Skip to content
Publicly Available Published by De Gruyter July 27, 2016

New Approaches to Wipe Sampling Methods for Antineoplastic and Other Hazardous Drugs in Healthcare Settings

  • Thomas H. Connor

    Dr. Connor received his doctorate from the University of Texas Medical Branch and was on the faculty of the University of Texas, School of Public Health. He is currently a Research Biologist in the Division of Applied Research and Technology at the National Institute for Occupational Safety and Health. He developed the NIOSH Alert on Hazardous Drugs. He was awarded the American Society of Health-System Pharmacists’ Board of Directors’ Award and the International Society of Oncology Pharmacy Practitioners’ Achievement Award. His research has focused on occupational exposure to hazardous drugs and he has published and lectured extensively on this topic.

    EMAIL logo
    and Jerome P. Smith

    Jerome P Smith is a research chemist in the Division of Applied Research and Technology at the National Institute for Occupational Safety and Health (NIOSH). He has developed immunoassays employing fluorescent covalent microsphere immunoassay for multiple analytes including drugs of abuse and antineoplastic drugs, enzyme-linked immunosorbent assay for anthrax antibodies, lateral flow immunoassay for onsite detection of drugs of abuse and antineoplastic drugs, and surface plasmon resonance for methamphetamine residues. He has a B.A. in Chemistry from the College of St. Thomas in St. Paul, Minnesota and a Ph.D. in analytical chemistry from the University of Colorado, Boulder.

Abstract

Purpose: At the present time, the method of choice to determine surface contamination of the workplace with antineoplastic and other hazardous drugs is surface wipe sampling and subsequent sample analysis with various analytical techniques. The purpose of this article is to review current methodology for determining the level of surface contamination with hazardous drugs in healthcare settings and to discuss recent advances in this area. In addition it will provide some guidance for conducting surface wipe sampling and sample analysis for these drugs in healthcare settings.

Methods: Published studies on the use of wipe sampling to measure hazardous chemicals, including antineoplastic drugs on surfaces were reviewed. These studies include the use of well-documented chromatographic techniques for sample analysis in addition to newly evolving technology that provides rapid analysis of specific antineoplastic drugs.

Results: Methodology for the analysis of surface wipe samples for hazardous drugs are reviewed, including the purposes, technical factors, sampling strategy, materials required, and limitations. The use of lateral flow immunoassay (LFIA) and fluorescence covalent microbead immunosorbent assay (FCMIA) for surface wipe sample evaluation is also discussed.

Conclusions: Current recommendations are that all healthcare settings where antineoplastic and other hazardous drugs are handled include surface wipe sampling as part of a comprehensive hazardous drug-safe handling program. Surface wipe sampling may be used as a method to characterize potential occupational dermal exposure risk and to evaluate the effectiveness of implemented controls and the overall safety program. New technology, although currently limited in scope, may make wipe sampling for hazardous drugs more routine, less costly, and provide a shorter response time than classical analytical techniques now in use.

Introduction

Occupational exposure to antineoplastic and other hazardous drugs is a concern for healthcare facilities in the U.S. and worldwide. Workplace contamination with these drugs, especially antineoplastic drugs, is a continuing issue in places where these drugs are prepared and administered to patients [1, 2]. At the present time, surface wipe sampling is used to determine the level of workplace contamination in healthcare settings where antineoplastic and other hazardous drugs are present [3]. Wipe sampling can help identify the need for, and effectiveness of engineering controls, the effect of improved work practice controls, and what type of personal protective equipment (PPE) is required [4].

In healthcare settings, dermal uptake is considered the most likely route of occupational exposure to most hazardous drugs, especially low-molecular-weight antineoplastic drugs. Wipe sampling is the most appropriate methodology available to evaluate potential worker exposure to these drugs [58]). Inhalation of aerosolized droplets or vapors, accidental hand-to-mouth ingestion following contact with contaminated surfaces, and needles or other sharps are also possible routes of exposure [9]). As stated by OSHA, in some cases, skin absorption may be a more important route of exposure than inhalation, especially for non-volatile hazardous chemicals which remain on work surfaces for long periods of time and may not be noticed by the employees (https://www.osha.gov/dts/chemicalsampling/data/CH_235600.html). Although not a direct measurement of exposure, wipe samples can provide information about contamination on surfaces which may lead to dermal uptake by workers.

