Usability and accuracy of high-resolution detectors for daily quality assurance for robotic radiosurgery

Britta Loutfi-Krauss
  • Corresponding author
  • Universitäts-klinikum Frankfurt, Klinik für Strahlentherapie und Onkologie, Theodor-Stern-Kai 7, 60590 Frankfurt, Germany
  • Email
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Marie-Christin Damme / Tenzin Sonam Stelljes
  • Carl von Ossietzky Universität, Universitätsklinik für Medizinische Strahlenphysik, Campus Pius Hospital, Georgstraße 12, 26121 Oldenburg, Germany
  • Email
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Mark KH Chan
  • Universitätsklinikum Schleswig-Holstein, Klinik für Strahlentherapie, Arnold-Heller-Straße 3, 24105 Kiel, Germany
  • Email
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Frank-Andre Siebert
  • Universitätsklinikum Schleswig-Holstein, Klinik für Strahlentherapie, Arnold-Heller-Straße 3, 24105 Kiel, Germany
  • Email
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Björn Poppe
  • Carl von Ossietzky Universität, Universitätsklinik für Medizinische Strahlenphysik, Campus Pius Hospital, Georgstraße 12, 26121 Oldenburg, Germany
  • Email
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Ulla Ramm
  • Universitäts-klinikum Frankfurt, Klinik für Strahlentherapie und Onkologie, Theodor-Stern-Kai 7, 60590 Frankfurt, Germany
  • Email
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Oliver Blanck
  • Universitätsklinikum Schleswig-Holstein, Klinik für Strahlentherapie, Arnold-Heller-Straße 3, 24105 Kiel, Germany
  • Saphir Radiochirurgie, Friedrich-Trendelenburg-Allee 2, 18273 Güstrow und Schleusenweg 2-16, 60528 Frankfurt am Main, Germany, e-mail:
  • Email
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
Published Online: 2017-09-07 | DOI: https://doi.org/10.1515/cdbme-2017-0057

Abstract

For daily CyberKnife QA a Winston-Lutz-Test (Automated-Quality-Assurance, AQA) is used to determine sub-millimeter deviations in beam delivery accuracy. This test is performed using gafchromic film, an extensive and user-dependent method requiring the use of disposables. We therefore analyzed the usability and accuracy of high-resolution detector arrays. We analyzed a liquid-filled ionization-chamber array (Octavius 1000SRS, PTW, Germany), which has a central resolution of 2.5mm. To test sufficient sensitivity, beam profiles with robot shifts of 0.1mm along the arrays' axes were measured. The detected deviation between the shifted and central profile were compared to the real robot's position. We then compared the results to the SRS-Profiler (SunNuclear, USA) with 4.0mm resolution and to the Nonius (QUART, Germany), a single-line diode detector with 2.8mm resolution. Finally, AQA variance and usability were analyzed performing a number of AQA tests over time, which required the use of specially designed fixtures for each array, and the results were compared to film. Concerning sensitivity, the 1000SRS detected the beam profile shifts with a maximum difference of 0.11mm (mean deviation = 0.03mm) compared to the actual robot shift. The Nonius and SRS-Profiler showed differences of up to 0.15mm and 0.69mm with mean deviation of 0.05mm and 0.18mm, respectively. Analyzing the variation of AQA results over time, the 1000SRS showed a comparable standard deviation to film (0.26mm vs. 0.18mm). The SRS-Profiler and the Nonius showed a standard deviation of 0.16mm and 0.24mm, respectively. The 1000SRS seems to provide equivalent accuracy and sensitivity to the gold standard film when performing daily AQA tests. Compared to other detectors in our study the sensitivity as well as the accuracy of the 1000SRS appears to be superior and more user-friendly. Furthermore, no significant modification of the standard AQA procedure is required when introducing 1000SRS for CyberKnife AQA.

Keywords: CyberKnife; Robotic Radiosurgery; Automatic Quality Assurance; High-Resolution Arrays; Gafchromic Film