Effects of different production technologies on mechanical and metallurgical properties of precious metal denture alloys

Paolo Ferro
  • Corresponding author
  • University of Padova, Department of Management and Engineering, Stradella S. Nicola, 3 I-36100 Vicenza, Italy
  • Email
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Eleonora Battaglia
  • University of Padova, Department of Management and Engineering, Stradella S. Nicola, 3 I-36100 Vicenza, Italy
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Stefano Capuzzi
  • University of Padova, Department of Management and Engineering, Stradella S. Nicola, 3 I-36100 Vicenza, Italy
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
/ Filippo Berto
  • NTNU, Department of Engineering Design and Materials, Richard Birkelands vei 2b, 7491, Trondheim, Norway
  • Other articles by this author:
  • De Gruyter OnlineGoogle Scholar
Published Online: 2017-12-29 | DOI: https://doi.org/10.1515/eng-2017-0043

Abstract

Background

Precious metal alloys can be supplied in traditional plate form or innovative drop form with high degree of purity.

Objective

The aim of the present work is to evaluate the influence of precious metal alloy form on metallurgical and mechanical properties of the final dental products with particular reference to metal-ceramic bond strength and casting defects.

Method

A widely used alloy for denture was selected; its nominal composition was close to 55 wt% Pd – 34 wt% Ag – 6 wt% In – 3 wt% Sn. Specimens were produced starting from the alloy in both plate and drop forms. A specific test method was developed to obtain results that could be representative of the real conditions of use. In order to achieve further information about the adhesion behaviour and resistance, the fracture surfaces of the samples were observed using ‘Scanning Electron Microscopy (SEM)’. Moreover, material defects caused by the moulding process were studied.

Results

The form of the alloy before casting does not significantly influence the shear bond strength between the metal and the ceramic material (p-value=0,976); however, according to SEM images, products from drop form alloy show less solidification defects compared to products obtained with plate form alloy. This was attributed to the absence of polluting additives used in the production of drop form alloy.

Conclusions

This study shows that the use of precious metal denture alloys supplied in drop form does not affect the metal-ceramic bond strength compared to alloys supplied in the traditional plate form. However, compared to the plate form, the drop form is found free of solidification defects, less expensive to produce and characterized by minor environmental impacts.

Keywords: Precious alloy; Denture; Defect; Scanning Electron Microscopy; Shear test; Process production; Bound strength

1 Introduction

Teeth play a critically important role in our lives. Loss of their function affects our ability to eat a balanced diet with negative effects on our systemic health. Loss of their aesthetics can negatively impact social function. However, both function and aesthetics can be restored.

Metal-ceramic restorations have been considered from mid-1950s [1]. Since that time, research and improvements in materials and techniques have dramatically increased the use of them. Nowadays, alternatives to these systems exist, mainly all-ceramic restorations that present advantages in terms of aesthetic properties [2, 3]. However, ceramics are brittle and subjected to premature failure, especially under repeated contact loading in a moist environment [4].

Densely sintered zirconia is significantly more stable as framework material than other ceramics; but, complications may occur such as discolorations, secondary caries and loss of retention [5, 6]. According to Donovan et al. [7], the major problem related to all-ceramics restorations is the greater incidence of veneering porcelain chipping (8–50% at 1–2 years) when compared with porcelain fused to metal restorations (4–10% at 10 years).

As reported by Pjetursson et al. [8], metal-ceramic fixed dental prostheses (FDPs) had lower failure rates than all-ceramic FDPs after a mean observation period of at least 3 years.

Eliasson et al. [9]indicates that FPDs fabricated using high-noble alloys have shown survival rates of 80 to 98%, 81 to 97% and 74 to 85% after 5, 10 and 15 years, respectively. Therefore, the metal–ceramic restoration still is the major and the most reliable type of dental restoration; especially, when a good adhesion between the two materials is achieved.

