Jump to ContentJump to Main Navigation
Show Summary Details
Weitere Optionen …

Current Directions in Biomedical Engineering

Joint Journal of the German Society for Biomedical Engineering in VDE and the Austrian and Swiss Societies for Biomedical Engineering

Editor-in-Chief: Dössel, Olaf

Wissenschaftlicher Beirat: Augat, Peter / Buzug, Thorsten M. / Haueisen, Jens / Jockenhoevel, Stefan / Knaup-Gregori, Petra / Kraft, Marc / Lenarz, Thomas / Leonhardt, Steffen / Malberg, Hagen / Penzel, Thomas / Plank, Gernot / Radermacher, Klaus M. / Schkommodau, Erik / Stieglitz, Thomas / Urban, Gerald A.


CiteScore 2018: 0.47

Source Normalized Impact per Paper (SNIP) 2018: 0.377

Open Access
Online
ISSN
2364-5504
Alle Formate und Preise
Weitere Optionen …

Neuro control – using neuro monitoring data to control networked active instruments

Christian Dietz
  • Korrespondenzautor
  • Institute of Micro Technology and Medical Device Technology, Technical University of Munich (TUM), Boltzmannstraße 15, 85748 Garching, Germany
  • E-Mail
  • Weitere Artikel des Autors:
  • De Gruyter OnlineGoogle Scholar
/ Jonas H. Pfeiffer
  • Institute of Micro Technology and Medical Device Technology, Technical University of Munich (TUM), Boltzmannstraße 15, 85748 Garching, Germany
  • E-Mail
  • Weitere Artikel des Autors:
  • De Gruyter OnlineGoogle Scholar
/ Max E. Dingler
  • Institute of Micro Technology and Medical Device Technology, Technical University of Munich (TUM), Boltzmannstraße 15, 85748 Garching, Germany
  • E-Mail
  • Weitere Artikel des Autors:
  • De Gruyter OnlineGoogle Scholar
/ Tim C. Lüth
  • Institute of Micro Technology and Medical Device Technology, Technical University of Munich (TUM), Boltzmannstraße 15, 85748 Garching, Germany
  • E-Mail
  • Weitere Artikel des Autors:
  • De Gruyter OnlineGoogle Scholar
Online erschienen: 07.09.2017 | DOI: https://doi.org/10.1515/cdbme-2017-0094

Abstract

Intraoperative neurophysiological monitoring (IONM) can be used to monitor the neural integrity and help the surgeon to reduce iatrogenic patient damage. We present a networked digital assistance system based on real-time IONM data. The system automatically reduces the power of active medical ablation instruments when they endanger neural structures. We integrated a nerve monitor into a real-time network of medical devices. In doing so, neurophysiological signals are available to all connected medical devices in real-time. With the IONM data a so-called function module Neuro Control calculates and controls the maximum allowed power for active instruments like an ultrasonic dissector.

Keywords: Neuro Control; neuromonitoring; OR.NET; SRTB; real-time network; medical devices; interoperability