Measurement of the energy spectrum of secondary neutrons in a proton therapy environment

Martin Dommert / Marcel Reginatto / Miroslav Zboril / Fine Fiedler / Stephan Helmbrecht
  • Helmholtz-Zentrum Dresden – Rossendorf, Bautzner Landstr. 400, 01328 Dresden, Germany
  • Currently: Senacor Technologies AG, Berlin, Germany
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Wolfgang Enghardt
  • Helmholtz-Zentrum Dresden – Rossendorf, Bautzner Landstr. 400, 01328 Dresden, Germany
  • OncoRay – National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden, Helmholtz-Zentrum Dresden – Rossendorf, Fetscherstr. 74, PF 41, 01307 Dresden, Germany
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Benjamin Lutz
Published Online: 2017-09-07 | DOI: https://doi.org/10.1515/cdbme-2017-0018

Abstract

Measurement of the energy spectrum of secondary neutrons were carried out at the OncoRay Proton Therapy facility in Dresden, following an approach originating in neutron metrology which is well suited for both the characterization of secondary neutron fields at proton therapy facilities and the validation of Monte Carlo simulations. For the experiment, a brass target was placed in the proton beam and Bonner spheres measurements were made at a distance of 2 m from the target and at different angles, 15° to 120°, with respect to the incoming proton beam. The measured spectra were compared to Monte Carlo simulations.

Keywords: proton therapy; secondary neutrons; neutron spectrometry; Bonner spheres

About the article

Published Online: 2017-09-07


Citation Information: Current Directions in Biomedical Engineering, Volume 3, Issue 2, Pages 83–86, ISSN (Online) 2364-5504, DOI: https://doi.org/10.1515/cdbme-2017-0018.

Export Citation

©2017 Martin Dommert et al., published by De Gruyter, Berlin/Boston. This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License. BY-NC-ND 4.0