Performance of the XN-2000 WPC channel-flagging to differentiate reactive and neoplastic leukocytosis

Peter Schuff-Werner 1 , Peter Kohlschein 2 , Aliaksandra Maroz 3 , Joachim Linssen 3 , Katrin Dreißiger 1 ,  and Christine Burstein 1
  • 1 Rostock University Medical Center, Institute of Clinical Chemistry and Laboratory Medicine, Germany
  • 2 Laborarztpraxis Osnabrück, Georgsmarienhütte, Germany
  • 3 Sysmex Europe GmbH, Norderstedt, Germany
Peter Schuff-Werner, Peter Kohlschein, Aliaksandra Maroz, Joachim Linssen, Katrin Dreißiger and Christine Burstein

Abstract

Background:

The distinction between reactive and neoplastic leukocytes, especially atypical lymphocytes suspected to be reactive or neoplastic, is a particular challenge in automated hematological cell differentiation. The aim of the study was to evaluate the performance of the XN analyzer supplemented with the WPC channel for differentiating between reactive and neoplastic leukocytosis.

Methods:

Blood samples of 253 patients with viral infections, lymphoma or leukemia were analyzed by the Sysmex XN-2000 analyzer equipped with the WPC channel. The results were compared to routine leukocyte differentiation using the routine Sysmex XE-2100 analyzer and automated digital microscopy (DM96). The combined information from standard morphology, immune phenotyping and clinical diagnosis served as a reference.

Results:

The XN WPC channel demonstrated an excellent performance for differentiating neoplastic (AUC=0.933) and reactive leukocytosis (AUC=0.900) as compared to morphological smear examination (AUC=0.949 and AUC=0.968, respectively) or to the differentiation results of our routine hematology analyzer (AUC=0.630 and AUC=0.635, respectively).

Conclusions:

Our data show that the combined WDF/WPC of the Sysmex XN-Series analyzer is advantageous in the automated differentiation of neoplastic and reactive leukocytosis, thus supporting the correct diagnostic decision in the daily laboratory routine.

  • 1.

    Simon MW. The atypical lymphocyte. Int Pediatr 2003;18:20–2.

  • 2.

    Baurmann H, Bettelheim P, Diem H, Gassmann W, Nebe T. Lymphocyte morphology in the peripheral blood film: proposal of a revised nomenclature and systematics. Lab Med 2011;35:261–70.

  • 3.

    Zini G, Bain B, Bettelheim P, Cortez J, d’Onofrio G, Faber E, et al. A European consensus report on blood cell identification: terminology utilized and morphological diagnosis concordance among 28 experts from 17 countries within the European LeukemiaNet network WP10, on behalf of the ELN Morphology Faculty. Br J Haematol 2010;151:359–64.

    • Crossref
    • Export Citation
  • 4.

    George TI. Malignant or benign leukocytosis. In: Hematology Education: the education program for annual congress of the European Hematology Association, The Hague, Netherlands: European Hematology Association. 2012;2:475–84.

  • 5.

    van der Meer W, Scott CS, de Keijzer MH. Automated flagging influences the inconsistency and bias of band cell and atypical lymphocyte morphological differentials. Clin Chem Lab Med 2004;42:371–7.

    • PubMed
    • Export Citation
  • 6.

    Barnes PW, McFadden SL, Machin SJ, Simson E. The international consensus group for hematology review: suggested criteria for action following automated CBC and WBC differential analysis. Lab Hematol 2005;11:83–90.

    • Crossref
    • Export Citation
  • 7.

    Briggs C, Longair I, Kumar P, Singh D, Machin SJ. Performance evaluation of the Sysmex haematology XN modular system. J Clin Pathol 2012;65:1024–30.

    • Crossref
    • PubMed
    • Export Citation
  • 8.

    Jones AS, Tailor H, Liesner R, Machin SJ, Briggs CJ. The value of the white precursor cell channel (WPC) on the Sysmex XN-1000 analyser in a specialist paediatric hospital. J Clin Pathol 2015;68:161–5.

    • Crossref
    • Export Citation
  • 9.

    Bruegel M, Nagel D, Funk M, Fuhrmann P, Zander J, Teupser D. Comparison of five automated hematology analyzers in a university hospital setting: Abbott Cell-Dyn Sapphire, Beckman Coulter DxH 800, Siemens Advia 2120i, Sysmex XE-2100, and Sysmex XN-2000. Clin Chem Lab Med 2015;53:1057–71.

  • 10.

    Seo J, Lee S, Kim S. Performance evaluation of the new hematology analyzer Sysmex XN-series. Int J Lab Hematol 2015;37:155–64.

    • Crossref
    • PubMed
    • Export Citation
  • 11.

    World Health Organization. International classification of diseases and delated health problems, 10th edition. 2010 [cited 2014]. Available from: http://www.who.int/classifications/icd/ICD10Volume2_en_2010.pdf.

  • 12.

    Swerdlow SH, Campo E, Harris NL, Jaffe ES, Pileri SA, editors. WHO classification of tumours of haematopoietic and lymphoid tissues, 4th ed. IARC: Lyon, 2008:439.

  • 13.

    Ruzicka K, Veitl M, Thalhammer-Scherrer R, Schwarzinger I. The new hematology analyzer Sysmex XE-2100: performance evaluation of a novel white blood cell differential technology. Arch Pathol Lab Med 2001;125:391–6.

    • PubMed
    • Export Citation
  • 14.

    Takagi Y, Kono M, Yamamoto S, Wada A, Morikawa T. Comparison of optical data from flow cytometry and microscopy of leukocytes after exposure to specific reagents. Microscopy (Oxf) 2015;64:305–10.

    • PubMed
    • Export Citation
  • 15.

    Gottfried EL. Lipids of human leukocytes: relation to celltype. J Lipid Res 1967;8:321–7.

    • PubMed
    • Export Citation
  • 16.

    Kuliszkiewicz-Janus M, Tuz MA, Kielbinski M, Jazwiec B, Niedoba J, Baczynski S. 31P MRS analysis of the phospholipid composition of the peripheral blood mononuclear cells (PBMC) and bone marrow mononuclear cells (BMMC) of patients with acute leukemia (AL). Cell Mol Biol Lett 2009;14:35–45.

    • PubMed
    • Export Citation
  • 17.

    El Kirat K, Morandat S. Cholesterol modulation of membrane resistance to Triton X-100 explored by atomic force microscopy. Biochim Biophys Acta 2007;1768:2300–9.

    • Crossref
    • PubMed
    • Export Citation
  • 18.

    Kawauchi S, Kono M, Takagi Y, Wada A, Morikawa T. The positions of normal leukocytes on the scattergram of the newly developed abnormal cell-detection channel of the XN-Series multi-parameter automated hematology analyzers. Sysmex J Int 2013;23:1–9.

Purchase article
Get instant unlimited access to the article.
$42.00
Log in
Already have access? Please log in.


or
Log in with your institution

Journal + Issues

Clinical Chemistry and Laboratory Medicine ( CCLM) publishes articles on novel teaching and training methods applicable to laboratory medicine. CCLM welcomes contributions on the progress in fundamental and applied research and cutting-edge clinical laboratory medicine. It is one of the leading journals in the field, with an impact factor of over three. CCLM is the official journal of nine national clinical societies and associated with EFLM.

Search