Even though various theories have been proposed and the variables affecting the metal-ceramic bond strength studied, the real mechanism of bonding is still not clear [10, 11]. As indicated by Anusavic et al. [12] the actual bond strength is difficult to determine with tests because of non-uniform stress distributions.

Even though a standard to evaluate the de-bonding of the ceramics by using a three-point bending text exists [13], shear test is the main alternative adopted in literature.

Both the ceramic fracture and bond strength assessment in bend tests are influenced by the Young’s modulus of the alloy. Therefore, it is not clear which characteristic of the metal-ceramic specimens, whether the bond strength or the modulus of elasticity of the metal, is actually tested [14].

There are two types of flexural tests reported in literature, 3-points [6, 15] and 4-point bending tests [16]; while, in the case of shear test, the geometry of the sample and the testing machine can vary.

Henriques et al. [17, 18] and De Melo et al. [19] used the planar interface shear bond strength test, while Shell et al. [20] and Neto et al. [21] used a modified rectangular parallel shear test.

Many types of metal can be used in metal-ceramic restorations [22]. The alloys used for restoration must meet certain minimum requirements for strength, stability, castability, corrosion/tarnish resistance, burnishability, polishability and biocompatibility. Metal-ceramic systems, compared to full metal restoration, have additional requirements that are necessary to achieve a good adhesion between the alloy and the ceramic. Requirements of these alloys include higher melting temperature, thermal compatibility with ceramics, oxide formation and sag resistance.

Gold is the oldest dental restorative material, having been used for dental repairs by Phoenicians and Romans [23]. Gold alloys, in particular Au-Pt-Pd alloys, are the first ones used for metal-ceramic restorations in the last 30 years [24]. As the cost of gold increased, new noble metals were introduced in dentistry. In particular, Pd alloys such as Pd-Ag and Pd-Cu alloys are the most widely used. In recent years, the use of alternative alloys such as base-metal alloys has increased worldwide. The choice of the based-alloy varies from country to country. In the USA nickel-chromium (Ni-Cr) alloys are frequently used, with or without beryllium (Be), while the alloys most often used in Europe and Japan are the cobalt-chromium (Co-Cr) alloys [21, 25, 26]. Titanium alloys are also an alternative [27, 28]. Their elastic moduli are nearly twice as high as those of the other systems, and their hardness may reach 350 HV [29]. Based metal alloys are often selected when cost is of major consideration, while noble metal casting alloys are generally used to fabricate the metal substructure, because of their advantages in terms of biocompatibility and excellent metal–ceramic bond. Moreover, casting of base metal alloys is more difficult than that of noble alloys because of their high melting temperatures and potentials for oxidation during casting [30]. Finally, due to the high hardness of many base metal alloys, a lot of time is required for the finishing operations in the dental laboratory.

Since the reduction of the production cost for noble-metal alloys is a key factor for the diffusion of these alloys in restoration, new technologies should be developed to achieve this result.

Precious metal alloys can be supplied in traditional plate form (PF) or innovative drop form (DF) with higher degree of purity and lower production cost. This research is aimed to investigate the effect of metal fabrication technique process on the metallurgical and mechanical properties of the alloy with particular attention to the metal-ceramic adhesion.

2 Materials and methods

2.1 Alloy preparation

Table 1 summarizes the chemical composition of the alloy used in this work, obtained by optical emission spectrometry, while in Table 2 the most relevant material properties are reported.

Table 1

Base alloy composition (wt%)

Table 2

Relevant material properties 31

Two forms of the same alloy were used to produce samples: plate alloy and drop alloy.

The plates are produced by melting the alloy at 1450° C (in a protective atmosphere) and pouring it into a copper stirrup with dimensions equal to 160x105x6 mm. The alloy is then rolled with four rolling passes in order to decrease its thickness to 1 mm. Before each rolling step the material is annealed at 950° C for 1 hour in argon atmosphere and subsequently quenched. Finally, the alloy is cut into several plates and cleaned with solvents to remove oil and impurities.

