in samples where H-index values are associated with a bias ranging between analytically and clinically significant cut-offs should be reported to the requesting clinicians, as (i) this result will not likely trigger inappropriate care, because such a variation is still clinically acceptable and (ii) timely communication of test results in which bias is not clinically relevant, may often be necessary for establishing the most rapid and appropriate therapeutic measures, especially in patients with critical or life-threatening conditions. Potassium is the most paradigmatic example. Failure to communicate hyper- or hypokalemia measured in mildly hemolyzed samples is probably worse than suppressing a non-clinically biased data, as this can lead to delayed treatment of a potentially serious disorder.

Therefore, whenever H-index values of the specimen are associated with a bias ranging between the analytically and clinically significant cut-offs, test results can be added to the laboratory report accompanied by a specific comment like “Value possibly decreased/increased by hemolysis. Consider recollecting another sample”. The comment can be either placed immediately below the numerical value, or added as a note at the end of the laboratory report. Both solutions seem suitable. In those cases where comments on laboratory reports tend to be ignored, missed or overlooked, it might be safer to replace test results with a specific sentence (e.g. “see comment”) and providing the value within the comment section along with information on the hemolysis level. The appropriate way of displaying test results and interference information has to be decided individually for each health care setting based on patient safety risk analysis. Moreover, it seems advisable to always describe the direction in which the test result is potentially biased (i.e. increased or decreased), as this will help clinicians to more appropriately interpret the data (Figure 2).

Summary of practical recommendations for managing hemolyzed specimens for clinical chemistry testing.
Figure 2:

Summary of practical recommendations for managing hemolyzed specimens for clinical chemistry testing.

4. Suppressing hemolysis-sensitive test results

Recommendations:

  • 4.1

    Results of hemolysis-sensitive tests measured in samples when H-index values are associated with a bias exceeding the clinically significant cut-offs calculated according to the test-specific RCV should be suppressed.

  • 4.2

    Instead of a result, a comment should be added to the test report, stating: “Hemolysis exceeding quality specifications of the test. Consider recollecting another sample”.

  • 4.3

    Test results measured on samples, for which H-index values are not associated with a bias exceeding the clinically significant cut-offs, should be reported to the requesting clinician (as already defined under 3.1–3.4).

Whenever the H-index values of the specimen are associated with a bias exceeding the clinically significant cut-offs calculated according to the test-specific RCV, the results of hemolysis-sensitive tests should be suppressed and replaced by a comment like “Hemolysis exceeding quality specifications of the test. Consider recollecting another sample” (Figure 2).

5. Suppressing all test results

Recommendations:

  • 5.1

    In samples with >10 g/L of cell-free hemoglobin, all results of clinical chemistry testing should be suppressed and another sample should be requested.

  • 5.2

    If test results from a grossly hemolyzed sample (>10 g/L of cell-free hemoglobin) are suppressed, the laboratory report should be accompanied by a comment stating: “Hemolyzed specimen. Consider recollecting another sample”.

The vast majority of studies have tested hemolysis interference up to 10 g/L of cell-free hemoglobin and no definitive evidence is available about the possible bias of clinical chemistry analytes above this threshold. Moreover, serum or plasma samples with such a high degree of hemolysis are not commonly observed in clinical laboratories [45]. Therefore, due to the lack of precise knowledge about the analytical and clinical consequences of bias in samples with higher degrees of hemolysis, it seems appropriate and more prudent to suggest that all results of clinical chemistry testing should be suppressed in these samples and another sample should be requested, using a comment like “Hemolyzed specimen. Consider recollecting another sample” (Figure 2). Notably, the injury of all blood cells and endothelia occurred during sample collection in serum or plasma grossly hemolyzed is probably so high that the vast majority of tests results will be thoroughly unreliable [46]. An overview of analytical and clinical significant cut-offs including the decisions on whether or not to release the test result with/without a comment are depictured in Figure 3.

Interferogram, showing analytical and clinical significant decision limits. RCV, reference change value.
Figure 3:

Interferogram, showing analytical and clinical significant decision limits.

