Preparation of a miniaturised iodide ion selective sensor using polypyrrole and pencil lead: effect of double-coating, electropolymerisation time, and current density

Mehrdad Gholami
  • Department of Chemistry, Marvdasht Branch, Islamic Azad University, P.O. Box 465, Marvdasht, Iran
  • Email:
/ Abdol-Majid Ghasemi
  • Engineering Institute of Agriculture Ministry, Fars Engineering Research Center (FERC), P.O. Box 71555414, Shiraz, Iran
  • Email:
/ Mohammad Loghavi
  • Engineering Institute of Agriculture Ministry, Fars Engineering Research Center (FERC), P.O. Box 71555414, Shiraz, Iran
  • Email:
/ Shima Behkami
  • Department of Chemistry, Marvdasht Branch, Islamic Azad University, P.O. Box 465, Marvdasht, Iran
  • Email:
/ Amaneh Ahamdi-Dokht-Faraghe
  • Engineering Institute of Agriculture Ministry, Fars Engineering Research Center (FERC), P.O. Box 71555414, Shiraz, Iran
  • Email:
Published Online: 2013-05-03 | DOI: https://doi.org/10.2478/s11696-013-0342-8

Abstract

Polypyrrole (PPy) is a conducting polymer which can be used for producing different ion-selective electrodes. An iodide-doped (PPy-iodide) was prepared electrochemically by anodic polymerisation of pyrrole in the presence of an iodide ion in an aqueous solution on the surface of a pencil lead. Polymerisation was investigated under galvanostatic conditions. The effects of electropolymerisation conditions on the characteristics of the potential response of the sensor were examined. Concentrations of pyrrole, iodide ions, and conditioning solution plus current density and the time of electropolymerisation were optimised in relation to the slope and linearity of calibration graphs. This electrode showed a Nernstian behaviour of 61.1 mV per decade for I− ion over a wide concentration range from 1.0 × 10−5 M to 1.0 × 10−1 M, with the limit of detection of 9.3 × 10−6 M. The response time of the electrode was from 3–5 s. The selectivity coefficients of the prepared sensors over a wide spectrum of interference anions were also evaluated, revealing that selectivity improves as a result of double-coating with PPy. A similar improvement was observed under lower current density and longer electropolymerisation time. This sensor was applied in the determination of iodide ions using titration potentiometry. This electrode can be used for the determination of iodide in drug preparations.

Keywords: ion selective electrodes; polypyrrole; conducting polymers; iodide anion; sensor

  • [1] Albano, D. R., & Sevilla, F., III (2007). Piezoelectric quartz crystal sensor for surfactant based on molecularly imprinted polypyrrole. Sensors and Actuators B: Chemical, 121, 129–134. DOI: 10.1016/j.snb.2006.09.006. http://dx.doi.org/10.1016/j.snb.2006.09.006 [Web of Science] [Crossref]

  • [2] Ansari, R. (2006). Polypyrrole conducting electroactive polymers: Synthesis and stability studies. E-Journal of Chemistry, 3, 186–201. DOI: 10.1155/2006/860413. http://dx.doi.org/10.1155/2006/860413 [Crossref]

  • [3] Aravamudhan, S., & Bhansali, S. (2008). Development of micro-fluidic nitrate-selective sensor based on doped-polypyrrole nanowires. Sensors and Actuators B: Chemical, 132, 623–630. DOI: 10.1016/j.snb.2008.01.046. http://dx.doi.org/10.1016/j.snb.2008.01.046 [Web of Science] [Crossref]

  • [4] Bendikov, T., Kim, J., & Harmon, T. C. (2005). Development and environmental application of a nitrate selective microsensor based on doped polypyrrole films. Sensors and Actuators B: Chemical, 106, 512–517. DOI: 10.1016/j.snb.2004.07.018. http://dx.doi.org/10.1016/j.snb.2004.07.018 [Crossref]

  • [5] Bobacka, J., Lewenstam, A., & Ivaska, A. (1993). Potentiometric response of poly (3-octylthiophene), poly(3-methylthiophene) and polythiophene in aqueous solutions. Talanta, 40, 1437–1444. DOI: 10.1016/0039-9140(93)80223-e. http://dx.doi.org/10.1016/0039-9140(93)80223-E [Crossref]