The drops are produced by melting the alloy at 1380 °C in a protective atmosphere and by sifting it through spinneret holes of diameter equal to 1,4 mm. Finally, the drops are rinsed with soft water and dried into a spin drier.

2.2 Specimens manufacturing

Both alloy forms were used to produce alloy-ceramic specimens with an opaque interlayer at the metal-ceramic interface. In particular, the alloy was casted adopting the lost-wax casting technique and all the samples were manufactured with the same geometry. Fig. 1 shows the sample geometry.

Geometry of the alloy-ceramic specimen (dimensions in mm)
Figure 1

Geometry of the alloy-ceramic specimen (dimensions in mm)

Alloy specimens were sandblasted with 110 μm alumina grains and then heat treated by increasing the temperature from 650°C up to 950°C with a heat rate of 65°C/min in order to oxidize the surfaces. Finally, a vaporization, which is a cleaning process aimed to eliminate Aluminium remains, was carried out at a pressure of 4,5 bar. An opaque interlayer was then applied manually in the form of a creamy paste mixed with an oily substance and dried in the oven starting from a temperature of 450°C up to 950°C (heat rate of approximately 60°C/min). At the end of the cycle, the specimens were removed from the oven and allowed to cool down. A ceramic coat (Shofu Vintage Halo) was then applied. The ceramic material used in this study was in the form of powder mixed with distilled water.

In order to allow the alignment between the crosshead hole and sample, the layers of opaque and ceramic material were applied at a distance of 5 mm from the upper end of the sample (Fig. 1). Since the opaque and ceramic depositions were applied manually, slight geometric irregularities at the edge between metal substrate and ceramic material were inevitable; thus, a grinding process was carried out in order to assure the repeatability of the test and a uniform contact between the crosshead and the opaque-ceramic material. In Fig. 2, the sample before and after grinding process is shown. Three specimens for each alloy form (PF and DF) were manufactured and mechanically tested.

Edge between metal and ceramic material (a) before and (b) after grinding process
Figure 2

Edge between metal and ceramic material (a) before and (b) after grinding process

2.3 Shear Bond Strength Test

Shear bond tests were carried out at room temperature by means of a universal testing machine (MTS 858 Mini Bionix) with a load cell capacity of 25 kN and a crosshead speed of 2 mm/min.

Fig. 3 shows a schematic of the testing system, which consists of the metal ceramic sample described in 2.2 (A) and a custom-made C45 steel crosshead (B) used to apply the load. The sample was secured in vertical position at the fixed part of the testing machine. The crosshead hole diameter was 5,80 mm (tolerance range H7, according to UNI EN 20286/1-2 [32]). In order to obtain a good matching between the crosshead and the sample, the latter was grinded to achieve a final diameter equal to 5,75 mm (tolerance range js9). After the alignment, a compressive load was applied on the metal-ceramic interface until fracture occurred.

Schematic of the testing system
Figure 3

Schematic of the testing system

Figure 4 shows the sample-crosshead alignment before (Figure 4a) and after testing (Figure 4b).

Shear bond strength test architecture (a) before and (b) after the ceramic detachment
Figure 4

Shear bond strength test architecture (a) before and (b) after the ceramic detachment

The shear bond strength (SBS) was determined according to Eq. 1:

SBS(MPa)=F(N)Cad(mm)Sp(mm)(1)

where, F is the force applied by the crosshead on the metal-ceramic interface, Cad is the ceramic axial development and Sp is the sample perimeter.

Both the SBS trend and its value at failure, which refers to the complete detachment between the ceramic material and the substrate, were assessed.

2.4 Analysis of the metal-ceramic interface and failure mode

The analyses of the fracture surfaces were performed by a Scanning Electron Microscope equipped with an energy-dispersive spectrometer (EDS). SEM micrographs were quantitatively analysed by using the Leica QWin image analyser.