RCV, reference change value.

6. Correcting data for the H-index

Recommendation:

  • 6.1

    Using corrective formulas for adjusting test results of hemolysis-sensitive tests from the serum or plasma concentration of cell-free hemoglobin is inaccurate and misleading, and is strongly discouraged.

Another approach for managing test results in hemolyzed samples has been proposed, entailing either correction of data of hemolysis-sensitive tests for the H-index or including comments about the possible bias estimated from the serum or plasma concentration of cell-free hemoglobin [47], [48], [49]. Nevertheless, independent studies showed that this practice should be highly discouraged as spurious hemolysis is not homogeneous and sufficiently predictable in different subjects (i.e. the biological source of interference has an extremely wide inter-individual variation) [50], [51], and is also strongly dependent on some biological characteristics of RBCs (i.e. volume, size heterogeneity, etc.) and mechanism of interference [52]. Therefore, using corrective formulas for adjusting test results of hemolysis-sensitive tests from serum or plasma concentration of cell-free hemoglobin is inaccurate and misleading, and we advise against routinely using this strategy.

7. Including H-index data in the laboratory report

Recommendations:

  • 7.1

    Degree of hemolysis should be converted from the instrument-specific arbitrary units into g/L of cell-free hemoglobin to improve harmonization.

  • 7.2

    Degree of hemolysis (expressed as g/L of cell-free hemoglobin) should be provided in the laboratory test report.

  • 7.3

    Clinical laboratories should use control materials specific for the H-indices for continuously monitoring the analytical performance of the H-index.

The new generation of clinical chemistry analyzers can now measure the H-index with a degree of accuracy and precision that is comparable to that of many clinical chemistry tests. Moreover, reliable evidence suggests that the H-index is highly correlated with the serum or plasma concentration of cell-free hemoglobin measured with the reference cyanmethemoglobin technique [53], [54]. Therefore, inclusion of H-index data in the laboratory report, preceded by translation of instrument-specific arbitrary units into g/L of hemoglobin, seems advisable for many reasons, such as (i) the precision and accuracy of this measure is satisfactory, so that it can be safely reported to clinicians; (ii) the provision of an objective measure of sample quality will provide an objective basis for supporting the decision to suppress results of hemolysis-sensitive tests; (iii) spurious hemolysis is also an index of phlebotomy quality [55], so that systematic reporting H-index data may help improving phlebotomy practices, highlighting those healthcare settings where sample hemolysis is higher compared to others; (iv) the use of H-index will ease participating to quality programs based on quality indicators such as that of the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) Working Group “Laboratory Error and Patient Safety” [56], [57]; (v) in rare cases of patients with in vivo hemolysis (i.e. hemolytic anemia), providing an estimation of cell-free hemoglobin in serum or plasma may be valuable information for clinicians.

Although the availability of quality control materials specific for the H-indices remains limited, many diagnostic companies have these products in their pipeline. Once these quality control materials become widely available in the diagnostic market, clinical laboratories should use them for continuously monitoring the analytical performance of the H-index likewise in the routine practice of clinical chemistry testing. In the meanwhile, the same procedures applied for other quality controls can be applied to monitoring the analytical performance of this test. Therefore, routine serum or plasma samples with different levels of cell-free hemoglobin values may be aliquoted and stored, to be used as IQAs, whilst failure of quality control measures should prevent using the test for undertaking clinically significant decisions.

Conclusions

Sample quality is still considered a cornerstone of total quality in laboratory diagnostics and its assessment should be seen as an almost unavoidable routine activity. Yet, some areas of uncertainty remain. For example, the CVI calculated in critical patients is very different from that estimated in ostensibly healthy subjects or in outpatients [34]. Therefore, too narrow limits of the RCV will lead to not releasing clinically useful data in patients with life-threatening conditions (e.g. results of coagulation testing in patients with liver disease) [58]. This important issue is not actually covered by the available data on CVI, but will probably be targeted by the “Biological variation database” prepared and maintained by the Task and Finish Group Biological Variation Database (TFG-BVD) and endorsed by the EFLM. The database will be hosted on the EFLM website, and will contain essential information about the biological variation and derived performance specifications for different measurands along with the evidence that supports it [59]. Then, provision of the H-index by different manufacturers is still plagued by poor harmonization, especially that referring to measurement procedure, variable definition of thresholds of interference after which test results may be “analytically” biased, and lack of information describing how interference cut-offs have been calculated.