In general, it is assumed that dermal absorption is more likely to occur with drugs with a molecular weight of <500 Daltons and less likely for those >1,000 Daltons [10, 11]. In addition, lipid-soluble compounds more readily penetrate the skin than water-soluble ones and uptake may be enhanced by the use of carrier solvents. Although many antineoplastic drugs are relatively small molecules with molecular weights <500 Daltons, newer antineoplastic drugs, such as monoclonal antibodies, can have a molecular weight >40,000 Daltons, which presumably limit their potential for dermal uptake from contaminated surfaces [1214]. N, N-Dimethylacetamide and other solvents used in some drugs may pose an additional health risk and/or may facilitate drug penetration of the skin [15].

McDevitt and colleagues at Johns Hopkins Hospital published the first study of surface contamination with antineoplastic drugs in the U.S [16]. In that study, cyclophosphamide contamination was demonstrated in both the pharmacy and patient treatment areas. In the mid-1990s, several studies conducted in the Netherlands were published by Sessink and colleagues [1720], which measured surface concentrations of several antineoplastic drugs by wipe sampling, including: cyclophosphamide, ifosfamide, methotrexate, and 5-fluorouracil. As in the U.S. study, surface contamination with these drugs was demonstrated in areas where they were prepared and administered to patients. The studies by Sessink also documented uptake of the drugs based on their measurement in the urine of healthcare workers. Connor et al. [21] measured surface contamination in three cancer hospitals in the United States and three in Canada. Their study documented surface contamination with cyclophosphamide, ifosfamide, and 5-fluorouracil in the pharmacy and patient treatment areas, and also in areas adjacent to the pharmacy. Since publication of these initial studies, world-wide reports have been published, demonstrating a universal issue with workplace surface contamination with antineoplastic drugs and uptake of the drugs by healthcare workers. In addition to surface wipe sampling, there have also been a limited number of dermal sampling studies reported in the literature (http://www.cdc.gov/niosh/topics/antineoplastic/). For this review, methodology used for wipe sampling is discussed in order to describe a framework for wipe sampling for hazardous drugs in healthcare settings.

Although surface wipe sampling is the method of choice to evaluate surface contamination within healthcare settings for antineoplastic drugs, the analytical methodology used for surface wipe sampling and analytical laboratory analysis for antineoplastic drugs has varied considerably [3, 22]. This article reviews methodology currently being used for surface wipe sampling for hazardous drugs and discusses new applications of available technology that can be employed to make wipe sampling less expensive and to provide near real-time results for wipe sampling.

Surface wipe sampling methods

Methods for surface wipe sampling have been developed for a number of hazardous agents, such as lead [23], asbestos, pesticides, polychlorinated biphenyls [24, 25], methamphetamine [26], antibiotics [27]), and similar methodology has been applied to sampling for hazardous drugs, primarily antineoplastic drugs [3]. Although there are many variations in application, some approaches are common to all methods. Once a sampling scheme has been decided upon, it is necessary to determine if methods are available for the drugs of concern. Several laboratories offer analytical services for these drugs and often provide sampling kits for a battery of the more commonly used drugs. Sampling locations tend to be agreed upon, but the surface areas to be sampled can vary considerably, from a “standard” of 100 cm2 to 500 cm2 or even larger [3]. Typically, a commercially available template is used to delineate the wiping area. A suitable solvent is either applied directly to the surface to be sampled or applied to the sampling material. Sampling materials vary among tissues, filter paper, or special sampling swabs, although other sampling media may also work well. The surface is then wiped in a pre-determined pattern, usually in one direction and then perpendicular to that direction. The wiping process may then be repeated for the same location using a second wipe to ensure better recovery [28]. The wipes are then placed in labelled containers and shipped to the analytical laboratory for analysis. In some case, the samples may have to be shipped on dry ice to maintain drug stability.