Two modes of failure were observed: (1) fracture occurring at metal-ceramic interface (no remains of opaque or ceramic were found on the metal substrate); (2) fracture occurring within the opaque-ceramic side (remains of ceramic or opaque were found on the metal substrate). Four SEM images were taken from each sample in order to obtain a panoramic view of the fracture surface. The images were then analysed using image analysis in order to quantify the metal-ceramic fracture surface after testing (Fig. 5 shows one of the four images obtained and analysed).

Fracture surface of a sample obtained with the DF precious alloy (a) before and (b) after visible metal area quantification
Figure 5

Fracture surface of a sample obtained with the DF precious alloy (a) before and (b) after visible metal area quantification

2.5 Statistical analysis

With the aim to determine whether there were any significant differences between the average values of two or more independent groups, data were analysed using a one-way ANOVA. P-values lower than 0,05 were considered statistically significant.

3 Results

3.1 Shear Bond Strength Test

Fig. 6 shows the results of the SBS test done on two specimens obtained from the two different alloy forms analyzed. All tests were interrupted at a crosshead displacement of 2 mm. No interference between crosshead hole and the sample was observed. The increasing of the shear stress with the increasing of the crosshead displacement is due to the attachment of the ceramic residues to the sample after the primary detachment at the first peak of the curve.

Example of shear bond strength results
Figure 6

Example of shear bond strength results

The mean results, assessed from all the tests carried out on plate and drop form specimens, are reported in Figure 7. The SBS recorded for plate form specimens was 14,36 ± 4,48 MPa (number of tests: 4); while, in the case of drop form, the value was 14,06 ± 3,48 MPa (number of tests: 4).

Mean values of SBS with corresponding standard deviations
Figure 7

Mean values of SBS with corresponding standard deviations

3.2 Metal-ceramic interface and failure mode

According to the area percentage of metal visible on the surface of the samples, specimens obtained with drop form alloy were found to present a slightly stronger metal-ceramic bond compared to specimens obtained with plate form alloy. As a matter of fact, the first ones showed less remnants of metal substrate than the second ones (Fig. 8). In particular, the average metal remain areas were 14,22% and 18,18% for drop and plate form, respectively.

Metal remain areas on (a) drop form and (b) plate form specimen
Figure 8

Metal remain areas on (a) drop form and (b) plate form specimen

Different EDS analyses were performed on samples of drop and plate form alloy in order to exclude the influence of base metal, ceramic material and oxide layer chemical composition differences on SBS results. An example is shown in Fig. 9.

SEM observations and EDS analyses of the oxide layer: (a) drop form alloy, (b) plate form alloy
Figure 9

SEM observations and EDS analyses of the oxide layer: (a) drop form alloy, (b) plate form alloy

Since the SBS results may also be influenced by the oxide and opaque thickness, different optical microscope (OM) micrographs were used to calculate their values (Fig. 10). In Table 3 such values are listed for both the alloy forms. With regard to the opaque layer, the minimum and the maximum thicknesses were assessed; thus, determining a range of variation. Regarding the oxide thickness value, the results listed in Table 3 are the average values of at least thirty measurements.

Metallographic images of the sample interfaces at 200× magnification: (a) drop form alloy, (b) plate form alloy
Figure 10

Metallographic images of the sample interfaces at 200× magnification: (a) drop form alloy, (b) plate form alloy

Table 3

Results of opaque and oxide layer assessments.

3.3 Bulk precious alloy analysis

In Fig. 11, SEM images of the metal substrate are shown. The sample obtained by plate form alloy (Fig. 11b) shows the presence of gas porosity probably due to contaminants (impurities, oil remains or solvents) used during the production process of the alloy. In contrast, the sample obtained by drop form alloy shows no solidification defects.