  • [6] Bobacka, J., Ivaska, A., & Lewenstam, A. (2003). Potentiometric ion sensors based on conducting polymers. Electroanalysis, 15, 366–374. DOI: 10.1002/elan.200390042. http://dx.doi.org/10.1002/elan.200390042 [Crossref]

  • [7] Bradner, F. P., & Shapiro, J. S. (1988). Improvement in the quality of polypyrrole films prepared electrochemically on a mercury anode. Synthetic Metals, 26, 69–77. DOI: 10.1016/0379-6779 (88)90336-0. http://dx.doi.org/10.1016/0379-6779(88)90336-0 [Crossref]

  • [8] Cadogan, A., Lewenstam, A., & Ivaska, A. (1992). Anionic responses of electrochemically synthesized polypyrrole films. Talanta, 39, 617–620. DOI: 10.1016/0039-9140(92)80070-t. http://dx.doi.org/10.1016/0039-9140(92)80070-T [Crossref]

  • [9] Diaz, A. F., & Lacroix, J. C. (1988). Synthesis of electroactive/conductive polymer films: Electro-oxidation of heteroaromatic compounds. New Journal of Chemistry, 12, 171–180.

  • [10] Ersin Karagözler, A., Yavuz Ataman, O., Galal, A., Xue, Z. L., Zimmer, H., & Mark, H. B., Jr. (1991). Potentiometric iodide ion sensor based on a conducting poly(3-methylthiophene) polymer film electrode. Analytica Chimica Acta, 248, 163–172. DOI: 10.1016/s0003-2670(00)80880-6. http://dx.doi.org/10.1016/S0003-2670(00)80880-6 [Crossref]

  • [11] García Alonso, J., de la Fuente Sánchez, C., Tascón García, M. L., Vázquez Barbado, M. D., & Sánchez Batanero, P. (1999). Electrocatalytical properties of a potassium hexacyanoferrate (II) modified polypyrrole electrode. Application to vitamin C determination. Analytical Letters, 32, 1961–1980. DOI: 10.1080/00032719908542945. http://dx.doi.org/10.1080/00032719908542945 [Crossref]

  • [12] Ge, H., Zhang, J., & Wallace, G. G. (1992). Use of overoxidised polypyrrole as a chromium(VI) sensor. Analytical Letters, 25, 429–441. DOI: 10.1080/00032719208016106. http://dx.doi.org/10.1080/00032719208016106 [Crossref]

  • [13] Ikariyama, Y., Galiatsatos, C., Heineman, W. R., & Yamauchi, S. (1987). Polypyrrole electrode as a detector of anionic substances. Sensors and Actuators, 12, 455–461. DOI: 10.1016/0250-6874(87)80064-1. http://dx.doi.org/10.1016/0250-6874(87)80064-1 [Crossref]

  • [14] Jovanovic, V. M., Dekanski, A., Vlajnic, G., & Jovanovic, M. S. (1997). Electrochemical and surface characterization of pH sensor based on bisulfate doped poly(pyrrole). Electroanalysis, 9, 564–569. DOI: 10.1002/elan.1140090712. http://dx.doi.org/10.1002/elan.1140090712 [Crossref]

  • [15] Lindfors, T., Bobacka, J., & Ivaska, A. (1997). Electrosynthesis of polypyrrole in iodide solution. Film growth, redox behaviour and potentiometric response. Analytica Chimica Acta, 355, 217–225. DOI: 10.1016/s0003-2670(97)00488-1. http://dx.doi.org/10.1016/S0003-2670(97)00488-1 [Crossref]

  • [16] Maksymiuk, K., Bobacka, J., Ivaska, A., & Lewenstam, A. (2000). Coupled redox and pH potentiometric responses of electrodes coated with polypyrrole. Analytical Letters, 33, 1339–1360. DOI: 10.1080/00032710008543126. http://dx.doi.org/10.1080/00032710008543126 [Crossref]

  • [17] Omastová, M., & Mičušík, M. (2012). Polypyrrole coating of inorganic and organic materials by chemical oxidative polymerization. Chemical Papers, 66, 392–414. DOI: 10.2478/s11696-011-0120-4. http://dx.doi.org/10.2478/s11696-011-0120-4 [Crossref]