Current analytical methods for antineoplastic drugs

A comprehensive review of analytical methods by Turci et al. [22] describes some of the analytical methods that have been employed. These have included gas chromatography (GC), high-performance liquid chromatography (HPLC), and ultra high-performance liquid chromatography (UPLC); all in combination with mass spectrometry (MS) or tandem mass spectrometry (MS/MS). Inductively coupled plasma mass spectrometry (ICP-MS) has been used for the analysis of total platinum in platinum-based drugs [22]. HPLC with mass spectrometry (LC-MS/MS) is the current method of choice for analysis of most antineoplastic drugs and does not require derivatization of the drug, as GC methods do [17]. Researchers using LC-MS/MS are reporting results in the pg/cm2 range. However, these methods are often very costly and it may take several weeks to obtain results using them. In addition, the number analytical laboratories in the U.S. that can perform these analyses is limited.

Analytical techniques for measurement of surface contamination by antineoplastic drugs such as HPLC–MS/MS are sensitive, specific, and accurate, but the initial equipment investment is high and a trained analytical chemist is required to operate the instrumentation, so these methods often cannot be used on a frequent basis due to cost [29]. Also, results from LC-MS/MS analyses are often obtained after a long delay, which may result in contamination existing for a long period of time before clean up can take place. Hence, these methods often cannot provide immediate feedback in developing work practices needed to lower exposure due to the time period between sample collection and obtaining the results of the analysis. The immunochemical techniques described below provide an opportunity for more cost-effective routine measurements and can provide direct reading near real-time on-site measurements.

Immunochemical technique for simultaneous detection of drug surface contamination by multiple drugs

A fluorescence covalent microbead immunosorbent assay (FCMIA) for performing multiple immunochemical determinations simultaneously has been developed by Luminex Corporation [30, 31]. FCMIA combines several classical methodologies: immunoassays, microspheres, and flow cytometry technology. In FCMIA, immunoassays are performed on sets of solid support 5.6 µm microspheres internal fluophores with different characteristic that allow multiple assays to be performed simultaneously (multiplexing). FCMIA has predominantly been used for multiple protein and nucleic acid analytes such as multiple antibodies in serum [32] multiple cytokines in serum [33], and multiple RNA and DNA viruses in patient samples [34]. FCMIA assays are simple to set up and the instrumentation requires limited training to operate. NIOSH has developed an assay and sampling technique capable of evaluating multiple drugs of abuse on surfaces using competitive FCMIA [35], which has been used for exposure assessments in law enforcement evidence vaults where relatively high surface levels were observed in the studies [36, 37].

A FCMIA has been demonstrated for use in the simultaneous measurement of multiple antineoplastic drugs [38]. The antineoplastic drug assay uses a competitive assay format similar to the drugs of abuse FCMIA assay, where a drug in solution competes with a microsphere bound drug–BSA conjugate for an anti-drug antibody. This results in less antidrug antibody being bound to the microsphere at higher drug concentrations. The anti-drug antibody bound to the microsphere is detected with a labeled secondary antibody (biotin labeled anti-mouse IgG, (Pierce Biotechnology, Inc., Rockford, IL), which in turn binds a fluorescent label (streptavidin R-PE, Molecular Probes, Eugene, OR). Thus, the streptavidin R-PE fluorescent signal from the microsphere decreases with increasing drug concentration [38]. The surface is sampled using a swab wetted in wash buffer (PBS-0.1 % tween). The sampling is done by carefully wiping the surface with the wetted swab in one direction with an overlapping pattern, then repeating the same wiping pattern in a direction perpendicular to the first direction, and finally repeating the original wiping pattern. The swab is then placed in a vial containing 1 ml of storage/blocking buffer (PBS-1 %BSA) and the swab is extracted with vigorous shaking for 2 min. The resulting solution is run in the assay without any further dilution.

In laboratory-based evaluations, there was no significant cross-reactivity between these drugs at the ranges studied, indicated by a lack of response in the assay to potentially competing analytes [38]. The limit of detection (LOD) for the three drugs are listed in Table 1. The main use of this assay would be for screening surfaces in pharmacies, patient treatment areas, and other areas in healthcare facilities for contamination by antineoplastic drugs. Of special interest are pharmacy areas such as biological safety cabinets (BSCs) or compounding aseptic containment isolators (CACIs) used for drug preparation as well as nursing areas where drugs are administered to patients. Instrumental methods are more sensitive and specific than this assay for several of the drugs (e. g. 5-FU), but this assay is simple and could be set up to be performed at the worksite to provide more timely results. The components of the assay could be supplied as a kit that requires limited training to use. Use of this assay may provide a lower cost method for routine application.