SEM images of metal substrate (a) drop form alloy, (b) plate form alloy
Figure 11

SEM images of metal substrate (a) drop form alloy, (b) plate form alloy

3.4 Statistical analysis of SBS data

There were no statistically significant differences between mean SBS values determined by one-way ANOVA (p > 0,05). Results of the two experimental conditions, are summarized in Table 4.

Table 4

Results of one-way ANOVA for tested conditions according to shear bond strength data.

4 Discussion

This study evaluates the effect of the alloy form (drop or plate form) used in the production of dental products on the metal-ceramic bond strength. Furthermore, some investigations were carried out in order to assess the metallurgical quality of the casting products obtained, starting from the two analyzed alloy forms.

4.1 Shear test

The standard [13] suggests a three-point-bending-test as testing method; it indicates that the minimum acceptable bond strength to achieve is 25 MPa. On the contrary, Anusavice [33] uses a shear test as testing method and suggests a minimum in vitro bond strength equal to 51 MPa. The SBS values obtained in this study are lower than both the limits suggested in the above-mentioned references. However, due to geometrical differences of the specimens employed in the present work compared to those presented in literature, no direct result comparison can be made.

Comparisons can be made only on results from the same test, as the adopted materials and the sample production technologies have been maintained the same. Moreover, the loading condition has been kept constant. In such conditions, the ANOVA analysis revealed that there is no statistically significant influence of the two alloy forms on the shear bond strength. The mechanical results are consistent with the thickness values of the oxide layers created on the metal surfaces of the samples obtained from both plate and drop form alloy.

It was found that the crack deflects from the joint to the ceramic material due to weakening effect induced by porosities which characterizes the ceramic bulk material (Fig. 5). Significant standard deviation values of the mechanical results may be attributed to slight alteration in the geometric dimensions due to the manual sample preparation and the brittle fracture characteristic which is very sensitive to porosity observed in the ceramic material.

4.2 Metal-ceramic interface and failure mode

Some conclusions about the metal–ceramic adhesion can be drawn by the analysis of the fracture surface. A metal– ceramic system under load will fail at the regions of weakest bonding. The highest bond strength leads to failure within the ceramic; in this case the failure is cohesive. Failures occurring at the interface between metal and oxide layer are adhesive. Depending on the type of fracture, whether it is cohesive or adhesive, the metal–ceramic bonding will be strong or weak, respectively [34]. In this study, according to references [35, 36, 37], the failure types were classified based on the presence of ceramic remnants on the metal substrate after shear tests (i.e. adhesive, cohesive or mixed).

Although adhesion failure is described as a typical failure mechanism of ceramic bonded to noble alloy substrates [38], stereomicroscope observations of the fracture surfaces on the samples analyzed revealed a mixed failure mode. In both cases (plate and drop form alloy), the area of visible metal after the test was lower than 20%; that means a prevalent cohesive failure and thus, a high adhesion strength between metal and ceramic. The difference between plate form and drop form is not significant.

Besides the material chemical compositions, the oxide layer thickness plays an important role in the adhesion between metal and ceramic material. Despite the fact that the average value of the oxide thickness in the samples obtained from plate form alloy is slightly higher than that of the samples obtained from drop form alloy, when making reference to the standard deviation values, it can be concluded that the two kinds of samples have undergone a comparable oxidation process.

4.3 Bulk precious alloy

Despite the same chemical composition, casting products obtained from alloy in plate form showed some solidification defects such as gas porosities. Such defects are linked to the rolling, blanking or continuous annealing used for the production of plates. These processes allow the oily residues, detergents, solvents and oxides to be incorporated and trapped in the micropores present in the metal. The contaminants thus bond indissolubly with the alloy, producing gas porosity in the final casting product. In this case, more expensive and non-conventional castings processes such as vacuum casting process should be used to avoid porosity. On the contrary, specimens obtained with drop form alloy show no solidification defects due to the cleaner drop form alloy production technology. Indeed, the process is contaminant-free, carbon-free and swarf-free. Thus, less or no porosities are generated.