The development and use of the H-index has introduced a paradigm shift in laboratory practice. The many advantages compared to visual inspection make its use highly advisable in all clinical laboratories. Irrespective of this virtually undisputable evidence, there is still fierce debate about the most effective and clinically safe strategy for dealing with sample hemolysis. In the authors’ expectations, the pragmatic approach suggested in this document (Figure 1; Table 1) should provide an acceptable compromise between quality and clinical needs. Our approach is in favor of reporting hemolysis-sensitive test results in cases when the bias is clinically acceptable, so providing fundamental clinical information to clinicians. On the other hand, in cases when the predicted bias is clinically significant and may thus jeopardize patient safety, test results suppression seems the most reliable approach. A useful example of this strategy, regarding practical management of potassium test results in a hemolyzed sample, is shown in Figure 4 [60]. Notably, this approach can be easily incorporated into automatic algorithms, thus saving operator time, preventing the risk of entering erroneous data in the LIS and promoting harmonization in the management of clinical chemistry testing performed using potentially unsuitable specimens.

Table 1:

Practical recommendations for managing test results in hemolyzed specimens.

Application of the suggested strategy for managing potassium results obtained in hemolyzed samples. *Cut-off values are fictional and have to be calculated for each individual setting of instrument and assay in use.
Figure 4:

Application of the suggested strategy for managing potassium results obtained in hemolyzed samples.

*Cut-off values are fictional and have to be calculated for each individual setting of instrument and assay in use.

References

  • 1.

    Lippi G, Plebani M, Graber ML. Building a bridge to safe diagnosis in health care. The role of the clinical laboratory. Clin Chem Lab Med 2016;54:1–3. CrossrefPubMedGoogle Scholar

  • 2.

    Lippi G, Baird GS, Banfi G, Bölenius K, Cadamuro J, Church S, et al. Improving quality in the preanalytical phase through innovation, on behalf of the European Federation for Clinical Chemistry and Laboratory Medicine (EFLM) Working Group for Preanalytical Phase (WG-PRE). Clin Chem Lab Med 2017;55:489–500. Google Scholar

  • 3.

    Dikmen ZG, Pinar A, Akbiyik F. Specimen rejection in laboratory medicine: necessary for patient safety? Biochem Med (Zagreb) 2015;25:377–85. PubMedGoogle Scholar

  • 4.

    Lippi, G, Cervellin G, Favaloro EJ, Plebani M. In vitro and in vivo hemolysis. An unresolved dispute in laboratory medicine. Berlin, Germany: Walter de Gruyter GmbH, 2012. Google Scholar

  • 5.

    Simundic AM, Nikolac N, Guder WG. Preanalytical variation and preexamination processes. In: Rifai N, Horvath R, Wittwer C, editors. Tietz textbook of clinical chemistry and molecular diagnostics, 6th ed. Amsterdam, The Netherlands: Elsevier, 2018:81–120. Google Scholar

  • 6.

    Lippi G, Plebani M, Di Somma S, Cervellin G. Hemolyzed specimens: a major challenge for emergency departments and clinical laboratories. Crit Rev Clin Lab Sci 2011;48:143–53. CrossrefGoogle Scholar

  • 7.

    Cadamuro J, Fiedler GM, Mrazek C, Felder TK, Oberkofler H, Kipman U, et al. In-vitro hemolysis and its financial impact using different blood collection systems. J Lab Med 2016;40:49–55. Google Scholar

  • 8.

