Calibration and development for increased accuracy of 3D range imaging cameras

Timo Kahlmann / Hilmar Ingensand
Published Online: 2008-06-27 | DOI: https://doi.org/10.1515/JAG.2008.001

Abstract

Range imaging has become a valuable technology for many kinds of applications in recent years. Numerous systematic deviations occur during the measurement process carried out by available systems. These systematics are partly excited by external and partly excited by internal influences. In this paper the following investigations will be presented in closer detail. First the statistics of the distance measurement of the analyzed range imaging cameras SwissRangerTM SR-2 and SR-3000 will be shown. Besides the question if the measurements are Gaussian distributed, the precision of the measurements will be shown. This aspect is of importance to answer the question if the mean value of a series of measurements leads to more precise data. Second diverse influencing parameters like the target's reflectivity and external as well as internal temperature are aimed. The dependency of the distance measurements with respect to the amplitude is one of the main aspects in this paper. A specialized set up has been developed in order to derive experimentally the detailed correlation, which is expressed in terms of linearity deviations.

Besides the results of some specific aspects, an overview of the recommended calibration procedure is given. The reader of this paper will be enabled to understand the calibration steps needed to gain highly accurate data from the investigated range imaging cameras. Due to the fact that range imaging cameras are on their way to become state of the art in 3D capturing of the environment, it is of importance to develop strategies for the calibration of such sensors in order to enable users to revert to these principles for the sake of simplicity. Therefore, these strategies long for sophisticated approaches and reliable results of investigations. This paper will introduce such an approach to be discussed within the scientific and user environment.

One of the main achievements of this work is the introduction of a method to significantly decrease the influence of temperature on the distance measurements by means of a differential measurement principle setup. The verification of the functionality is presented, as well.

Keywords.: Range imaging; 3D camera; calibration; flash ladar; SwissRanger; PMD; time-of-flight

About the article

Author information: Timo Kahlmann and Hilmar Ingensand, Institute of Geodesy and Photogrammetry, ETH Zurich, Switzerland. E-mail: , E-mail:


Received: 2007-09-25

Accepted: 2007-11-17

Published Online: 2008-06-27

Published in Print: 2008-04-01


Citation Information: Journal of Applied Geodesy, Volume 2, Issue 1, Pages 1–11, ISSN (Online) 1862-9024, ISSN (Print) 1862-9016, DOI: https://doi.org/10.1515/JAG.2008.001.

Export Citation

Citing Articles

Here you can find all Crossref-listed publications in which this article is cited. If you would like to receive automatic email messages as soon as this article is cited in other publications, simply activate the “Citation Alert” on the top of this page.

[1]
Gianluca Agresti, Ludovico Minto, Giulio Marin, and Pietro Zanuttigh
Information Fusion, 2018
[2]
Ying He, Bin Liang, Yu Zou, Jin He, and Jun Yang
Sensors, 2017, Volume 17, Number 1, Page 92
[3]
Derek D. Lichti and Changjae Kim
Remote Sensing, 2011, Volume 3, Number 12, Page 1014
[4]
Hongshan Yu, Jiang Zhu, Yaonan Wang, Wenyan Jia, Mingui Sun, and Yandong Tang
Sensors, 2014, Volume 14, Number 6, Page 10753
[5]
Carlo Dal Mutto, Pietro Zanuttigh, and Guido Maria Cortelazzo
IEEE Transactions on Pattern Analysis and Machine Intelligence, 2015, Volume 37, Number 11, Page 2260
[6]
Sergi Foix, Guillem Alenya, and Carme Torras
IEEE Sensors Journal, 2011, Volume 11, Number 9, Page 1917
[7]
Kourosh Khoshelham and Sander Oude Elberink
Sensors, 2012, Volume 12, Number 12, Page 1437
[8]
Derek D. Lichti, Jacky C. K. Chow, Edson Mitishita, Jorge Antonio Silva Centeno, Felipe Martins Marques da Silva, Roberto Arocha Barrios, and Ilich Contreras
Journal of Surveying Engineering, 2014, Volume 140, Number 2, Page 04014003
[9]
Valeria Garro, Carlo Dal Mutto, Pietro Zanuttigh, and Guido M. Cortelazzo
annals of telecommunications - annales des télécommunications, 2013, Volume 68, Number 11-12, Page 597
[10]
Sven Mersmann, Alexander Seitel, Michael Erz, Bernd Jähne, Felix Nickel, Markus Mieth, Arianeb Mehrabi, and Lena Maier-Hein
Medical Physics, 2013, Volume 40, Number 8, Page 082701
[11]
Derek D. Lichti and Xiaojuan Qi
Journal of Spatial Science, 2012, Volume 57, Number 2, Page 247
[12]
Derek D. Lichti, Xiaojuan Qi, and Tanvir Ahmed
ISPRS Journal of Photogrammetry and Remote Sensing, 2012, Volume 74, Page 101
[13]
Peter Dorninger, Clemens Nothegger, Norbert Pfeifer, and Gábor Molnár
Journal of Applied Geodesy, 2008, Volume 2, Number 4

Comments (0)

Please log in or register to comment.
Log in