5 Conclusions

Drop and plate alloy are both used in the dental sector but differences between them were not yet analysed in literature. In this paper, a comparison between samples obtained from both plate form and drop form alloy was made in terms of metallurgical and ceramic-alloy adhesion properties. The following conclusions can be drawn:

  • The shear test results show a good repeatability.

  • The ANOVA analysis reveals that there is no statistically significant influence of the two alloy forms on the SBS.

  • Both drop and plate alloy forms show a mixed failure mode, but more than 20% of the surface being covered by ceramic after fracture indicates a good adhesion between metal and ceramic material.

  • Samples from drop form alloy showed less defects than specimens obtained from plate form alloy. As a matter of fact, the production process for the creation of drop alloy is a direct process that involves instantaneous solidification of the molten metal, via the casting, within specific inert liquids. Drop alloy can thus be considered pure and uniform in composition. Furthermore, granulation significantly reduces production costs and consequently the end price.

  • Compared to plate alloy, drop alloy helps to reduce the environmental impact because it never come into contact with toxic substances and reduces casting times and production scrap.

Acknowledgement

The authors would like to thank Giacomo Mazzacavallo for his support during the experimental tests and Legor Group SpA for the materials supplied

References

  • [1]

    S.C. Brecker, “Porcelain baked to gold - a new medium In prosthodontics”, The Journal of prosthetic dentistry, Vol. 6, pp. 801-810, 1956 CrossrefGoogle Scholar

  • [2]

    F. Zarone, S. Russo, R. “Sorrentino From porcelain-fused-to-metal to zirconia: Clinical and experimental considerations”, Dental materials, Vol. 27, pp. 83-96, 2011 CrossrefGoogle Scholar

  • [3]

    N.V. Raptis, K.X. Michalakis, H. Hirayama, “Optical Behavior of Current Ceramic Systems” The International Journal of Periodontics & Restorative Dentistry, Vol. 26, pp. 31-41, 2006. Google Scholar

  • [4]

    B.L. Lawn, Y. Deng, V.P. Thompson, “Use of contact testing in the characterization and design of all-ceramic crown like layer structures: A review”, The Journal of prosthetic dentistry, Vol. 86, pp. 495-510, 2001 CrossrefGoogle Scholar

  • [5]

    A.J. Raigrodski, G.K. Meng, M.B. Hillstead, K.-H. Chung, “Survival and complications of zirconia-based fixed dental prostheses: A systematic review”, The Journal of prosthetic dentistry, Vol. 107, pp. 170-177, 2012 CrossrefGoogle Scholar

  • [6]

    M. Guazzato, M. Albakry, S.P. Ringer, M.V. Swain, “Strength, fracture toughness and microstructure of a selection of all-ceramic materials. Part II. Zirconia-based dental ceramics” Dental materials, Vol. 20, pp. 449-456, 2004 CrossrefGoogle Scholar

  • [7]

    T.E. Donovan, jr. EJS, “Porcelain-Fuse-To-Metal (PFM) alternatives”, Journal of Esthetic and Restorative Dentistry, Vol. 1, pp. 4-6, 2009 Google Scholar

  • [8]

    B.E. Pjetursson, I. Sailer, N.A. Makarov, M. Zwahlen, D.S. Thoma, “All-ceramic or metal-ceramic tooth-supported fixed dental prostheses (FDPs)? A systematic review of the survival and complication rates. Part II: Multiple-unit FDPs”, Dental materials, Vol. 31, pp. 624-639, 2015 Web of ScienceCrossrefGoogle Scholar

  • [9]

    A. Eliasson, C.-F. Arnelund, A. Johansson, “A clinical evaluation of cobalt-chromium metal-ceramic fixed partial dentures and crowns: A three to seven-year retrospective study”, The Journal of prosthetic dentistry, Vol. 98, pp. 6-16, 2007 CrossrefGoogle Scholar

  • [10]