    Badrick T, Barden H, Callen S, Dimeski G, Gay S, Graham P, et al. Consensus statement for the management and reporting of haemolysed specimens. Clin Biochem Rev 2016;37:140–2. PubMedGoogle Scholar

  • 9.

    Lippi G, Banfi G, Buttarello M, Ceriotti F, Daves M, Dolci A, et al. Recommendations for detection and management of unsuitable samples in clinical laboratories. Clin Chem Lab Med 2007;45:728–36. PubMedGoogle Scholar

  • 10.

    Simundic AM, Topic E, Nikolac N, Lippi G. Hemolysis detection and management of hemolysed specimens. Biochem Med (Zagreb) 2010;20:154–9. Google Scholar

  • 11.

    Cadamuro J, Mrazek C, Haschke-Becher E, Sandberg S. To report or not to report: a proposal on how to deal with altered test results in hemolytic samples. Clin Chem Lab Med 2017;55:1109–11. PubMedGoogle Scholar

  • 12.

    Lippi G, Cervellin G, Plebani M. Reporting altered test results in hemolyzed samples: is the cure worse than the disease? Clin Chem Lab Med 2017;55:1112–4. Google Scholar

  • 13.

    Cadamuro J, Simundic AM, Ajzner E, Sandberg S. A pragmatic approach to sample acceptance and rejection. Clin Biochem 2017;50:579–81. CrossrefPubMedGoogle Scholar

  • 14.

    Ryder KW, Glick MR. Erroneous laboratory results from hemolyzed, icteric, and lipemic specimens. Clin Chem 1993;39:175–6. PubMedGoogle Scholar

  • 15.

    Nikolac N. Lipemia: causes, interference mechanisms, detection and management. Biochem Med (Zagreb) 2014;24:57–67. PubMedGoogle Scholar

  • 16.

    International Organization for Standardization. ISO15189 Medical laboratories – Requirements for quality and competence. Geneva: ISO, 2012. Google Scholar

  • 17.

    Antonelli G, Padoan A, Aita A, Sciacovelli L, Plebani M. Verification of examination procedures in clinical laboratory for imprecision, trueness and diagnostic accuracy according to ISO 15189:2012: a pragmatic approach. Clin Chem Lab Med 2017;55:1501–8. PubMedGoogle Scholar

  • 18.

    Plebani M, Lippi G. Uncertainty, quality, safety and accreditation in laboratory medicine. J Lab Precis Med 2017;2:80. CrossrefGoogle Scholar

  • 19.

    Cornes MP, Church S, van Dongen-Lases E, Grankvist K, Guimarães JT, Ibarz M, et al. The role of European Federation of Clinical Chemistry and Laboratory Medicine Working Group for Preanalytical Phase in standardization and harmonization of the preanalytical phase in Europe. Ann Clin Biochem 2016;53:539–47. CrossrefGoogle Scholar

  • 20.

    Lippi G, Simundic AM. The EFLM strategy for harmonization of the preanalytical phase. Clin Chem Lab Med 2017. doi: 10.1515/cclm-2017-0277. [Epub ahead of print]. PubMedGoogle Scholar

  • 21.

    Simundic AM, Nikolac N, Ivankovic V, Ferenec-Ruzic D, Magdic B, Kvaternik M, et al. Comparison of visual vs. automated detection of lipemic, icteric and hemolyzed specimens: can we rely on a human eye? Clin Chem Lab Med 2009;47:1361–5. PubMedGoogle Scholar

  • 22.

    Luksic AH, Nikolac Gabaj N, Miler M, Dukic L, Bakliza A, Simundic AM. Visual assessment of hemolysis affects patient safety. Clin Chem Lab Med 2018;56:574–81. PubMedGoogle Scholar

  • 23.

    Simundic AM, Bilic-Zulle L, Nikolac N, Supak-Smolcic V, Honovic L, Avram S, et al. The quality of the extra-analytical phase of laboratory practice in some developing European countries and Mexico – a multicentric study. Clin Chem Lab Med 2011;49:215–28. PubMedGoogle Scholar

  • 24.