Table 1:

LOD and LOQ for FCMIA and LFIA compared to LC-MS/MS.

DrugLOD FCMIA (ng/cm2)LOQ FCMIA (ng/cm2)LOD LFIA (ng/cm2)Range LFIA (ng/cm2)LOD LC-MS/MS (ng/cm2)**
5-Fluorouracil0.932.800.10-10.06
Paclitaxel0.572.061NM*0.07
Doxorubicin0.00360.130.10–0.250.20

It would be desirable for the FCMIA assay to screen for a contamination of a wider array of drugs. This would require the development of antibodies and drug–protein conjugates for the additional drugs, which is a limitation of the assay. Also, cross-reactivity between the drugs being measured, as well as other drugs that might be present, would have to be assessed. The addition of other drugs would result in minimal increases in the time to perform the assay and the FCMIA assay could be used for more drugs at a relatively low cost.

Onsite detection techniques for drug surface contamination based on lateral flow immunoassay

Direct reading, field portable monitors to measure antineoplastic drugs on surfaces also can be used to help reduce potential exposure of healthcare workers to these drugs. Ideally, monitors need to be portable, near real time, sensitive, and easy to use. Monitors based on lateral flow immune assay (LFIA) fulfill the requirements of portability, sensitivity, rapid response, and ease of use. Lateral flow assays are used in many consumer products such as pregnancy tests and are being used in many point-of-care assays for clinical use [39]. The lateral flow assay cassettes typically have two lines, a test line which varies in intensity with the concentration of analyte and a control line which is relatively constant for all samples and has been mainly used to assure that the cassette is working properly [40].

The development of LFIAs for antineoplastic drugs was based on prior work done by NIOSH to detect methamphetamine contamination on surfaces [41]. The methamphetamine lateral flow cassette along with components for the sampling method is commercially available as a kit from SKC Inc. (MethChek 50, SKC Inc., Eighty Four, PA), which uses complete disappearance of the test line as the end point for detection. The methamphetamine kit has been used extensively by first responders such as police to evaluate contamination in sites where methamphetamine labs have been located [42]. In a further study of the methamphetamine lateral flow cassettes, comparison of the control line and test line was used as a visual end point and the methamphetamine cassettes were read with an electronic lateral flow reader [43]. Use of line comparison or an electronic reader results in a more sensitive measurement as well as semi-quantitative evaluation of contamination.

Based on success of the lateral flow methamphetamine detection kit, NIOSH investigated the development of lateral assays for antineoplastic drugs. The first assay developed was for 5-FU [44]. The wiping pattern was identical to that used in FCMIA. A cotton swab wetted in sampling buffer (in this case PBS- 1 % tween) was used and the swab was extracted in 1 ml sampling buffer. For the 5-FU cassette, line comparison was used as the visual endpoint (Figure 1) since complete disappearance of the test line would have resulted in an end point that was less sensitive than desired. The cassette was also evaluated using the electronic lateral flow reader. The 5-FU assay was evaluated by spiking ceramic, vinyl, composite, stainless steel, and glass surfaces of 100 cm2 area with 5-FU masses of 0, 5, 10, 25, 50, and 100 ng. The 5-FU cassette is capable of detecting 10 ng/100 cm2 (0.1 ng/cm2) using the electronic reader and 25 ng/100 cm2 (0.25 ng/cm2) using the visual line comparison method for the surfaces studied. Using measurements from the electronic lateral flow reader, the response from 0 to 100 ng/cm2 was fitted using the ratio of the control to the test line such that the cassette response could be used to predict the loading over this concentration range (Table 1). The response of the cassettes was compared to LC-MS/MS results for the same samples for validation and there was good correlation of the two methods, but the slope of the plots of the recovered mass from the 5-FU lateral flow measurement versus that for the LC-MS/MS varied from about 0.43 to 1.07 depending on the surface studied and the mass loading on the surface.

Figure 1: 5-Fluorouracil lateral flow assay cassettes with varying relative test line and control line intensities.
Figure 1:

5-Fluorouracil lateral flow assay cassettes with varying relative test line and control line intensities.