    W.C. Wagner, K. Asgar, W.C. Bigelow, R.A. Flinn, “Effect of interfacial variables on metal-porcelain bonding”, Journal of biomedical materials research, Vol. 27, pp. 531-537, 1993 CrossrefGoogle Scholar

  • [11]

    S.J. Marshalla, S.C. Bayne, R. Baier, A.P. Tomsiac, G.W. Marshalla, “A review of adhesion science”, Dental materials, Vol. 26, pp. e11-e6, 2010 Google Scholar

  • [12]

    K.J. Anusavice, P.H. Dehoff, C.W, Fairhurst, “Materials science: comparative evaluation of ceramic-metal bond tests using finite element stress analysis”, Journal of dental research, Vol. 59, pp. 608-613, 1980 CrossrefGoogle Scholar

  • [13]

    BSI. Dentistry-Compatibility testing Part 1: Metal-ceramic systems. 2012. Google Scholar

  • [14]

    I.A. Hammad, Y.F. Talic, “Design of bond strength tests for metal-ceramic complexes: Review of the literature”, The journal of prosthetic dentistry, Vol. 75, pp. 602-608, 1996 CrossrefGoogle Scholar

  • [15]

    T. Korkmaz, V. Asar, “Comparative evaluation of bond strength of various metal-ceramic restorations”, Materials & Design, Vol. 30, pp. 445-451, 2009 CrossrefWeb of ScienceGoogle Scholar

  • [16]

    S.E. Elsaka, I.M. Hamouda, Y.A. Elewady, O.B. Abouelatta, M.V. Swaind, “Influence of chromium interlayer on the adhesion of porcelain to machined titanium as determined by the strain energy release rate”, Journal of dentistry, Vol. 38, pp. 648-654, 2010 CrossrefWeb of ScienceGoogle Scholar

  • [17]

    B. Henriques, S. Goncalves, D. Soares, F.S. Silva, “Shear bond strength comparison between conventional porcelain fused to metal and new functionally graded dental restorations after thermal-mechanical cycling”, Journal of the mechanical behaviour of biomedical materials, Vol. 13, pp. 194-205, 2012 CrossrefGoogle Scholar

  • [18]

    B. Henriques, D. Soares, F.S. Silva, “Shear bond strength of a hot pressed Au–Pd–Pt alloy–porcelain dental composite”, Journal of the mechanical behaviour of biomedical materials, Vol. 4, pp. 1718-26, 2011 CrossrefGoogle Scholar

  • [19]

    R. Md Melo, A.C. Travassos, M.P. Neisser, “Shear bond strengths of a ceramic system to alternative metal alloys”, The Journal of prosthetic dentistry, Vol. 93, pp. 64-69, 2005 CrossrefGoogle Scholar

  • [20]

    J.S. Shell, J.P. Nielsen, “Study of the Bond between Gold Alloys and Porcelain”, Journal of dental research, Vol. 41, pp. 1424-1437, 1962 CrossrefGoogle Scholar

  • [21]

    A.J.F. Neto, H. Panzeri, F.D. Neves, R.Ad Prado, G. Mendoca, “Bond Strength of Three Dental Porcelains to Ni-Cr and Co-Cr-Ti Alloys”, Brazialian dental journal, Vol. 17, pp. 24-28, 2006 CrossrefGoogle Scholar

  • [22]

    H.W. Roberts, D.W. Berzins, B.K. Moore, D.G. Charlote, “Metal-Ceramic Alloys in Dentistry: A Review”, Journal of prosthodontics, Vol. 18, pp. 188-194, 2009 Web of ScienceCrossrefGoogle Scholar

  • [23]

    H. Knosp, R.J. Holliday, C.W. Corti, “Gold in dentistry: alloys, uses and performance”, Gold bulletin, Vol. 36, pp. 93-102, 2003 CrossrefGoogle Scholar

  • [24]