    Li L, Vecellio E, Gay S, Lake R, Mackay M, Burnett L, et al. Making sense of a haemolysis monitoring and reporting system: a nationwide longitudinal multimethod study of 68 Australian laboratory participant organizations. Clin Chem Lab Med 2018;56:565–73. Google Scholar

  • 25.

    Howanitz PJ, Lehman CM, Jones BA, Meier FA, Horowitz GL. Practices for Identifying and Rejecting Hemolyzed Specimens Are Highly Variable in Clinical Laboratories. Arch Pathol Lab Med 2015;139:1014–9. PubMedCrossrefGoogle Scholar

  • 26.

    Lippi G, Cadamuro J. Visual assessment of sample quality: quo usque tandem? Clin Chem Lab Med 2018;56:513–5. PubMedGoogle Scholar

  • 27.

    Lippi G. Systematic assessment of the hemolysis index: pros and cons. Adv Clin Chem 2015;71:157–70. PubMedCrossrefGoogle Scholar

  • 28.

    Sandberg S, Fraser CG, Horvath AR, Jansen R, Jones G, Oosterhuis W, et al. Defining analytical performance specifications: Consensus Statement from the 1st Strategic Conference of the European Federation of Clinical Chemistry and Laboratory Medicine. Clin Chem Lab Med 2015;53:833–5. PubMedGoogle Scholar

  • 29.

    Ceriotti F, Fernandez-Calle P, Klee GG, Nordin G, Sandberg S, Streichert T, et al. Criteria for assigning laboratory measurands to models for analytical performance specifications defined in the 1st EFLM Strategic Conference. Clin Chem Lab Med 2017;55:189–94. PubMedGoogle Scholar

  • 30.

    Perich C, Minchinela J, Ricós C, Fernández-Calle P, Alvarez V, Doménech MV, et al. Biological variation database: structure and criteria used for generation and update. Clin Chem Lab Med 2015;53:299–305. PubMedGoogle Scholar

  • 31.

    Pineda-Tenor D, Laserna-Mendieta EJ, Timón-Zapata J, Rodelgo-Jiménez L, Ramos-Corral R, Recio-Montealegre A, et al. Biological variation and reference change values of common clinical chemistry and haematologic laboratory analytes in the elderly population. Clin Chem Lab Med 2013;51:851–62. PubMedGoogle Scholar

  • 32.

    Carobene A, Marino I, Coşkun A, Serteser M, Unsal I, Guerra E, et al. The EuBIVAS Project: within- and between-subject biological variation data for serum creatinine using enzymatic and alkaline picrate methods and implications for monitoring. Clin Chem 2017;63:1527–36. CrossrefPubMedGoogle Scholar

  • 33.

    Harris EK, Yasaka T. On the calculation of a “reference change” for comparing two consecutive measurements. Clin Chem 1983;29:25–30. PubMedGoogle Scholar

  • 34.

    Fraser CG. Reference change values. Clin Chem Lab Med 2011;50:807–12. PubMedGoogle Scholar

  • 35.

    Simundic AM, Kackov S, Miler M, Fraser CG, Petersen PH. Terms and symbols used in studies on biological variation: the need for harmonization. Clin Chem 2015;61:438–9. CrossrefPubMedGoogle Scholar

  • 36.

    Bartlett WA, Braga F, Carobene A, Coşkun A, Prusa R, Fernandez-Calle P, et al. A checklist for critical appraisal of studies of biological variation. Clin Chem Lab Med 2015;53:879–85. PubMedGoogle Scholar

  • 37.

    Fraser CG, Lippi G, Plebani M. Reference change values may need some improvement but are invaluable tools in laboratory medicine. Clin Chem Lab Med 2012;50:963–4. Google Scholar

  • 38.

    Vermeer HJ, Thomassen E, de Jonge N. Automated processing of serum indices used for interference detection by the laboratory information system. Clin Chem 2005;51:244–7. Google Scholar

  • 39.