In addition to 5-FU, lateral flow assays have been developed for paclitaxel and doxorubicin. The paclitaxel assay detected 50–100 ng/100 cm2, so it was less sensitive than desired but could be used at higher levels of contamination. The doxorubicin assay detected as low as 10 ng/100 cm2 but had limited range of response, low line intensity, and longer development time. Our commercial partner has produced an improved doxorubicin antibody and doxorubicin protein conjugate in order to develop an improved assay which is now being tested.

The 5-FU lateral flow assay has been used to evaluate contamination in a limited number of hospital pharmacy and patient treatment areas in a pilot study and has agreed with measurement from LC-MS/MS for most of the samples taken (Smith et al. unpublished data). Ongoing evaluations to provide data for a more comprehensive comparison of the lateral flow assay with LC-MS/MS are underway. A large multinational medical equipment manufacturer has shown interest in the commercial development of the lateral flow assay and has developed an improved assay for doxorubicin which uses a portable reader that the manufacturer developed.

Future perspectives

Although the methodology for surface sampling is currently only available for a limited number of antineoplastic drugs, methods can be developed for other drugs depending on the availability of antibodies to them. Both LFIA and FCMIA have proven effective in other uses and settings [32, 35, 44] and their applicability to measuring surface contamination with antineoplastic drugs at low levels has been demonstrated [38, 43].

Conclusions

Surface wipe sampling is currently used to evaluate workplace contamination with antineoplastic drugs in healthcare and other settings where the drugs may be present. Using LC-MS/MS, surface levels of a few pg/cm2 can be measured. However, this method is costly and it may take a few weeks for the results to be reported to the facility, thus delaying necessary cleaning procedures and changes in work practices and engineering controls. In addition, most healthcare facilities do not have the capability to perform these assays in-house and must outsource the analyses. A simple, sensitive near real-time method to measure these drugs on work surfaces would enable healthcare workers and health and safety personnel to perform these measurements and obtain immediate results. Therefore, the results of routine cleaning or for cleaning a spill could be obtained within minutes instead of days or weeks.

USP General Chapter <800> recommends performing baseline wipe sampling for hazardous drugs and periodically afterwards (every six months or as often as needed) [45]. A battery of a relatively small number of available assays that could provide near real-time values for surface contamination would be beneficial to many healthcare other facilities where hazardous drugs are present.

Disclaimers

The findings and conclusions in this report are those of the author and do not necessarily represent the views of the National Institute for Occupational Safety and Health (NIOSH). Mention of any company or product does not constitute endorsement by NIOSH. In addition, citations to Web sites external to NIOSH do not constitute NIOSH endorsement of the sponsoring organizations or their programs or products. Furthermore, NIOSH is not responsible for the content of these Web sites. All Web addresses referenced in this document were accessible as of the publication date.

About the authors

Thomas H. Connor

Dr. Connor received his doctorate from the University of Texas Medical Branch and was on the faculty of the University of Texas, School of Public Health. He is currently a Research Biologist in the Division of Applied Research and Technology at the National Institute for Occupational Safety and Health. He developed the NIOSH Alert on Hazardous Drugs. He was awarded the American Society of Health-System Pharmacists’ Board of Directors’ Award and the International Society of Oncology Pharmacy Practitioners’ Achievement Award. His research has focused on occupational exposure to hazardous drugs and he has published and lectured extensively on this topic.

Jerome P. Smith

Jerome P Smith is a research chemist in the Division of Applied Research and Technology at the National Institute for Occupational Safety and Health (NIOSH). He has developed immunoassays employing fluorescent covalent microsphere immunoassay for multiple analytes including drugs of abuse and antineoplastic drugs, enzyme-linked immunosorbent assay for anthrax antibodies, lateral flow immunoassay for onsite detection of drugs of abuse and antineoplastic drugs, and surface plasmon resonance for methamphetamine residues. He has a B.A. in Chemistry from the College of St. Thomas in St. Paul, Minnesota and a Ph.D. in analytical chemistry from the University of Colorado, Boulder.

  1. Conflict of interest statement: Authors state no conflict of interest. All authors have read the journal’s Publication ethics and publication malpractice statement available at the journal’s website and hereby confirm that they comply with all its parts applicable to the present scientific work.