    R.M. German, “Gold Alloys for Porcelain-Fused-to-Metal Dental Restorations”, Gold bulletin, Vol. 13, pp. 57-62, 1980 CrossrefGoogle Scholar

  • [25]

    J.C. Watada, “Alloys for restoration”, The Journal of prosthetic dentistry, Vol. 87, pp. 351-363, 2002 Google Scholar

  • [26]

    O.L. Bezzon, Md.G.Cd Mattos, R.F. Ribeiro, J.M.D.dA Rollo, “Effect of beryllium on the castability and resistance of ceramometal bonds in nickel-chromium alloys”, The Journal of prosthetic dentistry, Vol. 80, pp. 570-574, 1998 CrossrefGoogle Scholar

  • [27]

    M. Walter, P.D. Reppel, K.B. Ning, W.B. Freesmeyer, “Six-year follow-up of titanium and high-gold porcelain-fused-to-metal fixed partial dentures”, Journal of oral rehability, Vol. 26, pp. 91-96, 1999 CrossrefGoogle Scholar

  • [28]

    L. Probster, U. Maiwald, H. Weber, “Three-point bending strenght of ceramics fused to cast titanium”, European Journal of Oral Sciences, Vol. 104, pp. 313-319, 1996 CrossrefGoogle Scholar

  • [29]

    J.M. Powers, R.L. Sakaguchi, “Restorative materials -metals”, Craig’s restorative dental materials, 13/e: Elsevier India;. pp. 199-250, 2006 Google Scholar

  • [30]

    S.F. Rosenstiel, M.F. Land, J. Fujimoto, Contemporary fixed prosthodontics: Elsevier Health Sciences; 2015. Google Scholar

  • [31]

    Legor. http://products.legor.com/downloadsheet/444/11/IT/-1/-1/none. 2016.

  • [32]

    BSI. ISO system of limits and fits — Part 1: Bases of tolerances, deviations and fits and Part 2: Tables of standard tolerance grades and limit deviations for holes and shafts. 1993. Google Scholar

  • [33]

    K.J. Anusavice, Phillips’ science of dental materials. Philadelphia: W.B Saunders; 2006. Google Scholar

  • [34]

    J.M. Powers, Sakaguchi RL. Craig’s restorative dental materials, 13/e: Elsevier India; 2006. Google Scholar

  • [35]

    J.Gd Santos, R.G. Fonseca, L.G. Adabo, C.A.S. Cruz, “Shear bond strength of metal–ceramic repair systems”, the Journal of prosthetic dentistry, Vol. 96, pp. 165-173, 2006 CrossrefGoogle Scholar

  • [36]

    V.Z.C. Vásqueza, M. Özcan, E.T. Kimpara, “Evaluation of interface characterization and adhesion of glass ceramics to commercially pure titanium and gold alloy after thermal- and mechanical-loading”, Dental materials, Vol. 25, pp. 221-231, 2009 CrossrefWeb of ScienceGoogle Scholar

  • [37]

    T. Akova, Y. Ucar, A. Tukay, M.C. Balkaya, W.A. Brantley, “Comparison of the bond strength of laser-sintered and cast base metal dental alloys to porcelain”, Dental materials, Vol. 24, pp. 1400-1404, 2008 Web of ScienceCrossrefGoogle Scholar

  • [38]

    D.R. Haselton, A.M. Diaz-Arnold, T.James, J. Dunne, “Shearbond strengths of 2 intraoral porcelain repair systems to porcelain or metal substrates”, The Journal of prosthetic dentistry, Vol. 86, pp. 526-531, 2001. CrossrefGoogle Scholar

About the article

Received: 2017-07-05

Accepted: 2017-11-03

Published Online: 2017-12-29


Citation Information: Open Engineering, Volume 7, Issue 1, Pages 394–402, ISSN (Online) 2391-5439, DOI: https://doi.org/10.1515/eng-2017-0043.

Export Citation

© 2017 TPaolo Ferro et al.. This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License. BY-NC-ND 4.0