    Monneret D, Mestari F, Atlan G, Corlouer C, Ramani Z, Jaffre J, et al. Hemolysis indexes for biochemical tests and immunoassays on Roche analyzers: determination of allowable interference limits according to different calculation methods. Scand J Clin Lab Invest 2015;75:162–9. PubMedCrossrefGoogle Scholar

  • 40.

    Goyal T, Schmotzer CL. Validation of hemolysis index thresholds optimizes detection of clinically significant hemolysis. Am J Clin Pathol 2015;143:579–83. CrossrefPubMedGoogle Scholar

  • 41.

    Feitosa MS, Bücker DH, Santos SM, Vasconcellos LS. Implementation of criteria for automatic release of clinical chemistry test results in a laboratory at an academic public hospital. J Bras Patol Med Lab 2016;52:149–56. Google Scholar

  • 42.

    Ko DH, Park HI, Hyun J, Kim HS, Park MJ, Shin DH. Utility of reference change values for delta check limits. Am J Clin Pathol 2017;148:323–9. PubMedCrossrefGoogle Scholar

  • 43.

    Lippi G. Interference studies: focus on blood cell lysates preparation and testing. Clin Lab 2012;58:351–5. PubMedGoogle Scholar

  • 44.

    Dimeski G. Interference testing. Clin Biochem 2008;29(Suppl. 1):S43–8. Google Scholar

  • 45.

    Carraro P, Servidio G, Plebani M. Hemolyzed specimens: a reason for rejection or a clinical challenge? Clin Chem 2000;46:306–7. Google Scholar

  • 46.

    Lippi G, Pavesi F, Benegiamo A, Pipitone S. What do hemolyzed whole-blood specimens look like? Analysis with a CellaVision DM96 automated image analysis system. J Lab Autom 2015;20:60–3. CrossrefPubMedGoogle Scholar

  • 47.

    Vermeer HJ, Steen G, Naus AJ, Goevaerts B, Agricola PT, Schoenmakers CH. Correction of patient results for Beckman Coulter LX-20 assays affected by interference due to hemoglobin, bilirubin or lipids: a practical approach. Clin Chem Lab Med 2007;45:114–9. PubMedGoogle Scholar

  • 48.

    Hawkins RC. Correction and reporting of potassium results in haemolysed samples. Ann Clin Biochem 2006;43:88–9. PubMedCrossrefGoogle Scholar

  • 49.

    Jeffery J, Sharma A, Ayling RM. Detection of haemolysis and reporting of potassium results in samples from neonates. Ann Clin Biochem 2009;46:222–5. CrossrefPubMedGoogle Scholar

  • 50.

    Mansour MM, Azzazy HM, Kazmierczak SC. Correction factors for estimating potassium concentrations in samples with in vitro hemolysis: a detriment to patient safety. Arch Pathol Lab Med 2009;133:960–6. PubMedGoogle Scholar

  • 51.

    Lippi G, Avanzini P, Pavesi F, Bardi M, Ippolito L, Aloe R, et al. Studies on in vitro hemolysis and utility of corrective formulas for reporting results on hemolyzed specimens. Biochem Med (Zagreb) 2011;21:297–305. PubMedGoogle Scholar

  • 52.

    Lippi G, Salvagno GL, Guidi GC. Mean corpuscular volume and red blood cell distribution width are independent predictors of serum potassium concentration in healthy individuals. Clin Chim Acta 2015;446:117–8. CrossrefPubMedGoogle Scholar

  • 53.

    Zwart A, van Assendelft OW, Bull BS, England JM, Lewis SM, Zijlstra WG. ICSH Recommendations for reference method for hemoglobinometry in human blood (ICSH standards 1995) and specifications for international hemoglobincyanamyde standard. J Clin Pathol 1996;49:271–4. CrossrefPubMedGoogle Scholar

  • 54.

    Lippi G, Luca Salvagno G, Blanckaert N, Giavarina D, Green S, Kitchen S, et al. Multicenter evaluation of the hemolysis index in automated clinical chemistry systems. Clin Chem Lab Med 2009;47:934–9. PubMedGoogle Scholar

  • 55.