References

1. Davis J, McLauchlan R, Connor TH. Exposure to hazardous drugs in healthcare: an issue that will not go away. J Oncol Pharm Pract 2011;17:9–13.10.1177/1078155210388462Search in Google Scholar

2. NIOSH. NIOSH list of antineoplastic and other hazardous drugs in healthcare settings. Cincinnati, OH: U.S. Department of Health and Human Services, Centers for Disease Control and Prevention, National Institute for Occupational Safety and Health, DHHS (NIOSH) Publication No. 2014–138, 2014. Available at: www.cdc.gov/niosh/docs/2014-138/.Search in Google Scholar

3. Connor TH, Zock MD, Snow AH. Surface wipe sampling for antineoplastic (chemotherapy) and other hazardous drug residue in healthcare settings: Methodology and recommendations. J Occup Environ Hyg 2016. doi:10.1080/15459624.2016.1165912.Search in Google Scholar

4. Ashley K, Brisson MJ, White KT. Review of standards for surface and dermal sampling. J ASTM Int 2011;8(6). DOI: 10.1520/JAI103678.Search in Google Scholar

5. Fransman W, Vermeulen R, Kromhout H. Occupational dermal exposure to cyclophosphamide in Dutch hospitals: a pilot study. Ann Occup Hyg 2004;48:237–44.Search in Google Scholar

6. Fransman W, Vermeulen R, Kromhout H. Dermal exposure to cyclophosphamide in hospitals during preparation, nursing and cleaning activities. Int Arch Occup Health 2005;78:403–12.10.1007/s00420-004-0595-1Search in Google Scholar

7. Hon CY, Teschke K, Demers PA, Venners S. Antineoplastic drug contamination on the hands of employees working throughout the hospital medication system. Ann Occup Hyg 2014;58:761–70.Search in Google Scholar

8. Kromhout H, Hoek F, Uitterhoeve R, Huijbers R, Overmars RF, Anzion R, et al. Postulating a dermal pathway for exposure to antineoplastic drugs among hospital workers: applying a conceptual model to the results of three workplace surveys. Ann Occup Hyg 2000;44:551–60.10.1016/S0003-4878(00)00050-8Search in Google Scholar

9. NIOSH. Personal Protective Equipment for Health Care Workers Who Work with Hazardous Drugs. DHHS (NIOSH) Publication No. 2009–106, 2008. Available at: http://www.cdc.gov/niosh/topics/antineoplastic/nioshpubs.html.Search in Google Scholar

10. Bos JD, Meinardi MM. The 500 Dalton rule for the skin penetration of chemical compounds and drugs. Exp Dermatol 2000;9:165–9.10.1034/j.1600-0625.2000.009003165.xSearch in Google Scholar PubMed

11. Kimmel TA, Sussman RG, Ku R, Adar AW. Developing acceptable surface limits for occupational exposure to pharmaceutical substances. J ASTM Int 2011;8(8). doi:10.1520JAI103480.Search in Google Scholar

12. Alexander M, King J, Bajel A, Doecke C, Fox P, Lingaratnam S, et al. Australian consensus guidelines for the safe handling of monoclonal antibodies for cancer treatment by healthcare personnel. Int Med J 2014;44:1018–26.10.1111/imj.12564Search in Google Scholar PubMed

13. Connor TH, MacKenzie BM. Should monoclonal antibodies and their conjugates be considered occupational hazards. Saf Considerations Oncol Pharm (Special edition) Fall 2011. Available at: www.ppme.eu.Search in Google Scholar

14. King J, Alexander M, Byrne J, MacMillan K, Mollo A, Kirsa S, et al. A review of the evidence for occupational exposure risks to novel anticancer agents – a focus on monoclonal antibodies. J Oncol Pharm Pract 2016;22:121–34.10.1177/1078155214550729Search in Google Scholar PubMed

15. Munro DD, Stoughton RB. Dimetylacetamide (DMAC) and dimethylformamide (DMFA) effect on percutaneous absorption. Arch Dermatol 1965;92:585–6.10.1001/archderm.1965.01600170101020Search in Google Scholar

16. McDevitt JJ, Lees PS, McDiarmid MA. Exposure of hospital pharmacists and nurses to antineoplastic agents. J Occup Med 1993;35:57–60.Search in Google Scholar

17. Sessink PJ, Anzion RB, Van den Broek PH, Bos RP. Detection of contamination with antineoplastic agents in a hospital pharmacy department. Pharm Weekbl Sci 1992;14:16–22.10.1007/BF01989220Search in Google Scholar