  • [18] Şahin, M., čSahin, Y., & Özcan, A. (2008). Ion chromatography-potentiometric detection of inorganic anions and cations using polypyrrole and overoxidized polypyrrole electrode. Sensors and Actuators B: Chemical, 133, 5–14. DOI: 10.1016/j.snb.2008.01.004 http://dx.doi.org/10.1016/j.snb.2008.01.004 [Web of Science] [Crossref]

  • [19] Şahin, M., Özcan, L., Usta, B., & Şahin, Y. (2009). Determination of ascorbic acid by polypyrrole potentiometric detector in ion chromatography. Biosensors and Bioelectronics, 24, 3492–3497. DOI: 10.1016/j.bios.2009.05.00 http://dx.doi.org/10.1016/j.bios.2009.05.005 [Web of Science] [Crossref]

  • [20] Shafiee-Dastjerdi, L., & Alizadeh, N. (2004). Coated wire linear alkylbenzenesulfonate sensor based on polypyrrole and improvement of the selectivity behavior. Analytica Chimica Acta, 505, 195–200. DOI: 10.1016/j.aca.2003.10.062. http://dx.doi.org/10.1016/j.aca.2003.10.062 [Crossref]

  • [21] Shin, M. C., & Kim, H. S. (1995). Effects of enzyme concentration and film thickness on the analytical performance of a polypyrrole/glucose oxidase biosensor. Analytical Letters, 28, 1017–1031. DOI: 10.1080/00032719508002676. http://dx.doi.org/10.1080/00032719508002676 [Crossref]

  • [22] Skotheim, T. A. (1986). Handbook of conducting polymers. (Vol. 1). New York, NY, USA: Marcel Dekker.

  • [23] Umezawa, Y., Bühlmann, P., Umezawa, K., Tohda, K., & Amemiya, S. (2000). Potentiometric selectivity coefficients of ion-selective electrodes. Part I. Inorganic cations (Technical Report). Pure and Applied Chemistry, 72, 1851–2082. DOI: 10.1351/pac200072101851. http://dx.doi.org/10.1351/pac200072101851 [Crossref]

  • [24] Wang, J., Chen, S. P., & Lin, M. S. (1989). Use of different electropolymerzation conditions for controlling the sizeexclusion selectivity at polyaniline, polypyrrole and polyphenol films. Journal of Electroanalytical Chemistry and Interfacial Electrochemistry, 273, 231–242. DOI: 10.1016/0022-0728(89)87016-0. http://dx.doi.org/10.1016/0022-0728(89)87016-0 [Crossref]

  • [25] Ying, M., Yuan, R., Song, Y. Q., & Li, Z. Q. (1997). Highly selective iodide poly(vinyl chloride) membrane electrode based on a nickel(II) tetraazaannulene macrocyclic complex. Analyst, 122, 1143–1146. DOI: 10.1039/a700544j. http://dx.doi.org/10.1039/a700544j [Crossref]

  • [26] Zanganeh, A. R., & Amini, M. K. (2007). A potentiometric and voltammetric sensor based on polypyrrole film with electrochemically induced recognition sites for detection of silver ion. Electrochimica Acta, 52, 3822–3830. DOI: 10.1016/j.electacta.2006.10.055. http://dx.doi.org/10.1016/j.electacta.2006.10.055 [Crossref] [Web of Science]

About the article

Published Online: 2013-05-03

Published in Print: 2013-08-01



Citation Information: Chemical Papers, ISSN (Online) 1336-9075, DOI: https://doi.org/10.2478/s11696-013-0342-8. Export Citation

Citing Articles

Here you can find all Crossref-listed publications in which this article is cited. If you would like to receive automatic email messages as soon as this article is cited in other publications, simply activate the “Citation Alert” on the top of this page.

[1]
Nur Rahimah Said, Majid Rezayi, Leila Narimani, Nassir N. Al-Mohammed, Ninie Suhana Abdul Manan, and Yatimah Alias
Electrochimica Acta, 2016, Volume 197, Page 10
[2]
Mehrdad Gholami, Majid Rezayi, Pooria Moozarm Nia, Ismail Yusoff, and Yatimah Alias
Measurement, 2015, Volume 69, Page 115

Comments (0)

Please log in or register to comment.
Log in