    Cadamuro J, von Meyer A, Wiedemann H, Klaus Felder T, Moser F, Kipman U, et al. Hemolysis rates in blood samples: differences between blood collected by clinicians and nurses and the effect of phlebotomy training. Clin Chem Lab Med 2016;54:1987–92. PubMedGoogle Scholar

  • 56.

    Sciacovelli L, Panteghini M, Lippi G, Sumarac Z, Cadamuro J, Galoro CA, et al. Defining a roadmap for harmonizing quality indicators in Laboratory Medicine: a consensus statement on behalf of the IFCC Working Group “Laboratory Error and Patient Safety” and EFLM Task and Finish Group “Performance specifications for the extra-analytical phases”. Clin Chem Lab Med 2017;55:1478–88. Google Scholar

  • 57.

    Lippi G, Sciacovelli L, Simundic AM, Plebani M. Innovative software for recording preanalytical errors in accord with the IFCC quality indicators. Clin Chem Lab Med 2017;55:e51–3. PubMedGoogle Scholar

  • 58.

    Lippi G, Adcock D, Simundic AM, Tripodi A, Favaloro EJ. Critical laboratory values in hemostasis: toward consensus. Ann Med 2017;49:455–61. CrossrefPubMedGoogle Scholar

  • 59.

    European Federation of Clinical Chemistry and Laboratory Medicine (EFLM). TFG-BVD “Biological variation database”. Available at: https://www.eflm.eu/site/page/a/1084. Accessed: 30 Oct 2017. 

  • 60.

    Lippi G, Salvagno GL, Montagnana M, Brocco G, Guidi GC. Influence of hemolysis on routine clinical chemistry testing. Clin Chem Lab Med 2006;44:311–6. PubMedGoogle Scholar

About the article

Corresponding author: Prof. Giuseppe Lippi, Section of Clinical Biochemistry, University Hospital of Verona, Piazzale LA Scuro, 37100 – Verona, Italy


Received: 2017-11-27

Accepted: 2017-12-15

Published Online: 2018-01-26

Published in Print: 2018-04-25


Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.

Research funding: None declared.

Employment or leadership: None declared.

Honorarium: None declared.

Competing interests: The funding organization(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.


Citation Information: Clinical Chemistry and Laboratory Medicine (CCLM), Volume 56, Issue 5, Pages 718–727, ISSN (Online) 1437-4331, ISSN (Print) 1434-6621, DOI: https://doi.org/10.1515/cclm-2017-1104.

Export Citation

©2018 Walter de Gruyter GmbH, Berlin/Boston.Get Permission

Citing Articles

Here you can find all Crossref-listed publications in which this article is cited. If you would like to receive automatic email messages as soon as this article is cited in other publications, simply activate the “Citation Alert” on the top of this page.

[1]
Fazıla Atakan Erkal, Güzin Aykal, Hayriye Melek Yalçınkaya, Nihal Aksoy, and Murat Özdemir
Turkish Journal of Biochemistry, 2019, Volume 0, Number 0
[2]
Eric S. Kilpatrick and Sverre Sandberg
Clinical Chemistry and Laboratory Medicine (CCLM), 2018, Volume 0, Number 0
[3]
Kjell Grankvist, Ruben Gomez, Mads Nybo, Gabriel Lima-Oliviera, and Alexander von Meyer
Diagnosis, 2018, Volume 0, Number 0
[4]
Giuseppe Lippi, Janne Cadamuro, Elisa Danese, Matteo Gelati, Martina Montagnana, Alexander von Meyer, Gian Luca Salvagno, Ana-Maria Simundic, and Pal Bela Szecsi
PLOS ONE, 2018, Volume 13, Number 7, Page e0200088
[5]
Giuseppe Lippi, Camilla Mattiuzzi, and Janne Cadamuro
LaboratoriumsMedizin, 2018, Volume 42, Number 3, Page 67

Comments (0)

Please log in or register to comment.
Log in