18. Sessink PJ, Friemèl NS, Anzion RB, Bos RP. Biological and environmental monitoring of occupational exposure of pharmaceutical plant workers to methotrexate. Int Arch Occup Environ Health 1994a;65:401–3.10.1007/BF00383251Search in Google Scholar

19. Sessink PJ, Van de Kerkhof MC, Anzion RB, Noordhoek J, Bos RP. Environmental contamination and assessment of exposure to antineoplastic agents by determination of cyclophosphamide in urine of exposed pharmacy technicians: Is skin absorption an important exposure route? Arch Environ Health 1994b;49:165–9.10.1080/00039896.1994.9940377Search in Google Scholar

20. Sessink PJ, Wittenhorst BC, Anzion RB, Rob RP. Exposure of pharmacy technicians to antineoplastic agents: Reevaluation after additional protective measures. Arch Environ Health 1997;52:240–4.10.1080/00039899709602893Search in Google Scholar

21. Connor TH, Anderson RW, Sessink PJ, Broadfield L, Power LA. Surface contamination with antineoplastic agents in six cancer treatment centers in the United States and Canada. Am J Health-Syst Pharm 1999;56:1427–32.10.1093/ajhp/56.14.1427Search in Google Scholar

22. Turci R, Sottani C, Spagnoli G, Minoia C. Biological and environmental monitoring of hospital personnel exposed to antineoplastic agents: a review of analytical methods. J Chromatogr B Analyt Technol Biomed Life Sci 2003;789:169–209.10.1016/S1570-0232(03)00100-4Search in Google Scholar

23. Brookhaven National Laboratory: IH75190: surface wipe sampling procedure. Upton, NY: Brookhaven National Laboratory, Safety & Health Services Division, Industrial Hygiene Group, 2014.Search in Google Scholar

24. ASTM: D6661-10: Standard practice for field collection of organic compounds from surfaces using wipe sampling. West Conshohocken, PA: ASTM International, 2010.Search in Google Scholar

25. EPA: a literature review of wipe sampling methods for chemical warfare agents and toxic industrial chemicals. EPA/600/R-07/004. Washington, DC: U.S. Environmental Protection Agency, Office of Research and Development, 2007.Search in Google Scholar

26. NIOSH. Methamphetamine on wipes by liquid chromatography/mass spectrometry: Method 9111. In: Ashley KA, O’Connor PF, editors. NIOSH manual of analytical methods, 5th ed. Cincinnati, OH: U.S. Department of Health and Human Services, Centers for Disease Control and Prevention, National Institute for Occupational Safety and Health, 2011. Available at: www.cdc.gov/niosh/nmam/.Search in Google Scholar

27. Nygren O, Lindahl R. Development of a method for screening spill and leakage of antibiotics on surfaces based on wipe sampling and HPLC-MS/MS analysis. J ASTM Int 2011;8(6). DOI: 10.1520/JAI103678.Search in Google Scholar

28. B’Hymer C, Connor T, Stinson D, Pretty J. Validation of an HPLC-MS/MS and wipe procedure for mitomycin C contamination. J Chromatogr Sci 2015;53:619–24.10.1093/chromsci/bmu095Search in Google Scholar PubMed PubMed Central

29. Pretty JR, Connor TH, Spasojevic I, Kurtz KS, McLaurin JL, B’Hymer C, et al. Sampling and mass spectrometric analytical methods for five antineoplastic drugs in the healthcare environment. J Oncol Pharm Pract 2012;18:23–36.10.1177/1078155210389215Search in Google Scholar PubMed PubMed Central

30. Fulton RJ, McDade RL, Smith PL, Kienker LJ, Kettman JR. Advanced multiplexed analysis with the FlowMetrixTM system. Clin Chem 1997;43:1749–56.10.1093/clinchem/43.9.1749Search in Google Scholar

31. Oliver K, Kettman J, Fulton R. Multiplexed analysis of human cytokines by use of the FlowMetrix System. Clin Chem 1998;44:2057–60.10.1093/clinchem/44.9.2057Search in Google Scholar

32. Biagini RE, Schlottmann SA, Sammons DL, Smith JP, Snawder JC, Striley CA. Method for simultaneous measurement of antibodies to 23 pneumococcal capsular polysaccharides. Clin Diagn Lab Immunol 2003;10:744–50.10.1128/CDLI.10.5.744-750.2003Search in Google Scholar PubMed PubMed Central

33. Bower M, Veraitch O, SZYdlo R, Charles P, Kelleher CP, Gazzard B. Changes during rituximab therapy in HIV-associated multicentric Castleman disease. Blood 2009;113:4521–4.10.1182/blood-2008-12-197053Search in Google Scholar PubMed

34. Ginocchio CC, Zhang F, Manji R, Arora S, Bornfreund M, Falk M. Evaluation of multiple test methods for the detection of the novel 2009 influenza A (H1N1) during the New York City outbreak. J Clin Virol 2009;45:191–5.10.1016/j.jcv.2009.06.005Search in Google Scholar PubMed PubMed Central

35. Smith JP, Sammons DL, Robertson SA, Biagini RE, Snawder JE. Measurement of multiple drugs in urine, water, and on surfaces using fluorescence covalent microbead immunosorbent assay. Toxicol Mech Methods 2010;20:587–93.10.3109/15376516.2010.518172Search in Google Scholar PubMed

36. Fent KW, Durgan S, Gibbins JD, Smith J, Niemeier MT. Police officers’ chemical exposures in a drug vault. Evid Technol Mag 2012 Jan-Feb;10:16–21.Search in Google Scholar

37. King BS, Musolin K, Choi J. Evaluation of potential employee exposures during crime and death investigations at a county coroner’s office. CDC/NIOSH HHE Report No. 2011-0146-3170, 2013 Mar.Search in Google Scholar

38. Smith JP, Sammons DL, Robertson SA, Pretty JR, DeBord DG, Connor TH, et al. Detection and measurement of surface contamination by multiple antineoplastic drugs using multiplex bead assay. J Oncol Pharm Pract 2016;22:60–7.10.1177/1078155214554407Search in Google Scholar PubMed PubMed Central

39. marketsandmarkets.com, Lateral flow assay market by product (reader, kits) application (clinical testing (pregnancy, infectious disease, cholesterol, cardiac marker), veterinary, drug development) technique (sandwich, competitive, multiplex) end user - global forecast to 2020, Report Code: MD 4158, 2016 Mar.Search in Google Scholar

40. Sajid M, Kawde A, Daud M. Designs formats and applications of lateral flow assay: a literature review. J Saudi Chem Soc 2015;19:689–705.10.1016/j.jscs.2014.09.001Search in Google Scholar

41. Snawder JE, Striley CA, Esswein EJ, Hessel J, Sammons DL, Robertson SA. Use of direct reading surface sampling methods for site characterization and remediation of methamphetamine contaminated properties. J ASTM Int 2011;8:297–312.10.1520/STP49758SSearch in Google Scholar

42. Hudson H, Snawder J, Esswein SC. Partnering and consumer orientation: techniques that move occupational safety and health research into practice. Soc Mark Q 2008;14 (Special Issue):99–104.10.1080/15245000802487564Search in Google Scholar

43. Smith JP, Sammons DL, Pretty JR, Kurtz KS, Robertson SA, DeBord DG, et al. Detection of 5-fluorouracil surface contamination in near real time. J Oncol Pharm Pract 2015;22:396–408.10.1177/1078155215585187Search in Google Scholar PubMed PubMed Central

44. Smith JP, Sammons DL, Robertson SA, Snawder JE. Enhanced performance of methamphetamine lateral flow cassettes using an electronic lateral flow reader. J Occup Environ Hyg 2015;12:45–50.10.1080/15459624.2014.935782Search in Google Scholar PubMed PubMed Central

45. U.S. Pharmacopeia (USP). Chapter: hazardous drugs – handling in healthcare settings. Rockville, MD: U.S. Pharmacopeia, 2016.Search in Google Scholar

Received: 2016-5-17
Revised: 2016-6-23
Accepted: 2016-6-23
Published Online: 2016-7-27
Published in Print: 2016-9-1

©2016 by De Gruyter Mouton

Downloaded on 28.3.2024 from https://www.degruyter.com/document/doi/10.1515/pthp-2016-0009/html
Scroll to top button